Dosimetry in Micro-computed Tomography: a Review of the Measurement Methods, Impacts, and Characterization of the Quantum GX Imaging System

PurposeX-ray micro-computed tomography (μCT) is a widely used imaging modality in preclinical research with applications in many areas including orthopedics, pulmonology, oncology, cardiology, and infectious disease. X-rays are a form of ionizing radiation and, therefore, can potentially induce damage and cause detrimental effects. Previous reviews have touched on these effects but have not comprehensively covered the possible implications on study results. Furthermore, interpreting data across these studies is difficult because there is no widely accepted dose characterization methodology for preclinical μCT. The purpose of this paper is to ensure in vivo μCT studies can be properly designed and the data can be appropriately interpreted.ProceduresStudies from the scientific literature that investigate the biological effects of radiation doses relevant to μCT were reviewed. The different dose measurement methodologies used in the peer-reviewed literature were also reviewed. The CT dose index 100 (CTDI100) was then measured on the Quantum GX μCT instrument. A low contrast phantom, a hydroxyapatite phantom, and a mouse were also imaged to provide examples of how the dose can affect image quality.ResultsData in the scientific literature indicate that scenarios exist where radiation doses used in μCT imaging are high enough to potentially bias experimental results. The significance of this effect may relate to the study outcome and tissue being imaged. CTDI100 is a reasonable metric to use for dose characterization in μCT. Dose rates in the Quantum GX vary based on the amount of material in the beam path and are a function of X-ray tube voltage. The CTDI100 in air for a Quantum GX can be as low as 5.1 mGy for a 50 kVp scan and 9.9 mGy for a 90 kVp scan. This dose is low enough to visualize bone both in a mouse image and in a hydroxyapatite phantom, but applications requiring higher resolution in a mouse or less noise in a low-contrast phantom benefit from longer scan times with increased dose.ConclusionsDose management should be considered when designing μCT studies. Dose rates in the Quantum GX are compatible with longitudinal μCT imaging.

[1]  L. Suva,et al.  Rapid Loss of Bone Mass and Strength in Mice after Abdominal Irradiation , 2011, Radiation research.

[2]  J Daniel Bourland,et al.  Risedronate prevents early radiation-induced osteoporosis in mice at multiple skeletal locations. , 2010, Bone.

[3]  P. Logan,et al.  Radiation-induced changes in bone. , 1998, Radiographics : a review publication of the Radiological Society of North America, Inc.

[4]  Gerald C. O'Sullivan,et al.  High Resolution In Vivo Bioluminescent Imaging for the Study of Bacterial Tumour Targeting , 2012, PloS one.

[5]  Nancy L Ford,et al.  Investigating the effect of longitudinal micro-CT imaging on tumour growth in mice , 2011, Physics in medicine and biology.

[6]  H. Yonekawa,et al.  Intestinal tumorigenesis in Min mice is enhanced by X-irradiation in an age-dependent manner. , 2005, Journal of radiation research.

[7]  I. Kashiwakura,et al.  Effects of ionizing radiation on differentiation of murine bone marrow cells into mast cells , 2015, Journal of radiation research.

[8]  O. Kovalchuk,et al.  Low dose irradiation profoundly affects transcriptome and microRNAme in rat mammary gland tissues , 2014, Oncoscience.

[9]  B. Cornelissen,et al.  Micro-CT for Anatomic Referencing in PET and SPECT: Radiation Dose, Biologic Damage, and Image Quality , 2011, The Journal of Nuclear Medicine.

[10]  D W Holdsworth,et al.  Fundamental image quality limits for microcomputed tomography in small animals. , 2003, Medical physics.

[11]  C. Cordon-Cardo,et al.  Impact of stromal sensitivity on radiation response of tumors implanted in SCID hosts revisited. , 2010, Cancer research.

[12]  Andre Obenaus,et al.  Radiation dose in rodent tissues during micro-CT imaging , 2004 .

[13]  Hristo N Nikolov,et al.  A quality assurance phantom for the performance evaluation of volumetric micro-CT systems , 2007, Physics in medicine and biology.

[14]  M. Wakabayashi,et al.  Different radiation susceptibility among five strains of mice detected by a skin reaction. , 2003, Journal of radiation research.

[15]  D. Yee,et al.  The Influence of Therapeutic Radiation on the Patterns of Bone Remodeling in Ovary-Intact and Ovariectomized Mice , 2013, Calcified Tissue International.

[16]  S. Judex,et al.  Low level irradiation in mice can lead to enhanced trabecular bone morphology , 2014, Journal of Bone and Mineral Metabolism.

[17]  K. Cengel,et al.  PTH prevents the adverse effects of focal radiation on bone architecture in young rats. , 2013, Bone.

[18]  E. Medico,et al.  Induction of MET by ionizing radiation and its role in radioresistance and invasive growth of cancer. , 2011, Journal of the National Cancer Institute.

[19]  Christoph Groden,et al.  Application of micro-CT in small animal imaging. , 2010, Methods.

[20]  K. Murase,et al.  Detection and Early Phase Assessment of Radiation-Induced Lung Injury in Mice Using Micro-CT , 2012, PloS one.

[21]  Jung Il Lee,et al.  Development of a small animal model to simulate clinical stereotactic body radiotherapy-induced central and peripheral lung injuries , 2014, Journal of radiation research.

[22]  L. Ghisolfi,et al.  Ionizing Radiation Induces Stemness in Cancer Cells , 2012, PloS one.

[23]  C T Badea,et al.  Micro-CT of rodents: state-of-the-art and future perspectives. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[24]  G. Suzuki,et al.  Expression of IL-1β mRNA in Mice after Whole Body X-Irradiation , 1995 .

[25]  Rik Huiskes,et al.  No effects of in vivo micro‐CT radiation on structural parameters and bone marrow cells in proximal tibia of wistar rats detected after eight weekly scans , 2007, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[26]  Ewa M Nowosielska,et al.  Immunological Mechanism of the Low-Dose Radiation-Induced Suppression of Cancer Metastases in a Mouse Model , 2010, Dose-response : a publication of International Hormesis Society.

[27]  Wolfhard Semmler,et al.  Gating in small-animal cardio-thoracic CT. , 2010, Methods.

[28]  Erik Tryggestad,et al.  Development of a novel preclinical pancreatic cancer research model: bioluminescence image-guided focal irradiation and tumor monitoring of orthotopic xenografts. , 2012, Translational oncology.

[29]  F. Slack,et al.  MicroRNAs in the ionizing radiation response and in radiotherapy. , 2013, Current opinion in genetics & development.

[30]  M. Goldman Ionizing radiation and its risks. , 1982, The Western journal of medicine.

[31]  Todd E Peterson,et al.  Longitudinal live animal micro-CT allows for quantitative analysis of tumor-induced bone destruction. , 2011, Bone.

[32]  Roel Van Holen,et al.  Low-Dose Micro-CT Imaging for Vascular Segmentation and Analysis Using Sparse-View Acquisitions , 2013, PloS one.

[33]  D J Brenner,et al.  Construction of mouse phantoms from segmented CT scan data for radiation dosimetry studies , 2015, Physics in medicine and biology.

[34]  Carl Virtanen,et al.  In Vivo Optical Imaging of Tumor and Microvascular Response to Ionizing Radiation , 2012, PloS one.

[35]  Dianna D Cody,et al.  In vivo respiratory-gated micro-CT imaging in small-animal oncology models. , 2004, Molecular imaging.

[36]  D. Parker Optimal short scan convolution reconstruction for fan beam CT , 1982 .

[37]  R Bogaerts,et al.  Development of micro-CT protocols for in vivo follow-up of mouse bone architecture without major radiation side effects. , 2011, Bone.

[38]  T. Chou,et al.  Immunotherapy of cancer. , 1970, British medical journal.

[39]  S. Powell,et al.  Repair of radiation damage to DNA , 2004, British Journal of Cancer.

[40]  H Weinans,et al.  Detecting and tracking local changes in the tibiae of individual rats: a novel method to analyse longitudinal in vivo micro-CT data. , 2004, Bone.

[41]  D L Parker,et al.  Optimal short scan convolution reconstruction for fanbeam CT. , 1982, Medical physics.

[42]  I. Iordachita,et al.  Evaluation of a cone beam computed tomography geometry for image guided small animal irradiation , 2015, Physics in medicine and biology.

[43]  John M. Boone,et al.  Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography , 2010 .

[44]  Matthew J Allen,et al.  Local irradiation alters bone morphology and increases bone fragility in a mouse model. , 2010, Journal of biomechanics.

[45]  B. Rice,et al.  Monitoring Bacterial Burden, Inflammation and Bone Damage Longitudinally Using Optical and μCT Imaging in an Orthopaedic Implant Infection in Mice , 2012, PloS one.

[46]  M. Barcellos-Hoff,et al.  Spatiotemporal characterization of ionizing radiation induced DNA damage foci and their relation to chromatin organization. , 2010, Mutation research.

[47]  T H Farncombe Software-based respiratory gating for small animal conebeam CT. , 2008, Medical physics.

[48]  Thomas L. Smith,et al.  Total-body irradiation produces late degenerative joint damage in rats , 2014, International journal of radiation biology.

[49]  M. Story,et al.  Time, Dose, and Volume Responses in a Mouse Pulmonary Injury Model Following Ablative Irradiation , 2016, Lung.

[50]  T. Bateman,et al.  Single‐Limb Irradiation Induces Local and Systemic Bone Loss in a Murine Model , 2015, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[51]  M J Paulus,et al.  High resolution X-ray computed tomography: an emerging tool for small animal cancer research. , 2000, Neoplasia.

[52]  J. Kiang,et al.  Skin wound trauma, following high-dose radiation exposure, amplifies and prolongs skeletal tissue loss. , 2015, Bone.

[53]  M McNitt‐Gray MO‐A‐ValB‐01: Tradeoffs in Image Quality and Radiation Dose for CT , 2006 .

[54]  William H. Judy,et al.  The Measurement, Reporting, and Management of Radiation Dose in CT , 2008 .

[55]  B. Marples,et al.  Exposure to low dose ionising radiation: Molecular and clinical consequences. , 2014, Cancer letters.

[56]  Takafumi Koike,et al.  Evaluation of X-ray doses and their corresponding biological effects on experimental animals in cone-beam micro-CT scans (R-mCT2) , 2015, Radiological Physics and Technology.

[57]  T. Bateman,et al.  A murine model for bone loss from therapeutic and space-relevant sources of radiation. , 2006, Journal of applied physiology.

[58]  J. Boone,et al.  Small-animal X-ray dose from micro-CT. , 2004, Molecular imaging.

[59]  S. Figueroa,et al.  TLD assessment of mouse dosimetry during microCT imaging. , 2008, Medical physics.

[60]  Ivailo E. Petrov,et al.  X-ray dose delivered during a longitudinal micro-CT study has no adverse effect on cardiac and pulmonary tissue in C57BL/6 mice , 2013, Acta radiologica.

[61]  J. H. Hubbell,et al.  Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest , 1995 .

[62]  Patrick L Chow,et al.  Monte carlo simulations of dose from microCT imaging procedures in a realistic mouse phantom. , 2005, Medical physics.

[63]  S. Hazra,et al.  The impact of ageing on adipose structure, function and vasculature in the B6D2F1 mouse: evidence of significant multisystem dysfunction , 2014, The Journal of physiology.

[64]  J. Carson,et al.  Detecting radiation-induced injury using rapid 3D variogram analysis of CT images of rat lungs. , 2013, Academic radiology.

[65]  P. Bruyndonckx,et al.  Longitudinal in vivo microcomputed tomography of mouse lungs: No evidence for radiotoxicity. , 2015, American journal of physiology. Lung cellular and molecular physiology.

[66]  S. Russell,et al.  Small Animal Absorbed Radiation Dose from Serial Micro-Computed Tomography Imaging , 2007, Molecular Imaging and Biology.

[67]  P. Jeggo,et al.  Radiation-induced DNA damage responses. , 2006, Radiation protection dosimetry.

[68]  A. Balmain,et al.  Guidelines for the welfare and use of animals in cancer research , 2010, British Journal of Cancer.

[69]  N. Sharma,et al.  DNA Damage Response Proteins and Oxygen Modulate Prostaglandin E2 Growth Factor Release in Response to Low and High LET Ionizing Radiation , 2015, Front. Oncol..

[70]  Emil Y. Sidky,et al.  Algorithm-Enabled Low-Dose Micro-CT Imaging , 2011, IEEE Transactions on Medical Imaging.

[71]  Raymond P. Rossi,et al.  Specification and Acceptance Testing of Computed Tomography Scanners , 1993 .

[72]  G. Allan Johnson,et al.  Left ventricle volume measurements in cardiac micro-CT: The impact of radiation dose and contrast agent , 2008, Comput. Medical Imaging Graph..

[74]  Edgar Ben-Josef,et al.  In vivo bioluminescent imaging of irradiated orthotopic pancreatic cancer xenografts in nonobese diabetic-severe combined immunodeficient mice: a novel method for targeting and assaying efficacy of ionizing radiation. , 2010, Translational oncology.

[75]  Characterization of X-ray Dose in Murine Animals Using microCT, a New Low-Dose Detector and nanoDot Dosimeters , 2012, PloS one.

[76]  Steven K Boyd,et al.  Radiation effects on bone architecture in mice and rats resulting from in vivo micro-computed tomography scanning. , 2008, Medical engineering & physics.

[77]  A. Taghian,et al.  Tumors arising in SCID mice share enhanced radiation sensitivity of SCID normal tissues. , 1992, Cancer research.

[78]  G Allan Johnson,et al.  Dual-energy micro-computed tomography imaging of radiation-induced vascular changes in primary mouse sarcomas. , 2013, International journal of radiation oncology, biology, physics.

[79]  P Zygmanski,et al.  Comprehensive quality assurance phantom for the small animal radiation research platform (SARRP). , 2015, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[80]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[81]  J. Kappes,et al.  Breast cancer metastasis to bone: evaluation of bioluminescent imaging and microSPECT/CT for detecting bone metastasis in immunodeficient mice , 2007, Clinical & Experimental Metastasis.

[82]  Daniel Kolditz,et al.  Dosimetry concepts for scanner quality assurance and tissue dose assessment in micro-CT. , 2012, Medical physics.

[83]  Alain Seret,et al.  Performance Evaluation and X-ray Dose Quantification for Various Scanning Protocols of the GE eXplore 120 Micro-CT , 2013, IEEE Transactions on Nuclear Science.

[84]  S. Hui,et al.  Characterization of Rotating Gantry Micro-CT Configuration for the In Vivo Evaluation of Murine Trabecular Bone , 2013, Microscopy and Microanalysis.

[85]  ICRU Report No. 87: Radiation dose and image-quality assessment in computed tomography. , 2012, Journal of the ICRU.

[86]  A. Giaccia,et al.  scid mutation in mice confers hypersensitivity to ionizing radiation and a deficiency in DNA double-strand break repair. , 1991, Proceedings of the National Academy of Sciences of the United States of America.

[87]  José Rino,et al.  Low Doses of Ionizing Radiation Promote Tumor Growth and Metastasis by Enhancing Angiogenesis , 2010, PloS one.

[88]  Rudi Deklerck,et al.  Evaluation of the radiation dose in micro-CT with optimization of the scan protocol. , 2010, Contrast media & molecular imaging.

[89]  R K Sachs,et al.  Radiation-induced cancer: a modern view. , 2012, The British journal of radiology.

[90]  W Semmler,et al.  Initial Experience with a Novel Low-Dose Micro-CT System , 2007, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.

[91]  M. Bala,et al.  Concerted action of Nrf2-ARE pathway, MRN complex, HMGB1 and inflammatory cytokines - Implication in modification of radiation damage , 2014, Redox biology.

[92]  Jiang Hsieh,et al.  Computed Tomography: Principles, Design, Artifacts, and Recent Advances, Fourth Edition , 2022 .

[93]  Ted A. Bateman,et al.  Early Increase in Osteoclast Number in Mice after Whole-Body Irradiation with 2 Gy X Rays , 2008, Radiation research.

[94]  M. Galanski,et al.  Phantom and cadaver measurements of dose and dose distribution in micro-CT of the chest in mice , 2011, Acta radiologica.

[95]  L. Caldas,et al.  A simple method for evaluation of half-value layer variation in CT equipment , 2006, Physics in medicine and biology.

[96]  Francesco De Carlo,et al.  Long-Term Dose Response of Trabecular Bone in Mice to Proton Radiation , 2008, Radiation research.

[97]  M. Löbrich,et al.  DNA Double-Strand Break Repair of Blood Lymphocytes and Normal Tissues Analysed in a Preclinical Mouse Model: Implications for Radiosensitivity Testing , 2008, Clinical Cancer Research.

[98]  Sayan Mukherjee,et al.  Assessing the Radiation Response of Lung Cancer with Different Gene Mutations Using Genetically Engineered Mice , 2013, Front. Oncol..

[99]  Hae-June Lee,et al.  Establishment of a murine model for radiation-induced bone loss using micro-computed tomography in adult C3H/HeN mice , 2013, Laboratory animal research.