A motion artifact generation and assessment system for the rapid testing of surface biopotential electrodes.

Dry electrodes can reduce cost while increasing the usability and comfort of wearable monitoring systems. They are, however, susceptible to motion artifacts. The present electrode testing methods lack reliability and do not separate the factors that affect the motion artifact. In this paper, we introduce a first generation motion artifact generation and assessment system that generates the speed, amplitude, and pattern-wise programmable movement of the electrode. The system simultaneously measures electrode-skin impedance, the motion artifact, and one channel of an electrocardiogram that contains the motion artifact and monitors the mounting force applied to the electrode. We demonstrate the system by comparing the applied movement and the measured signals for electrode movements up to 6 mm and movement frequencies from 0.4 Hz to 4 Hz. Results show that the impedance change and surface potential are visually clearly related to the applied motion, with average correlations of 0.89 and 0.64, respectively. The applied force, electrode location, and electrode structure all affect the motion artifact. The setup enables the motion of the electrode to be accurately controlled. The system can be used as a precursor to the testing of integrated systems because it enables thorough, repeatable, and robust motion artifact studies. The system allows a deeper insight into motion artifacts and the interplay of the various factors that affect them.

[1]  J. Webster,et al.  Minimizing Electrode Motion Artifact by Skin Abrasion , 1977, IEEE Transactions on Biomedical Engineering.

[2]  K. Kearney,et al.  Quantification of Motion Artifact in ECG Electrode Design , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[3]  John G. Webster,et al.  Electrocardiographic motion artifact versus electrode impedance , 2005, IEEE Transactions on Biomedical Engineering.

[4]  S. Leonhardt,et al.  Characterization of textile electrodes and conductors using standardized measurement setups , 2010, Physiological measurement.

[5]  J. Webster,et al.  Dry electrodes for electrocardiography , 2013, Physiological measurement.

[6]  J. Webster,et al.  The origin of skin-stretch-caused motion artifacts under electrodes. , 1996, Physiological measurement.

[7]  Kang-Hwi Lee,et al.  Heart Monitoring Garments Using Textile Electrodes for Healthcare Applications , 2009, Journal of Medical Systems.

[8]  S Odman Potential and impedance variations following skin deformation. , 1981, Medical & biological engineering & computing.

[9]  A. Gruetzmann,et al.  Novel dry electrodes for ECG monitoring , 2007, Physiological measurement.

[10]  Willis J. Tompkins,et al.  A Real-Time QRS Detection Algorithm , 1985, IEEE Transactions on Biomedical Engineering.

[11]  J P Clarys,et al.  Electromyography in sports and occupational settings: an update of its limits and possibilities , 2000, Ergonomics.

[12]  Sunyoung Kim,et al.  Correlation Between Electrode-Tissue Impedance and Motion Artifact in Biopotential Recordings , 2012, IEEE Sensors Journal.

[13]  E Marani,et al.  Low Frequency Changes in Skin Surface Potentials by Skin Compression: Experimental Results and Theories , 2003, Archives of physiology and biochemistry.

[14]  Wilhelm Stork,et al.  Reliable motion artifact detection for ECG monitoring systems with dry electrodes , 2008, 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[15]  Hui Zhang,et al.  A dynamic measurement system for evaluating dry bio-potential surface electrodes , 2013 .

[16]  Mario Lamontagne,et al.  Application of Electromyography in Sport Medicine , 2001 .

[17]  J. Lauter,et al.  Wearable approach for continuous ECG - and activity patient-monitoring , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[18]  Jari Hyttinen,et al.  Effect of pressure and padding on motion artifact of textile electrodes , 2013, Biomedical engineering online.

[19]  Steffen Leonhardt,et al.  Skin Electrode Impedance of Textile Electrodes for Bioimpedance Spectroscopy , 2007 .

[20]  Wilhelm Stork,et al.  An approach to reliable motion artifact detection for mobile long-term ECG monitoring systems using dry electrodes , 2007 .

[21]  David A. Tong,et al.  Adaptive reduction of motion artifact in the electrocardiogram , 2002, Proceedings of the Second Joint 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society] [Engineering in Medicine and Biology.

[22]  L. Kirkup,et al.  A direct comparison of wet, dry and insulating bioelectric recording electrodes. , 2000, Physiological measurement.

[23]  A. Johansson,et al.  Electro-mechanical stability of surface EMG sensors , 2007, Medical & Biological Engineering & Computing.

[24]  Refet Firat Yazicioglu,et al.  Motion artifact reduction in ambulatory ECG monitoring: an integrated system approach , 2011, Wireless Health.

[25]  Maria Romano,et al.  Relevance of motion artifact in electromyography recordings during vibration treatment. , 2009, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[26]  P.S. Hamilton,et al.  Comparison of methods for adaptive removal of motion artifact , 2000, Computers in Cardiology 2000. Vol.27 (Cat. 00CH37163).

[27]  Tao Xiaoming,et al.  Measurement of wearable electrode and skin mechanical interaction using displacement and pressure sensors , 2011, 2011 4th International Conference on Biomedical Engineering and Informatics (BMEI).

[28]  S. Ödman,et al.  Movement-induced potentials in surface electrodes , 1982, Medical and Biological Engineering and Computing.