Study for imaging of inside bone using FM-chirp pulse compression system.
暂无分享,去创建一个
We measured the attenuation of ultrasound in animal bone in vitro, and determined successfully the correct dynamic range for the detection of echo signals from inside the bone. It was possible to display a B-mode image of a 6-mm thick porcine rib bone using a high frequency (5 MHz) transducer. However, it was difficult to obtain the image of a large bone (20-mm thick femur). This paper describes a possibility of the improvement for the detection of echoes from inside the bone using the FM-chirp pulse. An ultrasonic wave generator that generates the FM-chirp pulse for a frequency range of 100 KHz to 5 MHz was fabricated. And the pulse was transmitted from a transducer and received by a hydrophone in the water. The received waveform and the frequency spectrum were displayed on an oscilloscope. We measured the amplitude spectrum of the received waves before and after inserting a sample bone between the transducer and the hydrophone. In the experiment, three types of focused transducers (0.5, 1 and 2.25 MHz) and three kinds of sample bones (spine, femur and rib of a pig) were used. From the results of the measurement, we could find the useful information for imaging of inside the bone.