PCR-I-High Resolution Positron Tomograph Using Analog Coding

PCR-I has been designed and built to demonstrate the concept of analog coding to provide high resolution positron tomographic images without interpolative motion. The instrument uses 360 bismuth germanate (BGO) detectors coupled through a lucite light pipe to 90 phototubes. The measured resolution at the center is 4.8 mm. A straightforward extension of the concept will lead to a tomograph capable of obtaining resolution below 3 mm for low energy positron emitters. The sensitivity of PCR-I is 46 000/s for a 20 cm diameter phantom uniformly filled with 1 μCi/cc of a positron emitter. Parameters related to dead time and random and prompt scatter coincidences have been studied. Phantom measurements demonstrate the resolution and uniformity of PCR-I. Animal studies carried out to date include 18F in rat skeleton, 11C-palmitate and 11C-labeled red cells in dog heart and blood flow, and blood volume in monkey brain using C15O2 and C15O.

[1]  R. H. Huesman,et al.  DYNAMIC POSITRON EMISSION TOMOGRAPHY IN MAN USING SMALL BISMUTH GERMANATE CRYSTALS , 1982 .

[2]  One Dimensional Scintillation Cameras for Positron Ect Ring Detectors , 1981 .

[3]  Ronald H. Huesman,et al.  Imaging Properties of a Positron Tomograph with 280 Bgo Crystals , 1981, IEEE Transactions on Nuclear Science.

[4]  E. Tanaka,et al.  Design Consideration for a Continuously Rotating Positron Computed Tomograph , 1982, IEEE Transactions on Nuclear Science.

[5]  D. Chesler,et al.  A Positron Tomograph Employing a One Dimension BGO Scintillation Camera , 1983, IEEE Transactions on Nuclear Science.

[6]  C. A. Burnham,et al.  A Stationary Positron Emission Ring Tomograph Using Bgo Detector and Analog Readout , 1984, IEEE Transactions on Nuclear Science.

[7]  E. J. Hoffman,et al.  Investigation of a Technique for Providing a Pseudo-Continuous Detector Ring for Positron Tomography , 1982, IEEE Transactions on Nuclear Science.

[8]  S Aronow,et al.  A hybrid positron scanner. , 1970, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  Rodney A. Brooks,et al.  Design Considerations for Positron Emission Tomography , 1981, IEEE Transactions on Biomedical Engineering.

[10]  T. Greitz,et al.  A Four Ring Positron Camera System for Emission Tomography of the Brain , 1982, IEEE Transactions on Nuclear Science.

[11]  Ronald H. Huesman,et al.  Imaging Properties of a Positron Tomograph with 280 Bgo Crystals , 1981, IEEE Transactions on Nuclear Science.

[12]  Gordon L. Brownell,et al.  A Multi-Crystal Positron Camera , 1972 .