Comparison of the TSI Model 8520 and Grimm Series 1.108 Portable Aerosol Instruments Used to Monitor Particulate Matter in an Iron Foundry

This study uses two real-time dust monitors, the TSI Model 8520 DustTrak and Grimm Series 1.108 Aerosol Spectrometer, to determine PM10 and PM2.5 levels simultaneously in an iron foundry. The SA Model 241 Dichotomous Sampler was used as a reference gravimetric method for comparing the measurement results obtained by these direct-reading instruments. The response to PM levels from DustTrak is higher than that of the Aerosol Spectrometer. The DustTrak provides an overestimation PM levels, and PM levels measured by an Aerosol Spectrometer are lower than actual concentrations. Calibration factors of the DustTrak and Aerosol Spectrometer are 0.74 and 1.33, respectively, when used to measure particulate matter at an iron foundry. Based on measurement results, the DustTrak provides a lower overestimation of PM10 levels than PM2.5 levels; that is, the response of the DustTrak increases as particle size decreases. In addition, measurement results suggest that the Aerosol Spectrometer provides precise measurements of PM10 and PM2.5, and measurement accuracy compared with the reference gravimetric method can be improved through a calibration factor.

[1]  Darrin K. Ott,et al.  Comparison of the Grimm 1.108 and 1.109 portable aerosol spectrometer to the TSI 3321 aerodynamic particle sizer for dry particles. , 2006, The Annals of occupational hygiene.

[2]  William A Heitbrink,et al.  The mapping of fine and ultrafine particle concentrations in an engine machining and assembly facility. , 2006, The Annals of occupational hygiene.

[3]  Martin Braniš,et al.  The contribution of ambient sources to particulate pollution in spaces and trains of the Prague underground transport system , 2006 .

[4]  Simon Kingham,et al.  Winter comparison of TEOM, MiniVol and DustTrak PM10 monitors in a woodsmoke environment , 2006 .

[5]  K. Pinkerton,et al.  Dust Generator for Inhalation Studies with Limited Amounts of Archived Particulate Matter , 2005 .

[6]  B. Tomkins,et al.  Development and Application of Protocols for the Determination of Response of Real-Time Particle Monitors to Common Indoor Aerosols , 2004, Journal of the Air & Waste Management Association.

[7]  Kerrie Mengersen,et al.  Characteristics of particle number and mass concentrations in residential houses in Brisbane, Australia , 2003 .

[8]  Gurumurthy Ramachandran,et al.  Characterizing Indoor and Outdoor 15 Minute Average PM 2.5 Concentrations in Urban Neighborhoods , 2003 .

[9]  C. Y. Chan,et al.  Commuter exposure to particulate matter in public transportation modes in Hong Kong , 2002 .

[10]  Phillip L. Williams,et al.  A comparison of two direct-reading aerosol monitors with the federal reference method for PM2.5 in indoor air , 2002 .

[11]  Laboratory and Field Evaluation of Measurement Methods for One-Hour Exposures to O3, PM25, and CO , 2001, Journal of the Air & Waste Management Association.

[12]  M. Kleeman,et al.  Comparison of Real-Time Instruments Used To Monitor Airborne Particulate Matter , 2001, Journal of the Air & Waste Management Association.

[13]  P. Catalano,et al.  Laboratory and field evaluation of measurement methods for one-hour exposures to O3, PM2.5, and CO. , 2001, Journal of the Air and Waste Management Association.

[14]  G Ramachandran,et al.  Comparison of Short-Term Variations (15-Minute Averages) in Outdoor and Indoor PM2.5 Concentrations , 2000, Journal of the Air & Waste Management Association.

[15]  A. Lehocky,et al.  Evaluation and field calibration of the Miniram PDM-3 aerosol monitor for measuring respirable and total coal dust. , 1999, American Industrial Hygiene Association journal.

[16]  James H. Vincent,et al.  Aerosol Instrumentation in Occupational Hygiene: An Historical Perspective , 1998 .

[17]  M. Kerker Light Scattering Instrumentation for Aerosol Studies: An Historical Overview , 1997 .

[18]  Phillip L. Williams,et al.  Comparison of Respirable Samplers to Direct-Reading Real-Time Aerosol Monitors for Measuring Coal Dust , 1996 .

[19]  Tung-Sheng Shih,et al.  Laboratory Testing of Three Direct Reading Dust Monitors , 1996 .

[20]  P. Görner,et al.  Photometer measurement of polydisperse aerosols , 1995 .

[21]  B. T. Chen,et al.  A New Technique to Calibrate Optical Particle Counters Aerodynamically , 1990 .

[22]  P. Kuusisto,et al.  Evaluation of the direct reading instruments for the measurement of aerosols , 1983 .

[23]  Winifred G. Palmer,et al.  Lung cancer in ferrous foundry workers: a review. , 1981, American Industrial Hygiene Association journal.