Factors Affecting Accuracy and Precision in PET Volume Imaging

Volume imaging positron emission tomographic (PET) scanners with no septa and a large axial acceptance angle offer several advantages over multiring PET scanners. A volume imaging scanner combines high sensitivity with fine axial sampling and spatial resolution. The fine axial sampling minimizes the partial volume effect, which affects the measured concentration of an object. Even if the size of an object is large compared to the slice spacing in a multiring scanner, significant variation in the concentration is measured as a function of the axial position of the object. With a volume imaging scanner, it is necessary to use a three-dimensional reconstruction algorithm in order to avoid variations in the axial resolution as a function of the distance from the center of the scanner. In addition, good energy resolution is needed in order to use a high energy threshold to reduce the coincident scattered radiation.

[1]  S. K. Hilal,et al.  Spherical positron emission tomograph (S-PET) I - performance analysis , 1984 .

[2]  M. Defrise,et al.  Three dimensional reconstruction of PET data from a multi-ring camera , 1989 .

[3]  David A. Chesler,et al.  Cylindrical PET detector design , 1988 .

[4]  D. Chesler,et al.  Simulation studies for cylindrical positron tomography , 1988 .

[5]  Paul Kinahan,et al.  The Theory of Three-Dimensional Image Reconstruction for PET , 1987, IEEE Transactions on Medical Imaging.

[6]  Christopher J. Thompson,et al.  The effect of collimation on scatter fraction in multi-slice PET , 1988 .

[7]  M. Daube-Witherspoon,et al.  Treatment of axial data in three-dimensional PET. , 1987, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[8]  M. Daube-Witherspoon,et al.  An Iterative Image Space Reconstruction Algorthm Suitable for Volume ECT , 1986, IEEE Transactions on Medical Imaging.

[9]  G Muehllehner,et al.  The high count rate performance of a two-dimensionally position-sensitive detector for positron emission tomography. , 1989, Physics in medicine and biology.

[10]  D. Mankoff,et al.  Continuous-slice PENN-PET: a positron tomograph with volume imaging capability. , 1990, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  E. Hoffman,et al.  Evaluating the Performance of Multiplane Positron Tomographs Designed for Brain Imaging , 1982, IEEE Transactions on Nuclear Science.

[12]  Rodney A. Brooks,et al.  Design Considerations for Positron Emission Tomography , 1981, IEEE Transactions on Biomedical Engineering.