Computational modelling of long bone fractures fixed with locking plates - How can the risk of implant failure be reduced?

BACKGROUND AND PURPOSE The Locking Compression Plate (LCP) is part of a new plate generation requiring an adapted surgical technique and new thinking about commonly used concepts of internal fixation using plates. Knowledge of the fixation stability provided by these new plates is very limited and clarification is still necessary to determine how the mechanical stability and the risk of implant failure can best be controlled. METHODS Upon validation, a finite element model of an LCP attached to a cylinder was developed to simulate and analyse the biomechanics of a transverse long bone fracture fixed with a locking plate. Of special interest were the factors influencing the mechanical conditions at the fracture site, the control of interfragmentary movement and implant failure. RESULTS Several factors were shown to influence stability in compression. Increasing translation and/or fracture angle post fixation reduced construct stability. Axial stiffness was also influenced by the working length and plate-bone distance. The fracture gap had no effect on the construct stability when no bone contact occurred during loading. Stress analysis of the LCP demonstrated that the maximum Von Mises stresses were found in the innermost screws at the screw-head junction. INTERPRETATION For the clinical use of the LCP as a locked internal fixator in fractures with an interfragmentary gap of 1 mm, at least two to four plate holes near the fracture gap should be omitted to allow fracture motion and bone contact to occur. This will also achieve a larger area of stress distribution on the plate and reduce the likelihood of fatigue failure due to cyclic loading.

[1]  P. Comte,et al.  Influence of Unoccupied Holes on the Fatigue Behaviour of Bone Fixation Plates , 1985 .

[2]  W. Ramotowski,et al.  Zespol. An original method of stable osteosynthesis. , 1991, Clinical orthopaedics and related research.

[3]  J. Cordey,et al.  Mechanical analysis of the bone to plate interface of the LC-DCP and of the PC-FIX on human femora. , 2000, Injury.

[4]  F. Kummer,et al.  Schuhli augmentation of plate and screw fixation for humeral shaft fractures: a laboratory study. , 1999, Journal of orthopaedic trauma.

[5]  P. Regazzoni,et al.  LISS-Traktor Kombination des “less invasive stabilization systems” (LISS) mit dem AO-Distraktor für distale Femur- und proximale Tibiafrakturen , 2001, Der Unfallchirurg.

[6]  C. Sommer,et al.  Guidelines for the clinical application of the LCP. , 2003, Injury.

[7]  I. Learmonth,et al.  "Biological" internal fixation of long bone fractures: a biomechanical study of a "noncontact" plate system. , 1998, Injury.

[8]  R. Nanda,et al.  Biomechanical testing of the locking compression plate: when does the distance between bone and implant significantly reduce construct stability? , 2007, Injury.

[9]  E. Schemitsch,et al.  Influence of the number of cortices on the stiffness of plate fixation of diaphyseal fractures. , 1999 .

[10]  S. Gilbert,et al.  The development of the distal femur Less Invasive Stabilization System (LISS). , 2001, Injury.

[11]  Ulrich Dieter,et al.  Biomechanical testing of the LCP--how can stability in locked internal fixators be controlled? , 2003, Injury.

[12]  M. Zlowodzki,et al.  Less Invasive Stabilization System (LISS) for fractures of the proximal tibia: indications, surgical technique and preliminary results of the UMC Clinical Trial. , 2003, Injury.

[13]  John M. O’Byrne,et al.  Locking compression plate breakage and fracture non-union: a finite element study of three patient-specific cases , 2012, European Journal of Orthopaedic Surgery & Traumatology.

[14]  E. Chao,et al.  The effects of physiologic dynamic compression on bone healing under external fixation. , 1990, Clinical orthopaedics and related research.

[15]  B. Hanson,et al.  Locking compression plate loosening and plate breakage: a report of four cases. , 2004, Journal of orthopaedic trauma.

[16]  Ching-Kong Chao,et al.  Increasing bending strength of tibial locking screws: mechanical tests and finite element analyses. , 2007, Clinical biomechanics.

[17]  D. Pichora,et al.  Finite element stress analysis of a hybrid fracture fixation plate. , 1996, Medical engineering & physics.