Design studies for a PET detector module using a PIN photodiode to measure depth of interaction

We present design studies of a multi-layer PET detector module that uses an 8/spl times/8 array of 3 mm square PIN photodiodes to both identify the crystal of interaction and measure the depth of interaction. Each photodiode is coupled to one end of a 3/spl times/3/spl times/30 mm BGO crystal, with the opposite ends of 64 such crystals attached to a single 1" square photomultiplier tube that provides a timing signal and energy discrimination. Each BGO crystal is coated with a lossy reflector, so the ratio of light detected in the photodiode and photomultiplier tube depends on the interaction depth in the crystal, and is used to determine this depth of interaction on an event by event basis, A test module with one 3/spl times/3/spl times/30 mm BGO crystal, one 3 mm square PIN photodiode, and one photomultiplier tube is operated at -20/spl deg/C with an amplifier peaking time of 4 /spl mu/s, and a depth of interaction resolution of 5 to 8 mm fwhm measured. Simulations predict that this virtually eliminates radial elongation in a 60 cm diameter BGO tomograph. The photodiode signal corresponding to 511 keV energy deposit varies linearly with excitation position, ranging from 1250 electrons (e/sup -/) at the end closest to the photodiode to 520 e/sup -/ at the opposite end. The electronic noise is a position independent 330 e/sup -/ fwhm, so the signal to noise ratio is sufficient to reliably identify the crystal of interaction in a 64 element module. >

[1]  V. Perez-Mendez,et al.  Design Considerations for a High Spatial Resolution Positron Camera with Dense Drift Space MWPC's , 1982, IEEE Transactions on Nuclear Science.

[2]  P. Bartzakos,et al.  A depth-encoded PET detector , 1990 .

[3]  C. Melcher,et al.  Temperature Dependence of Fluorescence Decay Time and Emission Spectrum of Bismuth Germanate , 1985, IEEE Transactions on Nuclear Science.

[4]  R. Nutt,et al.  Performance of a PET detector module utilizing an array of silicon photodiodes to identify the crystal of interaction , 1992 .

[5]  Bojan T. Turko,et al.  Initial characterization of a position-sensitive photodiode/BGO detector for PET (positron emission tomography) , 1988 .

[6]  D. P. Saylor,et al.  Design of an efficient position sensitive gamma ray detector for nuclear medicine. , 1986, Physics in medicine and biology.

[7]  C. Melcher,et al.  Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillator , 1991, Conference Record of the 1991 IEEE Nuclear Science Symposium and Medical Imaging Conference.

[8]  Roger Lecomte,et al.  Design of a high resolution positron emission tomograph using solid state scintillation detectors , 1988 .

[9]  Wai-Hoi Wong,et al.  Designing a Stratified Detection System for PET Cameras , 1986, IEEE Transactions on Nuclear Science.

[10]  Joel S. Karp,et al.  Depth-of-interaction determination in NaI(Tl) and BGO scintillation crystals using a temperature gradient , 1987 .

[11]  K. Shimizu,et al.  Development of 3-D detector system for positron CT , 1988 .

[12]  K. Shimizu,et al.  High resolution block detectors for PET , 1990 .