Measuring retinal blood flow in rats using Doppler optical coherence tomography without knowing eyeball axial length.

PURPOSE Doppler optical coherence tomography (OCT) is widely used for measuring retinal blood flow. Existing Doppler OCT methods require the eyeball axial length, in which empirical values are usually used. However, variations in the axial length can create a bias unaccounted for in the retinal blood flow measurement. The authors plan to develop a Doppler OCT method that can measure the total retinal blood flow rate without requiring the eyeball axial length. METHODS The authors measured the retinal blood flow rate using a dual-ring scanning protocol. The small and large scanning rings entered the eye at different incident angles (small ring: 4°; large ring: 6°), focused on different locations on the retina, and detected the projected velocities/phase shifts along the probing beams. The authors calculated the ratio of the projected velocities between the two rings, and then used this ratio to estimate absolute flow velocity. The authors tested this method in both Intralipid phantoms and in vivo rats. RESULTS In the Intralipid flow phantom experiments, the preset and measured flow rates were consistent with the coefficient of determination as 0.97. Linear fitting between preset and measured flow rates determined the fitting slope as 1.07 and the intercept as -0.28. In in vivo rat experiments, the measured average total retinal blood flow was 7.02 ± 0.31 μl/min among four wild-type rats. The authors' measured flow rates were consistent with results in the literature. CONCLUSIONS By using a dual-ring scanning protocol with carefully controlled incident angle difference between the two scanning rings in Doppler OCT, the authors demonstrated that it is feasible to measure the absolute retinal blood flow without knowing the eyeball axial length.

[1]  Ruikang K. Wang,et al.  Optical microangiography of retina and choroid and measurement of total retinal blood flow in mice , 2012, Biomedical optics express.

[2]  Leopold Schmetterer,et al.  Angle independent flow assessment with bidirectional Doppler optical coherence tomography. , 2013, Optics letters.

[3]  David A Atchison,et al.  Eye shape in emmetropia and myopia. , 2004, Investigative ophthalmology & visual science.

[4]  T. Duong,et al.  Reduced ocular blood flow as an early indicator of diabetic retinopathy in a mouse model of diabetes. , 2012, Investigative ophthalmology & visual science.

[5]  J. Izatt,et al.  Retinal blood flow measurement by circumpapillary Fourier domain Doppler optical coherence tomography. , 2008, Journal of biomedical optics.

[6]  T. Grosvenor,et al.  Reduction in Axial Length with Age: An Emmetropizing Mechanism for the Adult Eye? , 1987, American journal of optometry and physiological optics.

[7]  Amiram Grinvald,et al.  INCREASED RETINAL BLOOD FLOW VELOCITY IN PATIENTS WITH EARLY DIABETES MELLITUS , 2012, Retina.

[8]  J. Fujimoto,et al.  Ultrahigh-resolution ophthalmic optical coherence tomography , 2001, Nature Medicine.

[9]  David Huang,et al.  Retinal blood flow detection in diabetic patients by Doppler Fourier domain optical coherence tomography. , 2009, Optics express.

[10]  Martin F. Kraus,et al.  Total retinal blood flow measurement with ultrahigh speed swept source/Fourier domain OCT , 2011, Biomedical optics express.

[11]  T. Gardner,et al.  Diabetic retinopathy: more than meets the eye. , 2002, Survey of ophthalmology.

[12]  W. Freeman,et al.  Retinal blood flow measurements in branch retinal vein occlusion using scanning laser Doppler flowmetry. , 1998, American journal of ophthalmology.

[13]  Carmen A. Puliafito,et al.  Automatic retinal blood flow calculation using spectral domain optical coherence tomography , 2007 .

[14]  T. Kenny,et al.  CORRIGENDUM: Quantum Limit of Quality Factor in Silicon Micro and Nano Mechanical Resonators , 2014, Scientific Reports.

[15]  Ji Yi,et al.  Visible-light optical coherence tomography for retinal oximetry. , 2013, Optics letters.

[16]  Savleen Kaur,et al.  Reduced ocular blood flow in asymmetric glaucoma: cause or effect? , 2014, International Ophthalmology.

[17]  Harsha Radhakrishnan,et al.  Multiparametric optical coherence tomography imaging of the inner retinal hemodynamic response to visual stimulation , 2013, Journal of biomedical optics.

[18]  Bruce J. Tromberg,et al.  A comparison of Doppler optical coherence tomography methods , 2012, Biomedical optics express.

[19]  Leopold Schmetterer,et al.  Doppler Optical Coherence Tomography , 2014, Progress in Retinal and Eye Research.

[20]  Cuixia Dai,et al.  Absolute retinal blood flow measurement with a dual-beam Doppler optical coherence tomography. , 2013, Investigative ophthalmology & visual science.

[21]  Ji Yi,et al.  A combined method to quantify the retinal metabolic rate of oxygen using photoacoustic ophthalmoscopy and optical coherence tomography , 2014, Scientific Reports.

[22]  Joseph A Izatt,et al.  In vivo total retinal blood flow measurement by Fourier domain Doppler optical coherence tomography. , 2007, Journal of biomedical optics.