Efficacy of robust optimization plan with partial‐arc VMAT for photon volumetric‐modulated arc therapy: A phantom study

Abstract This study investigated position dependence in planning target volume (PTV)‐based and robust optimization plans using full‐arc and partial‐arc volumetric modulated arc therapy (VMAT). The gantry angles at the periphery, intermediate, and center CTV positions were 181°–180° (full‐arc VMAT) and 181°–360° (partial‐arc VMAT). A PTV‐based optimization plan was defined by 5 mm margin expansion of the CTV to a PTV volume, on which the dose constraints were applied. The robust optimization plan consisted of a directly optimized dose to the CTV under a maximum‐uncertainties setup of 5 mm. The prescription dose was normalized to the CTV D99% (the minimum relative dose that covers 99% of the volume of the CTV) as an original plan. The isocenter was rigidly shifted at 1 mm intervals in the anterior‐posterior (A‐P), superior‐inferior (S‐I), and right‐left (R‐L) directions from the original position to the maximum‐uncertainties setup of 5 mm in the original plan, yielding recalculated dose distributions. It was found that for the intermediate and center positions, the uncertainties in the D99% doses to the CTV for all directions did not significantly differ when comparing the PTV‐based and robust optimization plans (P > 0.05). For the periphery position, uncertainties in the D99% doses to the CTV in the R‐L direction for the robust optimization plan were found to be lower than those in the PTV‐based optimization plan (P < 0.05). Our study demonstrated that a robust optimization plan's efficacy using partial‐arc VMAT depends on the periphery CTV position.

[1]  M. V. van Herk,et al.  The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[2]  K. Abbink,et al.  24 , 1871, You Can Cross the Massacre on Foot.

[3]  Karl Otto,et al.  Volumetric modulated arc therapy: IMRT in a single gantry arc. , 2007, Medical physics.

[4]  N. Kamikonya,et al.  Dosimetric and delivery characterizations of full-arc and half-arc volumetric-modulated arc therapy for maxillary cancer , 2012, Journal of radiation research.

[5]  Timothy C Y Chan,et al.  Accounting for range uncertainties in the optimization of intensity modulated proton therapy , 2007, Physics in medicine and biology.

[6]  A. Lomax,et al.  Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: the potential effects of calculational uncertainties , 2008, Physics in medicine and biology.

[7]  Yunfei Hu,et al.  Evaluation of RayStation robust optimisation for superficial target coverage with setup variation in breast IMRT , 2016, Australasian Physical & Engineering Sciences in Medicine.

[8]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[9]  Simon J Thomas,et al.  The effect of optimization on surface dose in intensity modulated radiotherapy (IMRT). , 2004, Physics in medicine and biology.

[10]  A. Lomax,et al.  Intensity modulation methods for proton radiotherapy. , 1999, Physics in medicine and biology.

[11]  Gudrun Goitein,et al.  The clinical potential of intensity modulated proton therapy. , 2004, Zeitschrift fur medizinische Physik.

[12]  Steven J Frank,et al.  PTV-based IMPT optimization incorporating planning risk volumes vs robust optimization. , 2013, Medical physics.

[13]  Thomas Bortfeld,et al.  Visualization of a variety of possible dosimetric outcomes in radiation therapy using dose-volume histogram bands. , 2012, Practical radiation oncology.

[14]  T. Eade,et al.  Volumetric-modulated arc therapy in head and neck radiotherapy: a planning comparison using simultaneous integrated boost for nasopharynx and oropharynx carcinoma. , 2011, Clinical oncology (Royal College of Radiologists (Great Britain)).

[15]  Harold O. Wyckoff,et al.  International Commission ON Radiation Units and Measurements (ICRU). , 1974, The American journal of roentgenology, radium therapy, and nuclear medicine.

[16]  Wei Chen,et al.  Including robustness in multi-criteria optimization for intensity-modulated proton therapy , 2011, Physics in medicine and biology.

[17]  A J Lomax,et al.  Intensity modulated proton therapy and its sensitivity to treatment uncertainties 2: the potential effects of inter-fraction and inter-field motions , 2008, Physics in medicine and biology.

[18]  Xiaoqin Jiang,et al.  Planning analysis for locally advanced lung cancer: dosimetric and efficiency comparisons between intensity-modulated radiotherapy (IMRT), single-arc/partial-arc volumetric modulated arc therapy (SA/PA-VMAT) , 2011, Radiation oncology.

[19]  John N Tsitsiklis,et al.  A robust approach to IMRT optimization , 2006, Physics in medicine and biology.

[20]  Anders Forsgren,et al.  Minimax optimization for handling range and setup uncertainties in proton therapy. , 2011, Medical physics.

[21]  Robert J. Shalek,et al.  Determination of Absorbed Dose in a Patient Irradiated by Beams of X or Gamma Rays in Radiotherapy Procedures , 1977 .