Results of Total Hip Arthroplasties in the Young Patient; Further Evidence for a Barrier Against Articular wear Debris by Hydroxyapatite Coatings

We examined the hypothesis that the circumferential osseous apposition around HA-coated implants forms a protective barrier against articular wear debris. Sixty-five hydroxyapatite-coated total hip arthroplasties in 57 patients (age <50years) with polyethylene-metal articulation were evaluated regarding PE-wear, osteolysis, and clinical outcome at a minimum of 10 years follow-up. There was no correlation between PE-wear and osteolysis of the femoral zones or cup zones I and III. A strong Pearson correlation was found between polyethylene wear and osteolysis around cup zone II, where the cup only consisted of polyethylene (p<0,01). The aseptic failure rate was 1.5% for the femoral component and 4.5% for the cup after 10 years of follow–up. The average Harris Hip Score was 90 and the average Engh score for fixation was 23 after 10 years. Around HA-coated parts of the prosthesis bone formation remained stable, regardless of the degree of polyethylene wear. The average linear polyethylene wear was 0,16 mm/year. The circumferential osseous apposition of the HA-coated implants possibly formed a protective barrier against articular wear debris. The use of cups with a backside gap resulted in PE-wear associated osteolysis in cup zone II and may be considered to be best avoided.

[1]  M. Archibeck,et al.  Cementless total hip arthroplasty in patients 50 years or younger. , 2006, The Journal of arthroplasty.

[2]  Shin-Yoon Kim,et al.  Early Failure of Hemispheric Hydroxyapatite-coated Acetabular Cups , 2006, Clinical orthopaedics and related research.

[3]  T. Wright,et al.  Mechanisms for pumping fluid through cementless acetabular components with holes. , 2005, The Journal of arthroplasty.

[4]  H. Rubash,et al.  The acetabular insert-metal backing interface: an additional source of polyethylene wear debris. , 2005, The Journal of arthroplasty.

[5]  M. Silva,et al.  In vivo wear of bilateral total hip replacements: conventional versus crosslinked polyethylene , 2005, Archives of Orthopaedic and Trauma Surgery.

[6]  K. Bendix,et al.  Superior sealing effect of hydroxyapatite in porous-coated implants , 2005, Acta orthopaedica.

[7]  P. Herberts,et al.  Highly Cross-linked Polyethylene in Total Hip Arthroplasty: Randomized Evaluation of Penetration Rate in Cemented and Uncemented Sockets Using Radiostereometric Analysis , 2004, Clinical orthopaedics and related research.

[8]  R.G.T. Geesink,et al.  Osteoconductive Coatings for Total Joint Arthroplasty , 2002, Clinical orthopaedics and related research.

[9]  M. Roffman,et al.  The effect of abduction differences between cup and contralateral acetabular angle on polyethylene component wear. , 2002, Orthopedics.

[10]  E. Garcia-Cimbrelo,et al.  The Harris-Galante total hip arthroplasty: a minimum 8-year follow-up study. , 2001, The Journal of arthroplasty.

[11]  L. Dorr,et al.  Total Hip Arthroplasty with Use of the Metasul Metal-on-Metal Articulation: Four to Seven-Year Results* , 2000, The Journal of bone and joint surgery. American volume.

[12]  H. Oonishi,et al.  Wear of high-dose gamma-irradiated polyethylene in total hip replacements , 2000, Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association.

[13]  Naoto Saito,et al.  Poor bone quality or hip structure as risk factors affecting survival of total-hip arthroplasty , 2000, The Lancet.

[14]  T. Bauer,et al.  Deformation of the acetabular polyethylene liner and the backside gap. , 1999, The Journal of arthroplasty.

[15]  H. Amstutz,et al.  The role of acetabular component screw holes and/or screws in the development of pelvic osteolysis , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[16]  J Fisher,et al.  Quantification of third-body damage and its effect on UHMWPE wear with different types of femoral head. , 1998, The Journal of bone and joint surgery. British volume.

[17]  C. Shih,et al.  Measurement of polyethylene wear in cementless total hip arthroplasty. , 1997, The Journal of bone and joint surgery. British volume.

[18]  W. Hozack,et al.  Hydroxyapatite-Coated Femoral Stems. A Matched-Pair Analysis of Coated and Uncoated Implants* , 1996, The Journal of bone and joint surgery. American volume.

[19]  V. Fornasier,et al.  Migration of polyethylene wear debris in hip arthroplasties: a canine model. , 1995, Journal of applied biomaterials : an official journal of the Society for Biomaterials.

[20]  R. Geesink,et al.  Six-year results of hydroxyapatite-coated total hip replacement. , 1995, The Journal of bone and joint surgery. British volume.

[21]  John A. Anderson,et al.  Hydroxyapatite‐Coated Hip Prostheses: Early Results From an International Study , 1995, Clinical orthopaedics and related research.

[22]  C. Rimnac,et al.  Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components. , 1994, The Journal of bone and joint surgery. British volume.

[23]  S. Medendorp,et al.  An indirect comparison of third-body wear in retrieved hydroxyapatite-coated, porous, and cemented femoral components. , 1994, Clinical orthopaedics and related research.

[24]  R. E. Jensen,et al.  Loss of hydroxyapatite coating on retrieved, total hip components. , 1993, The Journal of arthroplasty.

[25]  M. Røkkum,et al.  Femoral remodeling after arthroplasty of the hip. Prospective randomized 5-year comparison of 120 cemented/uncemented cases of arthrosis. , 1993, Acta orthopaedica Scandinavica.

[26]  R. Bloebaum,et al.  Osteolysis from a press-fit hydroxyapatite-coated implant. A case study. , 1993, The Journal of arthroplasty.

[27]  M. Ritter,et al.  Polyethylene wear in cemented metal-backed acetabular cups. , 1993, The Journal of bone and joint surgery. British volume.

[28]  C. Engh,et al.  Producing and avoiding stress shielding. Laboratory and clinical observations of noncemented total hip arthroplasty. , 1992, Clinical orthopaedics and related research.

[29]  T. Bauer,et al.  Hydroxyapatite-coated femoral stems. Histological analysis of components retrieved at autopsy. , 1991, The Journal of bone and joint surgery. American volume.

[30]  R G Geesink,et al.  Hydroxyapatite-coated total hip prostheses. Two-year clinical and roentgenographic results of 100 cases. , 1990, Clinical orthopaedics and related research.

[31]  C. Engh,et al.  Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. , 1990, Clinical orthopaedics and related research.

[32]  B. Morrey,et al.  Effect of femoral head size on wear of the polyethylene acetabular component. , 1990, The Journal of bone and joint surgery. American volume.

[33]  R. Geesink Hydroxyapatite-coated total hip protheses : two-year clinical roentenografic results of 100 cases , 1990 .

[34]  C. Klein,et al.  Bonding of bone to apatite-coated implants. , 1988, The Journal of bone and joint surgery. British volume.

[35]  C. Engh,et al.  THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL RESULTS , 1987 .

[36]  C. Engh,et al.  Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. , 1987, The Journal of bone and joint surgery. British volume.

[37]  P. Walker,et al.  Bony ingrowth fixation of the acetabular component in canine hip joint arthroplasty. , 1983, Clinical orthopaedics and related research.

[38]  H. Amstutz,et al.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. , 1979, Clinical orthopaedics and related research.

[39]  L. Riley,et al.  Ectopic ossification following total hip replacement. Incidence and a method of classification. , 1973, The Journal of bone and joint surgery. American volume.

[40]  M. B. Coventry,et al.  Factors influencing the results in 2,012 total hip arthroplasties. , 1973, Clinical orthopaedics and related research.