Scanning tunneling microscope calibration and reconstruction of real image: Drift and slope elimination

Drift, slope of sample, and indeterminate sensitivity of piezoceramics are considered as the origin of the linear scanning tunneling microscope (STM) image distortions. A special algorithm of STM scanning and reconstruction of image of unknown atomic structures is suggested and tested. This algorithm allows us to measure three components of the drift velocity and two angles, characterizing the average slope of scanning surface. On the one hand, using this algorithm, one can perform the STM calibration by a known surface structure (for example, the highly oriented pyrolytic graphite surface) even in the presence of a drift. This enables us to determine all the three piezoceramics constants of STM piezoscanner and the deviation of the real scanner axes X and Y from orthogonality. On the other hand, using such a calibrated STM and the described algorithm, it is possible to obtain the real STM image and make measurements for unknown surfaces with atomic resolution without the distortions mentioned above. As i...

[1]  A. M. Simpson,et al.  Thermal expansion and piezoelectric response of PZT Channel 5800 for use in low‐temperature scanning tunneling microscope designs , 1987 .

[2]  E. Stoll,et al.  Scanning tunneling microscopy of individual molecules of copper phthalocyanine adsorbed on polycrystalline silver surfaces , 1987 .

[3]  J. G. Adler,et al.  Comparison of various low‐temperature scanning tunneling microscopy designs for use in spectroscopy and topography , 1991 .

[4]  J. Clarke,et al.  Scanning tunneling microscopy of silver, gold, and aluminum monomers and small clusters on graphite , 1988 .

[5]  P. Wyder,et al.  High‐stability scanning tunneling microscope , 1985 .

[6]  Toshiro Higuchi,et al.  Crystalline lattice for metrological applications and positioning control by a dual tunneling‐unit scanning tunneling microscope , 1991 .

[7]  E. P. Stoll Restoration of STM images distorted by time-dependent piezo driver aftereffects , 1992 .

[8]  H. van Kempen,et al.  Calibration and characterization of piezoelectric elements as used in scanning tunneling microscopy , 1991 .

[9]  Dieter W. Pohl,et al.  Some design criteria in scanning tunneling microscopy , 1986, IBM J. Res. Dev..

[10]  M. Taborelli,et al.  Deformations and nonlinearity in scanning tunneling microscope images , 1991 .

[11]  J. White,et al.  Diffraction grating calibration of scanning tunneling microscope piezoscanners , 1990 .

[12]  T. Hasegawa,et al.  Initial stage of Au adsorption onto a Si(111) surface studied by scanning tunneling microscopy , 1991 .