Validation of weighted frequency-difference EIT using a three-dimensional hemisphere model and phantom.

Frequency-difference (FD) electrical impedance tomography (EIT) using a weighted voltage difference has recently been proposed for imaging haemorrhagic stroke, abdominal bleeding and tumors. Although its feasibility was demonstrated through two-dimensional numerical simulations and phantom experiments, we should validate the method in three-dimensional imaging objects. At the same time, we need to investigate its robustness against geometrical modeling errors in boundary shapes and electrode positions. We performed a validation study of the weighted FD method through three-dimensional numerical simulations and phantom experiments. Adopting hemispherical models and phantoms whose admittivity distributions change with frequency, we investigated the performance of the method to detect an anomaly. We found that the simple FD method fails to detect the anomaly, whereas reconstructed images using the weighted FD method clearly visualize the anomaly. The weighted FD method is robust against modeling errors of boundary-shape deformations and displaced electrode positions. We also found that the method is capable of detecting an anomaly surrounded by a shell-shaped obstacle simulating the skull. We propose the weighted FD method for future studies of animal and human experiments.

[1]  Thom F. Oostendorp,et al.  The conductivity of the human skull: results of in vivo and in vitro measurements , 2000, IEEE Transactions on Biomedical Engineering.

[2]  R H Bayford,et al.  Multi-frequency electrical impedance tomography (EIT) of the adult human head: initial findings in brain tumours, arteriovenous malformations and chronic stroke, development of an analysis method and calibration , 2006, Physiological measurement.

[3]  D. C. Barber,et al.  Three-dimensional electrical impedance tomography , 1996, Nature.

[4]  R. W. Lau,et al.  The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. , 1996, Physics in medicine and biology.

[5]  H Griffiths,et al.  A dual-frequency electrical impedance tomography system. , 1989, Physics in medicine and biology.

[6]  R H Bayford,et al.  Electrical impedance tomography spectroscopy (EITS) for human head imaging. , 2003, Physiological measurement.

[7]  H Griffiths,et al.  A dual-frequency applied potential tomography technique: computer simulations. , 1987, Clinical physics and physiological measurement : an official journal of the Hospital Physicists' Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics.

[8]  D. Cohen,et al.  Demonstration of useful differences between magnetoencephalogram and electroencephalogram. , 1983, Electroencephalography and clinical neurophysiology.

[9]  Eung Je Woo,et al.  Calibration methods for a multi-channel multi-frequency EIT system , 2007, Physiological measurement.

[10]  Eung Je Woo,et al.  Frequency-difference electrical impedance tomography (fdEIT): algorithm development and feasibility study , 2008, Physiological measurement.

[11]  Eung Je Woo,et al.  Validation of a multi-frequency electrical impedance tomography (mfEIT) system KHU Mark1: impedance spectroscopy and time-difference imaging , 2008, Physiological measurement.

[12]  Eung Je Woo,et al.  Frequency-difference EIT (fdEIT) using weighted difference and equivalent homogeneous admittivity: validation by simulation and tank experiment , 2009, Physiological measurement.

[13]  R Bayford,et al.  Design and calibration of a compact multi-frequency EIT system for acute stroke imaging. , 2006, Physiological measurement.

[14]  L. Geddes,et al.  The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologist , 1967, Medical and biological engineering.

[15]  D S Holder,et al.  An electrode addressing protocol for imaging brain function with electrical impedance tomography using a 16-channel semi-parallel system , 2009, Physiological measurement.

[16]  Eung Je Woo,et al.  Multi-frequency EIT system with radially symmetric architecture: KHU Mark1 , 2007, Physiological measurement.

[17]  D. A. Driscoll,et al.  Current Distribution in the Brain From Surface Electrodes , 1968, Anesthesia and analgesia.

[18]  Eung Je Woo,et al.  Multi-frequency time-difference complex conductivity imaging of canine and human lungs using the KHU Mark1 EIT system , 2009, Physiological measurement.

[19]  David Atkinson,et al.  Use of anisotropic modelling in electrical impedance tomography; Description of method and preliminary assessment of utility in imaging brain function in the adult human head , 2008, NeuroImage.

[20]  David S. Holder,et al.  Electrical Impedance Tomography : Methods, History and Applications , 2004 .