The sensitivity of 100 nm RADFETs with zero gate bias up to dose of 230 Gy(Si)

Abstract The RADFETs (pMOS dosimeters) were irradiated by ionizing radiation from 60 Co gamma-ray source to dose of 230 Gy(Si) without gate bias, and their reproducibility and sensitivity to radiation were investigated. The completely automatic system containing switching matrix have been used, enabling very reliable measurements of RADFET electrical characteristics. Two sample types from each RADFET chip, with the same gate oxide thickness, but different geometry (channel width and length), were investigated. The samples have shown good reproducibility of the threshold voltage shift during irradiation, i.e., the radiation sensitivity was similar for all samples. Using MG and CP techniques, it has been shown that the slow switching (border) trap density is negligible, but the fixed trap density saturates and the fast switching trap density is linear.

[1]  A. Holmes-Siedle,et al.  Dosimetric Silica Films: The Influence of Fields on the Capture of Positive Charge , 1982, IEEE Transactions on Nuclear Science.

[2]  M. Pejovic,et al.  pMOS dosimeter with two-layer gate oxide operated at zero and negative bias , 1994 .

[3]  Philippe Godignon,et al.  The electron irradiation effects on silicon gate dioxide used for power MOS devices , 2001, Microelectron. Reliab..

[4]  A. Faigón,et al.  Electrically Erasable Metal–Oxide–Semiconductor Dosimeters , 2007, IEEE Transactions on Nuclear Science.

[5]  A. Holmes-Siedle Predicting end-of-life performance of microelectronics in space , 1994 .

[6]  F. B. McLean A Framework for Understanding Radiation-Induced Interface States in SiO2 MOS Structures , 1980, IEEE Transactions on Nuclear Science.

[7]  M. Glaser,et al.  Conception of an integrated sensor for the radiation monitoring of the CMS experiment at the large hadron collider , 2004, IEEE Transactions on Nuclear Science.

[8]  Muhammad Arsalan,et al.  Wireless Dosimeter: System-on-Chip Versus System-in-Package for Biomedical and Space Applications , 2008, IEEE Transactions on Circuits and Systems II: Express Briefs.

[9]  Milic M. Pejovic,et al.  Gamma-ray irradiation and post-irradiation responses of high dose range RADFETs , 2001, RADECS 2001. 2001 6th European Conference on Radiation and Its Effects on Components and Systems (Cat. No.01TH8605).

[10]  A. Albadri,et al.  The empirical dependence of radiation-induced charge neutralization on negative bias in dosimeters based on the metal-oxide-semiconductor field-effect transistor , 2006 .

[11]  A. Holmes-Siedle,et al.  Linearity of pMOS radiation dosimeters operated at zero bias , 1985 .

[12]  Y. Asano,et al.  Radiation monitoring in Mrad range using radiation-sensing field-effect transistors , 2005 .

[13]  E. Vlaikova,et al.  Investigation of p-type MOS structure irradiated with 23 MeV electrons , 2006 .

[14]  A. Holmes-Siedle,et al.  Thick oxide pMOSFET dosimeters for high energy radiation , 1988 .

[15]  M. Pejovic,et al.  pMOS dosimetric transistors with two-layer gate oxide , 1997 .

[16]  P. Winokur,et al.  Simple technique for separating the effects of interface traps and trapped‐oxide charge in metal‐oxide‐semiconductor transistors , 1986 .

[17]  Goran S. Ristić,et al.  Influence of ionizing radiation and hot carrier injection on metal-oxide-semiconductor transistors , 2008 .

[18]  A. Holmes-Siedle,et al.  Mapping CMOS radiation tolerance data on a 4-lane chart , 1994 .

[19]  G. Groeseneken,et al.  A reliable approach to charge-pumping measurements in MOS transistors , 1984, IEEE Transactions on Electron Devices.

[20]  M. Pejovic,et al.  Physico-chemical processes in metal–oxide–semiconductor transistors with thick gate oxide during high electric field stress , 2007 .

[21]  L. Adams,et al.  RADFET: A review of the use of metal-oxide-silicon devices as integrating dosimeters , 1986 .

[22]  Energy and integrated dose dependence of MOSFET dosimeter sensitivity for irradiation energies between 30kV and Co60. , 2006, Medical physics.

[23]  Evaluation of commercial programmable floating gate devices as radiation dosimeters , 2009 .

[24]  Aleksandar B. Jakšić,et al.  Comparison between post-irradiation annealing and post-high electric field stress annealing of n-channel power VDMOSFETs , 2003 .

[25]  L. Adams,et al.  MOS dosimeters-improvement of responsivity , 1991, RADECS 91 First European Conference on Radiation and its Effects on Devices and Systems.

[26]  A. Elliot,et al.  The use of charge pumping currents to measure surface state densities in MOS transistors , 1976 .

[27]  Andrew Holmes-Siedle,et al.  A Simple Model for Predicting Radiation Effects in MOS Devices , 1978, IEEE Transactions on Nuclear Science.

[28]  M. Heyns,et al.  X-ray irradiation effect on the reliability of ultrathin gate oxides and oxynitrides , 1999 .

[29]  Alberto J. Palma,et al.  Readout techniques for linearity and resolution improvements in MOSFET dosimeters , 2010 .

[30]  L. August Design Criteria for a High-Dose MOS Dosimeter for Use in Space , 1984, IEEE Transactions on Nuclear Science.