Simulation study of PET detector limitations using continuous crystals

Continuous crystals can potentially obtain better intrinsic detector spatial resolution compared to pixelated crystals, additionally providing depth of interaction (DoI) information from the light distribution. To achieve high performance sophisticated interaction position estimation algorithms are required. There are a number of algorithms in the literature applied to different crystal dimensions and different photodetectors. However, the different crystal properties and photodetector array geometries have an impact on the algorithm performance. In this work we analysed, through Monte Carlo simulations, different combinations of realistic crystals and photodetector parameters to better understand their influence on the interaction position estimation accuracy, with special emphasis on the DoI. We used an interaction position estimation based on an analytical model for the present work. Different photodetector granulation schemes were investigated. The impact of the number of crystal faces readout by photodetectors was studied by simulating scenarios with one and two photodetectors. In addition, crystals with different levels of reflection and aspect ratios (AR) were analysed. Results showed that the impact of photodetector granularity is mainly shown near the edges and specially in the corners of the crystal. The resulting intrinsic spatial resolution near the centre with a 12 × 12 × 10 mm(3) LYSO crystal was 0.7-0.9 mm, while the average spatial resolution calculated on the entire crystal was 0.77 ± 0.18 mm for all the simulated geometries with one and two photodetectors. Having front and back photodetectors reduced the DoI bias (Euclidean distance between estimated DoI and real DoI) and improved the transversal resolution near the corners. In scenarios with one photodetector, small AR resulted in DoI inaccuracies for absorbed events at the entrance of the crystal. These inaccuracies were slightly reduced either by increasing the AR or reducing the amount of reflected light, and highly mitigated using two photodetectors. Using one photodetector, we obtained a piecewise DoI error model with a DoI resolution of 0.4-0.9 mm for a 1.2 AR crystal, and we observed that including a second photodetector or reducing the amount of reflections reduced the DoI bias but did not significantly improve the DoI resolution. Translating the piecewise DoI error model obtained in this study to image reconstruction we obtained a spatial resolution variability of 0.39 mm using 85% of the FoV, compared to 2.59 mm and 1.87 mm without DoI correction or with a dual layer system, respectively.

[1]  K. Kamada,et al.  Scintillation properties of Ce doped Gd2Lu1(Ga,Al)5O12 single crystal grown by the micro-pulling-down method , 2012 .

[2]  G. Vandersteen,et al.  Nonlinear least-squares modeling of 3D interaction position in a monolithic scintillator block , 2010, Physics in medicine and biology.

[3]  Simon R Cherry,et al.  Experimental characterization and system simulations of depth of interaction PET detectors using 0.5 mm and 0.7 mm LSO arrays , 2009, Physics in medicine and biology.

[4]  Sibylle Ziegler,et al.  Positron Emission Tomography: Principles, Technology, and Recent Developments , 2005 .

[5]  I. Buvat,et al.  A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology , 2012, Physics in medicine and biology.

[6]  R S Miyaoka,et al.  Depth of interaction decoding of a continuous crystal detector module , 2007, Physics in medicine and biology.

[7]  M. G. Bisogni,et al.  Characterization of a PET detector head based on continuous LYSO crystals and monolithic, 64-pixel silicon photomultiplier matrices , 2010, Physics in medicine and biology.

[8]  J. Benlloch,et al.  Dependency of Energy-, Position- and Depth of Interaction Resolution on Scintillation Crystal Coating and Geometry , 2008, IEEE Transactions on Nuclear Science.

[9]  Sascha Moehrs,et al.  A detector head design for small-animal PET with silicon photomultipliers (SiPM) , 2006, Physics in medicine and biology.

[10]  A. Sebastia,et al.  Depth of /spl gamma/-ray interaction within continuous crystals from the width of its scintillation light-distribution , 2005, IEEE Transactions on Nuclear Science.

[11]  Harrison H. Barrett,et al.  Multiple-Hit Parameter Estimation in Monolithic Detectors , 2013, IEEE Transactions on Medical Imaging.

[12]  L. Shepp,et al.  Maximum Likelihood Reconstruction for Emission Tomography , 1983, IEEE Transactions on Medical Imaging.

[13]  F. Beekman,et al.  Optical simulation of monolithic scintillator detectors using GATE/GEANT4 , 2010, Physics in medicine and biology.

[14]  R S Miyaoka,et al.  Study of PET Detector Performance With Varying SiPM Parameters and Readout Schemes , 2011, IEEE Transactions on Nuclear Science.

[15]  Piergiorgio Cerello,et al.  Development of a PET detector module with Depth of Interaction capability , 2013 .

[16]  S. Tavernier,et al.  Neural network-based position estimators for PET detectors using monolithic LSO blocks , 2004, IEEE Transactions on Nuclear Science.

[17]  Magdalena Rafecas,et al.  High resolution detectors based on continuous crystals and SiPMs for small animal PET , 2013 .

[18]  H. Barrett,et al.  Maximum-Likelihood Methods for Processing Signals From Gamma-Ray Detectors , 2009, IEEE Transactions on Nuclear Science.

[19]  Herbert Löhner,et al.  A practical method for depth of interaction determination in monolithic scintillator PET detectors , 2011, Physics in medicine and biology.

[20]  R. Siddon Fast calculation of the exact radiological path for a three-dimensional CT array. , 1985, Medical physics.

[21]  M. Maas,et al.  Experimental characterization of monolithic-crystal small animal PET detectors read out by APD arrays , 2006, IEEE Transactions on Nuclear Science.

[22]  Jurgen Seidel,et al.  Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner. , 2006, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[23]  Simulation studies of optical photons in monolithic block scintillators , 2011, 2011 IEEE Nuclear Science Symposium Conference Record.

[24]  Magnus Dahlbom,et al.  Parallax correction in PET using depth of interaction information , 1997, 1997 IEEE Nuclear Science Symposium Conference Record.

[25]  M. G. Bisogni,et al.  High performance detector head for PET and PET/MR with continuous crystals and SiPMs , 2013 .

[26]  Kanai Shah,et al.  Continuous depth-of-interaction encoding using phosphor-coated scintillators , 2009, Physics in medicine and biology.

[27]  Takahiro Matsumoto,et al.  A SiPM-based isotropic-3D PET detector X'tal cube with a three-dimensional array of 1 mm3 crystals , 2011, Physics in medicine and biology.

[28]  Thomas F. Coleman,et al.  An Interior Trust Region Approach for Nonlinear Minimization Subject to Bounds , 1993, SIAM J. Optim..

[29]  H. Barrett,et al.  Calibration Method for ML Estimation of 3D Interaction Position in a Thick Gamma-Ray Detector , 2009, IEEE Transactions on Nuclear Science.

[30]  M. Rafecas,et al.  Characterization and readout of MADPET-II detector modules: validation of a unique design concept for high resolution small animal PET , 2005, IEEE Transactions on Nuclear Science.

[31]  R. Cuerdo,et al.  Study and optimization of positioning algorithms for monolithic PET detectors blocks , 2012 .