The effects of formulation parameters on the product performance characteristics of solution metered dose inhalers (MDIs) were determined using ethanol as the cosolvent and HFA 134a as the propellant. Solubility of beclomethasone dipropionate (BDP) was determined in various blends of 134a and ethanol and was shown to increase with ethanol concentration. Product performance was assessed using the APS Model 3306 Impactor Inlet in conjunction with APS Model 3320 Aerodynamic Particle Sizer (APS). Nine solution formulations containing various BDP and ethanol concentrations were studied. Chemical analysis of the Impactor Inlet was performed in order to determine the "respirable" deposition of the MDI system. With increased ethanol concentration, the throat deposition and plate deposition increased and the respirable deposition decreased. The mass median aerodynamic diameter (MMAD) increased with the increasing drug concentration, but did not show a significant increase with an increase in ethanol concentration. This indicates that the efficiency of solution MDIs decreases with increased ethanol concentration. A Maximum Respirable Mass (MRM) was calculated based on the drug solubility at a particular ethanol concentration and the respirable deposition for a 50mcl valve and QVAR actuator for that ethanol concentration. The MRM represents the maximum amount of a given drug that can be delivered to the lungs theoretically and is very sensitive to the solubility profile of the drug. The MRM increased with the increasing ethanol concentration in the formulation until a plateau was reached at an ethanol concentration of 10-15% w/w. The MRM initially increases with increase in ethanol concentration due to the increase in drug solubility. However, at higher ethanol concentrations the increase in drug solubility was negated by a decrease in the respirable deposition. This study illustrates the importance of considering both formulation properties and product performance characteristics when optimizing a metered dose inhaler drug delivery system.
[1]
P. Myrdal,et al.
A theoretical and experimental analysis of formulation and device parameters affecting solution MDI size distributions.
,
2004,
Journal of pharmaceutical sciences.
[2]
Abhishek Gupta,et al.
On-line high-performance liquid chromatography method for analyte quantitation from pressurized metered dose inhalers.
,
2004,
Journal of chromatography. A.
[3]
S. W. Stein,et al.
Optimized dose delivery of the peptide cyclosporine using hydrofluoroalkane-based metered dose inhalers.
,
2004,
Journal of pharmaceutical sciences.
[4]
P. Hairston,et al.
An Evaluation of Mass-Weighted Size Distribution Measurements with the Model 3320 Aerodynamic Particle Sizer
,
2002
.
[5]
M. Igawa,et al.
High frequency and large deposition of acid fog on high elevation forest.
,
2002,
Environmental science & technology.
[6]
N. C. Miller,et al.
Test method for evaluation of loss of prime in metered-dose aerosols.
,
1988,
Journal of pharmaceutical sciences.
[7]
M. Newhouse,et al.
Site of beta-adrenergic receptors in the respiratory tract: use of fenoterol administered by two methods.
,
1978,
Chest.
[8]
W. Kirk.
In vitro method of comparing clouds produced from inhalation aerosols for efficiency in penetration of airways.
,
1972,
Journal of pharmaceutical sciences.