The noncontact tonometer. Clinical evaluation on normal and diseased eyes.

The measurement of intraocular pressure with a noncontact tonometer was correlated to a Goldmann tonometer in 20 normal persons, 20 glaucomatous patients, and 8 patients suffering from corneal diseases. In normal persons the error in measurement for the noncontact tonometer was related to their skill in fixation, and in eye patients to height of pressure and corneal state. Acceptable correlation was found between noncontact tonometry and Goldmann applanation tonometry when the cornea was normal and the pressure below 35 mmHg (Goldmann), otherwise noncontact tonometry was only a guide, and in the presence of corneal disease, unreliable. Good fixation reduced the method error. The standard deviation was 1.09 mmHg at poor fixation and 0.60 mmHg at good fixation. Repeated measurements on the same eye with noncontact tonometry did not alter the intraocular tension.

[1]  J. Stepanik Das AO-non contact-Tonometer , 1974, Albrecht von Graefes Archiv für klinische und experimentelle Ophthalmologie.

[2]  N. Ducrey,et al.  Le tonomètre à air , 1975 .

[3]  B. Grolman,et al.  A noncontact applanation tonometer. Description and clinical evaluation. , 1974, Archives of ophthalmology.

[4]  B. Grolman,et al.  A noncontact applanation tonometer. , 1973, The Sight-saving review.

[5]  B Grolman,et al.  A NEW TONOMETER SYSTEM* , 1972, American journal of optometry and archives of American Academy of Optometry.

[6]  C. Krakau,et al.  ON REPEATED TONOMETRY , 1971, Acta ophthalmologica.

[7]  C. Krakau A Vibration Tonometer , 1970 .

[8]  R. Moses Repeated applanation tonometry. , 1961, Ophthalmologica. Journal international d'ophtalmologie. International journal of ophthalmology. Zeitschrift fur Augenheilkunde.