暂无分享,去创建一个
In Dual Head Coincidence Imaging (DHCI), rotational weights must be applied to detected coincidence events to compensate for varying dwell-times. The dwell-time depends on both the transaxial and axial distance coordinates describing a coincidence line in parallel projection space. An exact formula is derived for the dwell-time of a DHCI camera with continuous detector rotation given a coincidence line described in parallel projection coordinates. The formula is verified by simulation. In addition, the boundary of the region where no data are measured is described. This missing-data region differs from that of a cylindrical PVI system whose diameter is the same as the head separation of the DHCI system and whose axial extent is the same as the axial dimension of the rotating planar detectors in the DHCI system.
[1] D. Townsend,et al. Increased sensitivity and field of view for a rotating positron camera. , 1984, Physics in Medicine and Biology.
[2] K. Erlandsson,et al. Fast accurate iterative reconstruction for low-statistics positron volume imaging. , 1998, Physics in medicine and biology.