Round Window Membrane Implantation with an Active Middle Ear Implant: A Study of the Effects on the Performance of Round Window Exposure and Transducer Tip Diameter in Human Cadaveric Temporal Bones

Objectives: To assess the importance of 2 variables, transducer tip diameter and resection of the round window (RW) niche, affecting the optimization of the mechanical stimulation of the RW membrane with an active middle ear implant (AMEI). Materials and Methods: Ten temporal bones were prepared with combined atticotomy and facial recess approach to expose the RW. An AMEI stimulated the RW with 2 ball tip diameters (0.5 and 1.0 mm) before and after the resection of the bony rim of the RW niche. The RW drive performance, assessed by stapes velocities using laser Doppler velocimetry, was analyzed in 3 frequency ranges: low (0.25–1 kHz), medium (1–3 kHz) and high (3–8 kHz). Results: Driving the RW produced mean peak stapes velocities (HEV) of 0.305 and 0.255 mm/s/V at 3.03 kHz, respectively, for the 1- and 0.5-mm tips, with the RW niche intact. Niche drilling increased the HEV to 0.73 and 0.832 mm/s/V for the 1- and 0.5-mm tips, respectively. The tip diameter produced no difference in output at low and medium frequencies; however, the 0.5-mm tip was 5 and 6 dB better than the 1-mm tip at high frequencies before and after niche drilling, respectively. Drilling the niche significantly improved the output by 4 dB at high frequencies for the 1-mm tip, and by 6 and 10 dB in the medium- and high-frequency ranges for the 0.5-mm tip. Conclusion: The AMEI was able to successfully drive the RW membrane in cadaveric temporal bones using a classical facial recess approach. Stimulation of the RW membrane with an AMEI without drilling the niche is sufficient for successful hearing outputs. However, the resection of the bony rim of the RW niche significantly improved the RW stimulation at medium and higher frequencies. Drilling the niche enhances the exposure of the RW membrane and facilitates positioning the implant tip.

[1]  Matthias Bornitz,et al.  TORP-Vibroplasty: A New Alternative for the Chronically Disabled Middle Ear , 2008, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[2]  Matthias Bornitz,et al.  Noise Exposure of the Inner Ear During Drilling a Cochleostomy for Cochlear Implantation , 2007, The Laryngoscope.

[3]  J. Nadol Techniques for human temporal bone removal: Information for the scientific community , 1996, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.

[4]  J M Aran,et al.  Piezoelectric middle ear implant preserving the ossicular chain. , 1995, Otolaryngologic clinics of North America.

[5]  Christian Streitberger,et al.  Coupling the Vibrant Soundbridge to Cochlea Round Window: Auditory Results in Patients With Mixed Hearing Loss , 2009, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[6]  R A Ruth,et al.  The round window electromagnetic implantable hearing aid approach. , 1995, Otolaryngologic clinics of North America.

[7]  John K. Niparko,et al.  Otologics middle ear transducer™ ossicular stimulator: Performance results with varying degrees of sensorineural hearing loss , 2004, Acta oto-laryngologica.

[8]  F. Linthicum,et al.  Morphometry of the human cochlear wall and implications for cochlear surgery* , 2004, Acta oto-laryngologica.

[9]  A. Huber,et al.  The incudo-malleolar joint and sound transmission losses , 2002, Hearing Research.

[10]  Simonetta Monini,et al.  Totally implantable middle ear device for rehabilitation of sensorineural hearing loss: preliminary experience with the Esteem®, Envoy , 2009, Acta oto-laryngologica.

[11]  John J. Rosowski,et al.  Measurements of stapes velocity in live human ears , 2009, Hearing Research.

[12]  J. Magnan,et al.  New Perspectives for Middle Ear Implants: First Results in Otosclerosis With Mixed Hearing Loss , 2007, The Laryngoscope.

[13]  S. Tringali,et al.  Fully implantable hearing device with transducer on the round window as a treatment of mixed hearing loss. , 2009, Auris, nasus, larynx.

[14]  E. Wever,et al.  LII The Functions of the round Window , 1948, The Annals of otology, rhinology, and laryngology.

[15]  K. Koka,et al.  Electrocochleographic and mechanical assessment of round window stimulation with an active middle ear prosthesis , 2010, Hearing Research.

[16]  M. Kringlebotn,et al.  The equality of volume displacements in the inner ear windows. , 1995, The Journal of the Acoustical Society of America.

[17]  E. Truy,et al.  European Results With Totally Implantable Carina Placed on the Round Window: 2-Year Follow-Up , 2009, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[18]  M. Hoffer,et al.  Otologics Fully Implantable Hearing System: Phase I Trial 1-Year Results , 2008, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[19]  Sigfrid D. Soli,et al.  Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window , 2006, International journal of audiology.

[20]  John J. Rosowski,et al.  The Effect of Methodological Differences in the Measurement of Stapes Motion in Live and Cadaver Ears , 2006, Audiology and Neurotology.

[21]  S. Mattheis,et al.  Fully Implantable Hearing Aids in Patients With Congenital Auricular Atresia , 2007, The Laryngoscope.

[22]  Jonathan H. Spindel,et al.  The Basis For A Round Window Electromagnetic Implantable Hearing Aid , 1991, Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society Volume 13: 1991.

[23]  Stefan Stenfelt,et al.  Round window membrane motion with air conduction and bone conduction stimulation , 2004, Hearing Research.

[24]  E. Truy,et al.  Fully implantable hearing device as a new treatment of conductive hearing loss in Franceschetti syndrome. , 2008, International journal of pediatric otorhinolaryngology.

[25]  Stefan Stenfelt,et al.  Fluid volume displacement at the oval and round windows with air and bone conduction stimulation. , 2004, The Journal of the Acoustical Society of America.

[26]  Sunil Puria,et al.  Measurements of human middle ear forward and reverse acoustics: implications for otoacoustic emissions. , 2003, The Journal of the Acoustical Society of America.

[27]  S. Merchant,et al.  Testing a Method for Quantifying the Output of Implantable Middle Ear Hearing Devices , 2007, Audiology and Neurotology.

[28]  Eric Truy,et al.  A Middle Ear Implant, the Symphonix Vibrant Soundbridge: Retrospective Study of the First 125 Patients Implanted in France , 2003, Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology.

[29]  C. G. Wright,et al.  Cochlear Implant Electrode Insertion: The Round Window Revisited , 2007, The Laryngoscope.

[30]  S. Tringali,et al.  A Pilot Study of the Safety and Performance of the Otologics Fully Implantable Hearing Device: Transducing Sounds via the Round Window Membrane to the Inner Ear , 2008, Audiology and Neurotology.

[31]  J. Aran,et al.  Piezo-electric middle ear implant hearing aid experimental model in guinea-pig , 1992, Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[32]  J. Fredrickson,et al.  The Otologics MET ossicular stimulator. , 2001, Otolaryngologic clinics of North America.

[33]  Herman A. Jenkins,et al.  U.S. Phase I Preliminary Results of Use of the Otologics MET Fully-Implantable Ossicular Stimulator , 2007, Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery.