Efficacy of a Miniature Centrifugal Rotary Pump (TinyPump) for Transfusion-Free Cardiopulmonary Bypass in Neonatal Piglets

We have developed a miniaturized semiclosed cardiopulmonary bypass (CPB) circuit incorporating a centrifugal blood pump (TinyPump) with a volume of 5 ml. The current study was undertaken to evaluate the hemolytic performance of the TinyPump in comparison with the BioPump and to investigate the impact of different CPB circuit volumes on hemodilution, coagulation, and the inflammatory response. Twelve 1-week-old piglets (3.4 ± 0.2 kg) were used. The circuit comprised a centrifugal pump, a membrane oxygenator, and a cardiotomy reservoir. Cardiopulmonary bypass was conducted with mild hypothermia at 150 ml/kg/min for 3 hours. Transfusion was not performed. Priming volume was 68 ml for the circuit with the TinyPump and 111 ml for the circuit with the BioPump. Although the TinyPump required higher speed, plasma free hemoglobin levels after CPB were not different between the groups. After CPB, the TinyPump group had a significantly higher hematocrit (27% ± 3% vs. 23% ± 3%) and lower platelet reduction rate, lower thrombin-antithrombin complex levels, and lower interleukin-6 levels. Better lung compliance with less water content was observed in the TinyPump group. The TinyPump maintained CPB with acceptable hemolysis and lower inflammatory responses. This miniaturized CPB circuit may make transfusion-free open heart surgery feasible in neonates and would help to prevent postoperative organ dysfunction.

[1]  K. Watterson,et al.  Effects of cardiopulmonary bypass on neonatal and paediatric inflammatory profiles. , 1997, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[2]  K. Ishihara,et al.  In vitro and ex vivo blood compatibility study of 2-methacryloyloxyethyl phosphorylcholine (MPC) copolymer-coated hemodialysis hollow fibers , 2003, Journal of Artificial Organs.

[3]  T. Hall,et al.  The pathophysiology of cardiopulmonary bypass. The risks and benefits of hemodilution. , 1995, Chest.

[4]  Michael J Cromer,et al.  A minimal priming technique that allows for a higher circulating hemoglobin on cardiopulmonary bypass , 1998, Perfusion.

[5]  B. Spiess,et al.  Endothelial cell injury in cardiovascular surgery: the procoagulant response. , 1996, The Annals of thoracic surgery.

[6]  Setsuo Takatani,et al.  Feasibility of a Miniature Centrifugal Rotary Blood Pump for Low-Flow Circulation in Children and Infants , 2005, ASAIO journal.

[7]  R. Hopkins,et al.  Interleukin-6 levels in serum and lung lavage fluid of children undergoing open heart surgery correlate with postoperative morbidity , 1998, Intensive Care Medicine.

[8]  O. Ponzio,et al.  Comparison of the inflammatory response between miniaturized and standard CPB circuits in aortic valve surgery. , 2006, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[9]  E. Moore,et al.  Interleukin-6 in the injured patient. Marker of injury or mediator of inflammation? , 1996, Annals of surgery.

[10]  E. Darling,et al.  Mini-circuit cardiopulmonary bypass with vacuum assisted venous drainage: feasibility of an asanguineous prime in the neonate , 1999, Perfusion.

[11]  R. Hetzer,et al.  Transfusion-free cardiopulmonary bypass in Jehovah's Witness patients weighing less than 5 kg. , 2005, The journal of extra-corporeal technology.

[12]  S. Clark Lung Injury after Cardiopulmonary Bypass , 2006, Perfusion.

[13]  Frank Merkle,et al.  Perfusion technique for nonhaemic cardiopulmonary bypass prime in neonates and infants under 6 kg body weight , 2004, Perfusion.

[14]  R. Ungerleider,et al.  Optimizing response of the neonate and infant to cardiopulmonary bypass. , 2003, Seminars in thoracic and cardiovascular surgery. Pediatric cardiac surgery annual.

[15]  E. Verrier,et al.  Endothelial cell injury in cardiovascular surgery: ischemia-reperfusion. , 1996, The Annals of thoracic surgery.

[16]  J. Hazelzet,et al.  Relation between cytokines and routine laboratory data in children with septic shock and purpura , 1994, Intensive Care Medicine.

[17]  R. Groom,et al.  Repair of hypoplastic left heart syndrome of a 4.25-kg Jehovah’s witness , 2002, Perfusion.

[18]  K. Ishihara,et al.  Feasibility of a tiny centrifugal blood pump (TinyPump) for pediatric extracorporeal circulatory support. , 2007, Artificial organs.