PURPOSE
To compare periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) multishot fast spin-echo diffusion-weighted magnetic resonance (MR) imaging with single-shot echo-planar diffusion-weighted MR imaging for image quality and visualization of acute cerebral infarction.
MATERIALS AND METHODS
Seventy subjects (35 men, 35 women; mean age, 55 years +/- 24 [SD]) who were suspected of having acute cerebral infarction (symptom duration, 2.8 days +/- 2.7) underwent PROPELLER and echo-planar MR imaging (b = 1,000 sec/mm(2)). Two neuroradiologists compared unlabeled images for presence of artifacts, visualization of infarction, and their preference of images. Interobserver agreement and image comparison were assessed by using the kappa statistic and the chi(2) test, respectively.
RESULTS
PROPELLER MR imaging reduced susceptibility artifacts (n = 70 subjects), which limited visualization of temporal (echo-planar, n = 64; PROPELLER, n = 0; P <.01, chi(2) test), frontal (echo-planar, n = 58; PROPELLER, n = 1; P <.01), and parietal lobes (echo-planar, n = 5; PROPELLER, n = 0; P <.05) and cerebellum (echo-planar, n = 36; PROPELLER, n = 0; P <.01) and brainstem (echo-planar, n = 23; PROPELLER, n = 0; P <.01). Acute infarction (n = 31 subjects) was better demonstrated at PROPELLER MR imaging (PROPELLER better, n = 18; echo-planar better, n = 1; PROPELLER and echo-planar equal, n = 12; P <.01, chi(2) test). PROPELLER MR imaging was preferred in all (n = 70) but one case in which the lesion lay within the intersection gap (PROPELLER preferred, n = 69; echo-planar preferred, n = 1; P <.01, chi(2) test).
CONCLUSION
With a short increase in imaging time, PROPELLER MR imaging offers better image quality and detection of acute cerebral infarction than does echo-planar MR imaging.
[1]
W. Mali,et al.
Diffusion-weighted magnetic resonance imaging in acute stroke.
,
1998,
Stroke.
[2]
J. Pekar,et al.
Echo-planar imaging of intravoxel incoherent motion.
,
1990,
Radiology.
[3]
J G Pipe,et al.
PROPELLER MRI: Clinical testing of a novel technique for quantification and compensation of head motion
,
2001,
Journal of magnetic resonance imaging : JMRI.
[4]
J. Kurhanewicz,et al.
Diffusion-weighted MR imaging of acute stroke: correlation with T2-weighted and magnetic susceptibility-enhanced MR imaging in cats.
,
1990,
AJNR. American journal of neuroradiology.
[5]
James G Pipe,et al.
Multishot diffusion‐weighted FSE using PROPELLER MRI
,
2002,
Magnetic resonance in medicine.
[6]
D Revel,et al.
Dental materials and magnetic resonance imaging.
,
1991,
Investigative radiology.
[7]
M. Raichle,et al.
Anatomic Localization and Quantitative Analysis of Gradient Refocused Echo-Planar fMRI Susceptibility Artifacts
,
1997,
NeuroImage.
[8]
E Kanal,et al.
Aneurysm clips: evaluation of MR imaging artifacts at 1.5 T.
,
1998,
Radiology.
[9]
J. Pipe.
Motion correction with PROPELLER MRI: Application to head motion and free‐breathing cardiac imaging
,
1999,
Magnetic resonance in medicine.