A Rotational Cylindrical fMRI Phantom for Image Quality Control

Purpose A novel phantom for image quality testing for functional magnetic resonance imaging (fMRI) scans is described. Methods The cylindrical, rotatable, ~4.5L phantom, with eight wedge-shaped compartments, is used to simulate rest and activated states. The compartments contain NiCl2 doped agar gel with alternating concentrations of agar (1.4%, 1.6%) to produce T1 and T2 values approximating brain grey matter. The Jacard index was used to compare the image distortions for echo planar imaging (EPI) and gradient recalled echo (GRE) scans. Contrast to noise ratio (CNR) was compared across the imaging volume for GRE and EPI. Results The mean T2 for the two agar concentrations were found to be 106.5±4.8, 94.5±4.7 ms, and T1 of 1500±40 and 1485±30 ms, respectively. The Jacard index for GRE was generally found to be higher than for EPI (0.95 versus 0.8). The CNR varied from 20 to 50 across the slices and echo times used for EPI scans, and from 20 to 40 across the slices for the GRE scans. The phantom provided a reproducible CNR over 25 days. Conclusions The phantom provides a quantifiable signal change over a head-size imaging volume with EPI and GRE sequences, which was used for image quality assessment.

[1]  D J Holt,et al.  Striatal function in relation to negative symptoms in schizophrenia , 2011, Psychological Medicine.

[2]  Paul S. Tofts,et al.  QA: Quality Assurance, Accuracy, Precision and Phantoms , 2004 .

[3]  David Thomasson,et al.  Head-repositioning does not reduce the reproducibility of fMRI activation in a block-design motor task , 2011, NeuroImage.

[4]  A. Engel,et al.  Single-trial EEG–fMRI reveals the dynamics of cognitive function , 2006, Trends in Cognitive Sciences.

[5]  R. Edden,et al.  Longitudinal and multi‐echo transverse relaxation times of normal breast tissue at 3 Tesla , 2010, Journal of magnetic resonance imaging : JMRI.

[6]  Frans A. J. Verstraten,et al.  Dynamics of visual recognition revealed by fMRI , 2006, NeuroImage.

[7]  Markus Barth,et al.  Contrast‐to‐noise ratio (CNR) as a quality parameter in fMRI , 2007, Journal of magnetic resonance imaging : JMRI.

[8]  Nicolaas Bloembergen,et al.  Proton Relaxation Times in Paramagnetic Solutions , 1957 .

[9]  Darren R. Fuerst Clinical applications of functional brain MRI, S. Rombouts, F. Barkhoff, P. Scheltens (Eds.). Oxford University Press, Great Clarendon Street, Oxford (2007), 386 pages, OX2 6DP, ISBN 978-0-19-856629-8, $245.00 (US) , 2009 .

[10]  Lee Friedman,et al.  Report on a multicenter fMRI quality assurance protocol , 2006, Journal of magnetic resonance imaging : JMRI.

[11]  Anders Nilsson,et al.  A two-compartment gel phantom for optimization and quality assurance in clinical BOLD fMRI. , 2008, Magnetic resonance imaging.

[12]  David J. McGonigle,et al.  Test–retest reliability in fMRI: Or how I learned to stop worrying and love the variability , 2012, NeuroImage.

[13]  N. Kanwisher,et al.  The Fusiform Face Area: A Module in Human Extrastriate Cortex Specialized for Face Perception , 1997, The Journal of Neuroscience.

[14]  Ralf Deichmann,et al.  Resting state fMRI reveals increased subthalamic nucleus–motor cortex connectivity in Parkinson's disease , 2011, NeuroImage.

[15]  Steve Majerus,et al.  The neural correlates of verbal short-term memory in Alzheimer's disease: an fMRI study. , 2009, Brain : a journal of neurology.

[16]  Michael B. Miller,et al.  How reliable are the results from functional magnetic resonance imaging? , 2010, Annals of the New York Academy of Sciences.

[17]  P. Tofts Quantitative MRI of the Brain , 2003 .

[18]  Helmut Laufs,et al.  A personalized history of EEG–fMRI integration , 2012, NeuroImage.

[19]  Yihong Yang,et al.  Implicit reference-based group-wise image registration and its application to structural and functional MRI , 2009, NeuroImage.

[20]  Stephen M. Smith,et al.  Multiplexed Echo Planar Imaging for Sub-Second Whole Brain FMRI and Fast Diffusion Imaging , 2010, PloS one.

[21]  D. Tank,et al.  4 Tesla gradient recalled echo characteristics of photic stimulation‐induced signal changes in the human primary visual cortex , 1993 .

[22]  Lutz Jäncke,et al.  Brain size and grey matter volume in the healthy human brain , 2002, Neuroreport.

[23]  Susumu Kanazawa,et al.  Production of a human-tissue-equivalent MRI phantom: optimization of material heating. , 2008, Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine.

[24]  George R Duensing,et al.  SmartPhantom--an fMRI simulator. , 2006, Magnetic resonance imaging.

[25]  E. Atalar,et al.  Monitoring and correcting spatio-temporal variations of the MR scanner’s static magnetic field , 2006, Magnetic Resonance Materials in Physics, Biology and Medicine.

[26]  M. Bronskill,et al.  T1, T2 relaxation and magnetization transfer in tissue at 3T , 2005, Magnetic resonance in medicine.

[27]  Gaby S Pell,et al.  Voxel-based relaxometry: a new approach for analysis of T2 relaxometry changes in epilepsy , 2004, NeuroImage.

[28]  W. Nobnop,et al.  Quality assurance. , 1998, Nursing standard (Royal College of Nursing (Great Britain) : 1987).

[29]  D. Tank,et al.  Brain magnetic resonance imaging with contrast dependent on blood oxygenation. , 1990, Proceedings of the National Academy of Sciences of the United States of America.

[30]  Nicolaas Bloembergen,et al.  Proton Relaxation Times in Paramagnetic Solutions. Effects of Electron Spin Relaxation , 1961 .

[31]  R W Cox,et al.  Software tools for analysis and visualization of fMRI data , 1997, NMR in biomedicine.

[32]  Peter A. Bandettini,et al.  Physiological noise effects on the flip angle selection in BOLD fMRI , 2011, NeuroImage.

[33]  P. Jezzard,et al.  Sources of distortion in functional MRI data , 1999, Human brain mapping.

[34]  Ranjan Maitra,et al.  A re-defined and generalized percent-overlap-of-activation measure for studies of fMRI reproducibility and its use in identifying outlier activation maps , 2010, NeuroImage.

[35]  M. Fukunaga,et al.  Sources of functional magnetic resonance imaging signal fluctuations in the human brain at rest: a 7 T study. , 2009, Magnetic resonance imaging.

[36]  S. Posse,et al.  Enhancement of BOLD‐contrast sensitivity by single‐shot multi‐echo functional MR imaging , 1999, Magnetic resonance in medicine.