Correlation of pre-operative MRI and intra-operative 3D ultrasound to measure brain tissue shift

The usefulness of stereotactic neurosurgery performed via a craniotomy is limited because the craniotomy leads to a brain tissue shift of 10 mm on average. We have recently completed an examination of 2D intra-operative ultrasound as a means of visualization and measurement of brain shift. A commercial 3D tracking system was used for real-time registration of the ultrasound video to pre-operative MIt images, and annotation of the images was used to measure the shift. More than 15 surgical cases have been performed thus far with the 2D system. We are now undertaking phantom studies with tracked 3D ultrasound, and have developed sophisticated tools for real-time overlay of ultrasound and MRI volumes. These tools include a virtual-reality view of the ultrasound probe with live ultrasound video superimposed over a 3D-rendered MRI of the brain, as well as 3D ultrasound/MRI transparency overlay views. Algorithms to automatically extract landmarks from MRJ and 3D ultrasound images are under development. We aim to use these landmarks to automatically generate nonlinear warp transformations to correct the pre-operative MRI as well as surgical target coordinates for brain shift. Portions of the C++ code developed for this project have been contributed to the open-source Visualization Toolkit (VTK).

[1]  R W Prager,et al.  Rapid calibration for 3-D freehand ultrasound. , 1998, Ultrasound in medicine & biology.

[2]  T. Peters,et al.  Intraoperative US in interactive image-guided neurosurgery. , 1998, Radiographics : a review publication of the Radiological Society of North America, Inc.

[3]  A. Jödicke,et al.  Intraoperative three-dimensional ultrasonography: an approach to register brain shift using multidimensional image processing. , 1998, Minimally invasive neurosurgery : MIN.

[4]  J. Patrick Intraoperative Brain Shift and Deformation: A Quantitative Analysis of Cortical Displacement in 28 Cases , 1998 .

[5]  T. Peters,et al.  Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery. , 2000, Medical physics.

[6]  Terry M. Peters,et al.  Level-set surface segmentation and registration for computing intrasurgical deformations , 1999, Medical Imaging.

[7]  Jason Trobaugh,et al.  The correction of stereotactic inaccuracy caused by brain shift using an intraoperative ultrasound device , 1997, CVRMed.

[8]  B. Rutt,et al.  Polyvinyl alcohol cryogel: An ideal phantom material for MR studies of arterial flow and elasticity , 1997, Magnetic resonance in medicine.

[9]  I Mano,et al.  New polyvinyl alcohol gel material for MRI phantoms , 1986, Magnetic resonance in medicine.

[10]  K. Takakura,et al.  Development of a frameless and armless stereotactic neuronavigation system with ultrasonographic registration. , 1997, Neurosurgery.

[11]  Terry M. Peters,et al.  Level-Set Surface Segmentation and Fast Cortical Range Image Tracking for Computing Intrasurgical Deformations , 1999, MICCAI.

[12]  M. Y. Wang,et al.  Measurement of Intraoperative Brain Surface Deformation Under a Craniotomy , 1998, MICCAI.

[13]  K Doi,et al.  Automated tracking and computer reproduction of vessels in DSA images. , 1990, Investigative radiology.

[14]  Fred L. Bookstein,et al.  Linear Methods for Nonlinear Maps , 1999 .

[15]  C Giorgi,et al.  Preliminary clinical experience with intraoperative stereotactic ultrasound imaging. , 1997, Stereotactic and functional neurosurgery.