Titration of Power Output During Radiofrequency Catheter Ablation of Atrioventricular Nodal Reentrant Tachycardia

Radiofrequency lesions in the anterior, superior aspect of the tricuspid annulus result in selective elimination of fast pathway function in patients with typical atrioventricular (AV) nodal reentry tachycardia. This technique is simple and effective, but has been associated with a significant risk of inadvertent complete AV block. The purpose of this study was to compare the safety and effectiveness of two different techniques for radiofrequency catheter ablation of the fast AV nodal pathway. Initially, a fixed power output was used at each target site. This method was compared retrospectively to a newer technique where power output was gradually incremented at each site. Radiofrequency power was initially applied at 10 watts for 10–15 seconds. If no junctional ectopy or a change in PR intervoi was seen, power output was incremented by 2 to 4 watts every 10 to 15 seconds up to a maximum of 30 watts. Thirty seven of 38 (96%) patients treated using this incremental power output were cured of their AV nodal reentry tachycardia. None of these patients developed inadvertent complete AV block. In contrast, 92% of historic controls treated with a fixed power output between 20 and 30 watts achieved a primary success and nine of these 89 (10%) historic controls developed inadvertent complete AV block (P = 0.04). There was no difference in the amplitudes of atrial, His, or ventricular electrograms at the effective sites between the two groups. Conclusions: (1) the anterior approach to radiofrequency catheter ablation of typical AV nodal reentry is associated with a significant risk of inadvertent complete AV block if a fixed power output is used; (2) starting at low power and gradually incrementing the output during radiofrequency energy application reduces the risk of complete AV block; (3) this incremental technique does not compromise efficacy.

[1]  D D Watson,et al.  Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free Wall , 1989, Pacing and clinical electrophysiology : PACE.

[2]  G. Kay,et al.  Selective Radiofrequency Ablation of the Slow Pathway for the Treatment of Atrioventricular Nodal Reentrant Tachycardia: Evidence for Involvement of Perinodal Myocardium Within the Reentrant Circuit , 1992, Circulation.

[3]  J. Langberg,et al.  Diagnosis and cure of the Wolff-Parkinson-White syndrome or paroxysmal supraventricular tachycardias during a single electrophysiologic test. , 1991, The New England journal of medicine.

[4]  J. Goy,et al.  Clinical efficacy of radiofrequency current in the treatment of patients with atrioventricular node reentrant tachycardia. , 1990, Journal of the American College of Cardiology.

[5]  R. Lazzara,et al.  Treatment of supraventricular tachycardia due to atrioventricular nodal reentry by radiofrequency catheter ablation of slow-pathway conduction. , 1992, The New England journal of medicine.

[6]  R N Hauer,et al.  Control of radiofrequency lesion size by power regulation. , 1989, Circulation.

[7]  D. Krum,et al.  Selective Transcatheter Ablation of the Fast and Slow Pathways Using Radiofrequency Energy in Patients With Atrioventricular Nodal Reentrant Tachycardia , 1992, Circulation.

[8]  D. Commenges,et al.  Elimination of Atrioventricular Nodal Reentrant Tachycardia Using Discrete Slow Potentials to Guide Application of Radiofrequency Energy , 1992, Circulation.

[9]  Michael D. Lesh,et al.  Catheter Modification of the Atrioventricular Junction With Radiofrequency Energy for Control of Atrioventricular Nodal Reentry Tachycardia , 1991, Circulation.