Complex impedance measurement system for environmental sensors characterization

In this paper description of complex impedance measurement system for environmental sensors characterization is described. Device can be used as standalone device but it also can be connected to the PC through serial communication. Color display is used for local graphical presentation of measured values while for online PC analysis specialized application was developed. System verification was done in frequency range from 5 to 100 kHz for impedance range from 100 'Ω to 100 k'Ω. Verification of impedance magnitude and phase angle measurement results as done as proof that developed system can be used for characterization of sensors for air, soil and water pollution monitoring developed within SENSEIVER project.

[1]  Mitar Simic,et al.  Realization of digital LCR meter , 2014, 2014 International Conference and Exposition on Electrical and Power Engineering (EPE).

[2]  Weisong Shi,et al.  Data Quality and Failures Characterization of Sensing Data in Environmental Applications , 2008, CollaborateCom.

[3]  Mitar Simic,et al.  Realization of Complex Impedance Measurement System Based on the Integrated Circuit AD5933 , 2013, 2013 21st Telecommunications Forum Telfor (TELFOR).

[4]  M. Simić Complex Impedance Measurement System for the Frequency Range from 5 kHz to 100 kHz , 2015 .