A scintillating gas detector for 2D dose measurements in clinical carbon beams

A two-dimensional position sensitive dosimetry system based on a scintillating gas detector has been developed for pre-treatment verification of dose distributions in hadron therapy. The dosimetry system consists of a chamber filled with an Ar/CF4 scintillating gas mixture, inside which two cascaded gas electron multipliers (GEMs) are mounted. A GEM is a thin kapton foil with copper cladding structured with a regular pattern of sub-mm holes. The primary electrons, created in the detector's sensitive volume by the incoming beam, drift in an electric field towards the GEMs and undergo gas multiplication in the GEM holes. During this process, photons are emitted by the excited Ar/CF4 gas molecules and detected by a mirror-lens-CCD camera system. Since the amount of emitted light is proportional to the dose deposited in the sensitive volume of the detector by the incoming beam, the intensity distribution of the measured light spot is proportional to the 2D hadron dose distribution. For a measurement of a 3D dose distribution, the scintillating gas detector is mounted at the beam exit side of a water-bellows phantom, whose thickness can be varied in steps. In this work, the energy dependence of the output signal of the scintillating gas detector has been verified in a 250 MeV/u clinical 12C ion beam by means of a depth–dose curve measurement. The underestimation of the measured signal at the Bragg peak depth is only 9% with respect to an air-filled ionization chamber. This is much smaller than the underestimation found for a scintillating Gd2O2S:Tb (‘Lanex’) screen under the same measurement conditions (43%). Consequently, the scintillating gas detector is a promising device for verifying dose distributions in high LET beams, for example to check hadron therapy treatment plans which comprise beams with different energies.

[1]  C. V. van Eijk,et al.  First Results of a Scintillating GEM Detector for 2-D Dosimetry in an Alpha Beam , 2007, IEEE Transactions on Nuclear Science.

[2]  Dieter Schardt,et al.  Tumor therapy with high-energy carbon ion beams , 2007 .

[3]  Sairos Safai,et al.  Development of an inorganic scintillating mixture for proton beam verification dosimetry. , 2004, Physics in medicine and biology.

[4]  J. D. Haas,et al.  Dose imaging in radiotherapy with an Ar–CF4 filled scintillating GEM , 2003 .

[5]  M. Fraga,et al.  The GEM scintillation in He–CF4, Ar–CF4, Ar–TEA and Xe–TEA mixtures , 2003 .

[6]  M. Charles,et al.  A wide dynamic range, high-spatial-resolution scanning system for radiochromic dye films. , 2001, Physics in medicine and biology.

[7]  E. Pedroni,et al.  Intensity modulated proton therapy: a clinical example. , 2001, Medical physics.

[8]  A. Policarpo,et al.  Performance of a tracking device based on the GEM scintillation , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[9]  E Pedroni,et al.  Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniques. , 2000, Medical physics.

[10]  M. Bock,et al.  Three-dimensional BANG gel dosimetry in conformal carbon ion radiotherapy. , 2000, Physics in medicine and biology.

[11]  Michael Krämer,et al.  A dosimetry system for fast measurement of 3D depth–dose profiles in charged-particle tumor therapy with scanning techniques , 2000 .

[12]  G. Kraft,et al.  Design and construction of a ripple filter for a smoothed depth dose distribution in conformal particle therapy. , 1999, Physics in medicine and biology.

[13]  O Jäkel,et al.  A system for three-dimensional dosimetric verification of treatment plans in intensity-modulated radiotherapy with heavy ions. , 1999, Medical physics.

[14]  P. Fonte,et al.  Breakdown limit studies in high-rate gaseous detectors , 1999 .

[15]  T. Kanai,et al.  Initial recombination in a parallel-plate ionization chamber exposed to heavy ions. , 1998, Physics in medicine and biology.

[16]  Sjirk N. Boon,et al.  Fast 2D phantom dosimetry for scanning proton beams. , 1998, Medical physics.

[17]  G. Mazza,et al.  Performances of a VLSI wide dynamic range current-to-frequency converter for strip ionization chambers , 1998 .

[18]  L E Olsson,et al.  MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gels. , 1990, Physics in medicine and biology.

[19]  F. Sauli,et al.  Drift and diffusion of electrons in gases: a compilation , 1984 .

[20]  Fabio Sauli,et al.  Principles of operation of multiwire proportional and drift chambers , 1977 .

[21]  Y. Reibel,et al.  CCD or CMOS camera noise characterisation , 2003 .

[22]  J. Gore,et al.  NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: a new approach to 3D dosimetry by MRI. , 1993, Magnetic resonance imaging.