Segmented interlocking nail: An in vivo evaluation of a novel humeral osteotomy fixation device in a caprine model

OBJECTIVES To describe a novel humeral fixation device, the insertion technique, healing of humeral osteotomies, and clinical outcomes in a caprine model over a six month period. METHODS Fourteen mature female Boer/Nubian cross goats with a mean body weight of 50.7 kg were implanted with a proprietary segmented interlocking nail (SILN) in both humeri. Each goat had one humerus randomly selected for mid-diaphyseal osteotomy. RESULTS Immediately after surgery all but one goat was able to stand, although none of the goats were weight bearing on the osteotomy limb. During the six month study, clinical lameness was always associated with the osteotomy limb. One month after surgery, lameness for twelve of the goats was grade 2/5 or better. At three months, 11 of the 14 did not exhibit any signs of lameness. On radiographic images, notable malalignment of the osteotomy was observed, although all osteotomies went to bone union. CLINICAL SIGNIFICANCE The results of this study suggest that despite misalignment, the SILN maintained adequate osteotomy fixation to achieve bone union in the research model studied, with reduced morbidity and early return to function with bilateral implantation. The SILN used in this study allowed intramedullary fixation of humeral diaphyseal osteotomies with a limited and safe surgical approach.

[1]  E. Chao,et al.  Structural properties of interlocking nails, canine femora, and femur-interlocking nail constructs. , 1996, Veterinary surgery : VS.

[2]  P. Hyvönen,et al.  Intramedullary nailing of humeral shaft fractures. A retrospective study of 126 cases. , 1999, Acta orthopaedica Scandinavica.

[3]  H. Lin,et al.  Effects of anesthesia induced and maintained by continuous intravenous administration of guaifenesin, ketamine, and xylazine in spontaneously breathing sheep. , 1993, American journal of veterinary research.

[4]  R. Buckley,et al.  Fixation of fractures of the shaft of the humerus by dynamic compression plate or intramedullary nail , 2000 .

[5]  I. Durall,et al.  Interlocking Nail Stabilisation of Humeral Fractures. Initial Experience in Seven Clinical Cases , 1994, Veterinary and Comparative Orthopaedics and Traumatology.

[6]  M T MODNY,et al.  THE PERFORATED CRUCIATE INTRAMEDULLARY NAIL: PRELIMINARY REPORT OF ITS USE IN GERIATRIC PATIENTS , 1953, Journal of the American Geriatrics Society.

[7]  I. Durall,et al.  Early experience with the use of an interlocking nail for the repair of canine femoral shaft fractures. , 1996, Veterinary surgery : VS.

[8]  S. Liu,et al.  Osteomyelitis and neoplasia associated with use of the Jonas intramedullary splint in small animals. , 1982, Journal of the American Veterinary Medical Association.

[9]  A. Cross,et al.  Intramedullary interlocking nail stabilisation of 21 humeral fractures in 19 dogs and one cat. , 2002, Australian veterinary journal.

[10]  T. Albert,et al.  Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern? , 2004, The spine journal : official journal of the North American Spine Society.

[11]  S. Perren,et al.  [Titanium, steel and their combination in bone surgery]. , 1975, Langenbecks Archiv fur Chirurgie.

[12]  G. McGwin,et al.  Intramedullary nailing of humeral shaft fractures with a locking flexible nail. , 2003, The Journal of bone and joint surgery. American volume.

[13]  R. Haut,et al.  Effect of bending direction on the mechanical behaviour of interlocking nail systems , 2009, Veterinary and Comparative Orthopaedics and Traumatology.

[14]  J. McElwain,et al.  Closed interlocking nailing of humeral shaft fractures with the Russell-Taylor nail. , 2000, Journal of orthopaedic trauma.

[15]  R. Haut,et al.  In vitro evaluation of the effect of fracture configuration on the mechanical properties of standard and novel interlocking nail systems in bending. , 2009, Veterinary surgery : VS.

[16]  A. Diop,et al.  An in vitro biomechanical study of bone plate and interlocking nail in a canine diaphyseal femoral fracture model. , 2001, Veterinary surgery : VS.

[17]  R. Haut,et al.  In vitro biomechanical comparison of a plate-rod combination-construct and an interlocking nail-construct for experimentally induced gap fractures in canine tibiae. , 2005, American journal of veterinary research.

[18]  P. Muir,et al.  Interlocking intramedullary nail stabilisation of a diaphyseal tibial fracture , 1993 .

[19]  T. Bhullar,et al.  Interlocking nailing of humeral shaft fractures: the Oxford experience 1991 to 1994. , 1996, Injury.

[20]  P. Stern,et al.  Humeral nonunion associated with metallosis secondary to use of a titanium flexible humeral intramedullary nail: a case report. , 2002, The Journal of bone and joint surgery. American volume.

[21]  R. Haut,et al.  In vitro mechanical comparison of screwed, bolted, and novel interlocking nail systems to buttress plate fixation in torsion and mediolateral bending. , 2007, Veterinary surgery : VS.

[22]  R. Haut,et al.  Mechanical comparison of an interlocking nail locked with conventional bolts to extended bolts connected with a type-IA external skeletal fixator in a tibial fracture model. , 2007, Veterinary surgery : VS.

[23]  K. A. Johnson,et al.  Interlocking nail treatment of diaphyseal long-bone fractures in dogs. , 1999, Journal of the American Veterinary Medical Association.

[24]  Kenneth A Johnson,et al.  Area Moment of Inertia for Comparison of Implant Cross-Sectional Geometry and Bending Stiffness , 1995, Veterinary and Comparative Orthopaedics and Traumatology.

[25]  R. Haut,et al.  In vitro mechanical evaluation of torsional loading in simulated canine tibiae for a novel hourglass-shaped interlocking nail with a self-tapping tapered locking design. , 2006, American journal of veterinary research.

[26]  P. Boerlin,et al.  A prospective study of postoperative surgical site infections in dogs and cats. , 2004, Veterinary surgery : VS.