An open-loop controlled active lung simulator for preterm infants.

We describe the underlying theory, design and experimental evaluation of an electromechanical analogue infant lung to simulate spontaneous breathing patterns of preterm infants. The aim of this work is to test the possibility to obtain breathing patterns of preterm infants by taking into consideration the air compressibility. Respiratory volume function represents the actuation pattern, and pulmonary pressure and flow-rate waveforms are mathematically obtained through the application of the perfect gas and adiabatic laws. The mathematical model reduces the simulation interval into a step shorter than 1 ms, allowing to consider an entire respiratory act as composed of a large number of almost instantaneous adiabatic transformations. The device consists of a spherical chamber where the air is compressed by four cylinder-pistons, moved by stepper motors, and flows through a fluid-dynamic resistance, which also works as flow-rate sensor. Specifically designed software generates the actuators motion, based on the desired ventilation parameters, without controlling the gas pneumatic parameters with a closed-loop. The system is able to simulate tidal volumes from 3 to 8 ml, breathing frequencies from 60 to 120 bpm and functional residual capacities from 25 to 80 ml. The simulated waveforms appear very close to the measured ones. Percentage differences on the tidal volume waveform vary from 7% for the tidal volume of 3 ml, down to 2.2-3.5% for tidal volumes in the range of 4-7 ml, and 1.3% for the tidal volume equal to 8 ml in the whole breathing frequency and functional residual capacity ranges. The open-loop electromechanical simulator shows that gas compressibility can be theoretically assessed in the typical pneumatic variable range of preterm infant respiratory mechanics.

[1]  A. L. Jung,et al.  Functional residual capacity and severity of respiratory distress syndrome in infants , 1980, Critical care medicine.

[2]  D. Morel,et al.  Volume-targeted modes of modern neonatal ventilators: how stable is the delivered tidal volume? , 2007, Intensive Care Medicine.

[3]  C. McEvoy,et al.  Functional residual capacity and passive compliance measurements after antenatal steroid therapy in preterm infants , 2001, Pediatric pulmonology.

[4]  Sergio Silvestri The influence of flow rate on breathing circuit compliance and tidal volume delivered to patients in mechanical ventilation. , 2006, Physiological measurement.

[5]  J D Horbar,et al.  Very low birth weight outcomes of the National Institute of Child Health and Human Development Neonatal Network. , 1991, Pediatrics.

[6]  Sergio Silvestri,et al.  A novel preterm respiratory mechanics active simulator to test the performances of neonatal pulmonary ventilators , 2002 .

[7]  J. E. W. Beneken,et al.  Computer-controlled mechanical lung model for application in pulmonary function studies , 1995, Medical and Biological Engineering and Computing.

[8]  M. Lange,et al.  Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships , 2007, European Journal of Applied Physiology.

[9]  O. Hjalmarson,et al.  Lung function at term reflects severity of bronchopulmonary dysplasia. , 2005, The Journal of pediatrics.

[10]  Peter G Davis,et al.  Breathing Patterns in Preterm and Term Infants Immediately After Birth , 2009, Pediatric Research.

[11]  R. Ramanathan,et al.  Lung protective ventilatory strategies in very low birth weight infants , 2008, Journal of Perinatology.

[12]  Robert M. Kacmarek,et al.  Adult ICU ventilators to provide neonatal ventilation: a lung simulator study , 2009, Intensive Care Medicine.

[13]  E. Bancalari,et al.  Methods and evidence on volume-targeted ventilation in preterm infants , 2008, Current opinion in pediatrics.

[14]  G. Schmalisch,et al.  Changes in pulmonary function in preterm infants recovering from RDS following early treatment with ambroxol: Results of a randomized trial , 1999, Pediatric pulmonology.

[15]  J. Stocks,et al.  The infant lung function model: a mechanical analogue to test infant lung function equipment. , 2001, The European respiratory journal.

[16]  J. Mortola,et al.  Dynamics of breathing in infants. , 1982, Journal of applied physiology: respiratory, environmental and exercise physiology.

[17]  V. V. Meka,et al.  Bellows-less lung system for the human patient simulator , 2006, Medical and Biological Engineering and Computing.

[18]  Sergio Silvestri,et al.  A novel device to simulate low respiratory static compliances of pre-term infants measuring expansion tidal volume , 2003, Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439).

[19]  R. W. Miller,et al.  Flow Measurement Engineering Handbook , 1983 .

[20]  J. Wagener,et al.  Pulmonary mechanics in healthy term neonates: variability in measurements obtained with a computerized system. , 1990, The Journal of pediatrics.

[21]  N. Desai,et al.  Methods of assessment and findings regarding pulmonary function in infants less than 1000 grams. , 1986, Clinics in perinatology.

[22]  J. Pillow,et al.  Injury and Inflammation from Resuscitation of the Preterm Infant , 2008, Neonatology.

[23]  A. Milner,et al.  Performance of neonatal ventilators in volume targeted ventilation mode , 2007, Acta paediatrica.

[24]  J. E. W. Beneken,et al.  A new approach to mechanical simulation of lung behaviour: Pressure-controlled and time-related piston movement , 2006, Medical and Biological Engineering and Computing.

[25]  S. Iijima,et al.  Clinical patterns in extremely preterm (22 to 24 weeks of gestation) infants in relation to survival time and prognosis. , 2009, American journal of perinatology.

[26]  Robert Babuska,et al.  Computer-controlled mechanical simulation of the artificially ventilated human respiratory system , 2003, IEEE Transactions on Biomedical Engineering.