Electrical admittance method for estimating fluid removal during artificial dialysis

Because the conductivity of blood changes remarkably during artificial dialysis, sometimes by more than 20%, changes in tissue admittance at low frequency are caused by changes not only in the extracellular fluid volume but also in blood conductivity. Therefore the changes in blood conductance due to artificial dialysis must be considered for the estimation of water removal by the admittance method. An accurate bio-impedance measurement system was developed. Measurement error was less than 1% at low frequency and 10% at high frequency. A new electrical bio-admittance method (NAM) was evaluated for the continous measurement of removed fluid volume, using a three parallel-compartment tissue model, consisting of intracellular, interstitial and blood compartments, which takes into account the blood conductivity change. NAM used the equivalent conductivity calculated from the leg admittances of patients, measured at 1 min intervals during various artificial dialysis procedures. The actual amount of excess water removed by ultra-filtration agreed with the NAM-estimated amount within an error of 20%. NAM was also applied to estimate the intra-and extracellular fluid changes. The results were consistent with the physiological changes known to occur during the various forms of dialysis.

[1]  Sverre Grimnes,et al.  Bioimpedance and Bioelectricity Basics , 2000 .

[2]  L. Geddes,et al.  Specific resistance of blood at body temperature II , 1976, Medical and biological engineering.

[3]  Art Lew On Analog Computer Generation of Continuous Functions , 1977, IEEE Transactions on Computers.

[4]  Tatsuo Togawa,et al.  Noninvasive Measurement of Hematocrit by Electrical Admittance Plethysmography Technique , 1980, IEEE Transactions on Biomedical Engineering.

[5]  Ronald S. Newbower,et al.  A Practical Analysis of the Electrical Conductivity of Blood , 1983, IEEE Transactions on Biomedical Engineering.

[6]  H Kanai,et al.  Study of the Relation between Fluid Distribution Change in Tissue and Impedance Change during Hemodialysis by Frequency Characteristics of the Flowing Blood , 1999, Annals of the New York Academy of Sciences.

[7]  Fluids and Electrolytes: A Conceptual Approach , 1980 .

[8]  P. Soeters,et al.  Determination of total body water by multifrequency bio-electric impedance: development of several models , 2006, Medical and Biological Engineering and Computing.

[9]  The Impedance Plethysmographic Sampling Field in the Human Calf , 1979, IEEE Transactions on Biomedical Engineering.

[10]  H Kanai,et al.  Electrical measurement of fluid distribution in human legs: estimation of extra- and intra-cellular fluid volume. , 1983, The Journal of microwave power.

[11]  H. Kanai,et al.  [Estimation of fluid distribution by impedance method]. , 1985, Iyo denshi to seitai kogaku. Japanese journal of medical electronics and biological engineering.

[12]  H. P. Schwan,et al.  Alternating Current Spectroscopy of Biological Substances , 1959, Proceedings of the IRE.

[13]  L. A. Geddes,et al.  The specific resistance of blood at body temperature , 1973, Medical and biological engineering.

[14]  Hiroshi Kanai,et al.  Electrical Characteristics of Flowing Blood , 1979, IEEE Transactions on Biomedical Engineering.

[15]  Willis J. Tompkins,et al.  Design of Microcomputer-Based Medical Instrumentation , 1981 .

[16]  H. Schwan,et al.  An electrical method to determine hematocrits. , 1960, IRE transactions on medical electronics.

[17]  E Gheorghiu On the Limits of Ellipsoidal Models when Analyzing Dielectric Behavior of Living Cells Emphasis on Red Blood Cells , 1999, Annals of the New York Academy of Sciences.

[18]  Measurement of body fluid volume change using multisite impedance measurements , 2006, Medical and Biological Engineering and Computing.

[19]  K. Nakayama,et al.  Measurement of Leg Blood Volume Change by Impedance Plethysmography with Special Reference to Microgravity Simulation Testing , 1988 .

[20]  H. Schwan,et al.  Biological Engineering , 1970 .

[21]  Relative Underestimation of Fluid Removal During Hemodialysis Hypotension Measured by Whole Body Bioimpedance , 1998, ASAIO journal.

[22]  L. Geddes,et al.  The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologist , 1967, Medical and biological engineering.