Different responses of Finapres and the oscillometric finger blood pressure monitor during intensive vasomotion.

Two different techniques for non-invasive beat-to-beat finger arterial blood pressure monitoring are compared in six healthy volunteers during local hand heating from 21 to 38 degrees C. The degree of peripheral vasoconstriction was established by recording the thumb pulp skin blood flow with a laser Doppler instrument. For time episodes without vasoconstriction no systematic difference in the readings of beat-to-beat mean blood pressure of the two monitors was found (the oscillometric device UT9201 minus Finapres difference was 0.3 mm Hg, SD 0.3). For the episodes with vasoconstriction the difference was statistically significant (6.7 mm Hg, SD 2.0). The oscillometric device minus Finapres difference and the laser Doppler signal were found to be inversely correlated, the correlation coefficient varying from -0.28 to -0.67. A disagreement between the readings of the instruments during intensive vasomotion is assumed to be caused mainly by the tendency of the oscillometric method to overestimate the finger mean blood pressure under the condition of peripheral vasoconstriction.

[1]  M. Eriksen,et al.  Fluctuations in blood flow to acral skin in humans: connection with heart rate and blood pressure variability. , 1993, The Journal of physiology.

[2]  K H Wesseling,et al.  Effects of peripheral vasoconstriction on the measurement of blood pressure in a finger. , 1985, Cardiovascular research.

[3]  K H Wesseling,et al.  Why use Finapres or Portapres rather than intra-arterial or intermittent non-invasive techniques of blood pressure measurement? , 1998, Journal of medical engineering & technology.

[4]  K. Wesseling,et al.  Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. , 1998, Cardiovascular research.

[5]  Y Iyriboz Oscillometric finger blood pressure versus brachial auscultative blood pressure recording. , 1990, The Journal of family practice.

[6]  T. Thien,et al.  LAM 100/Marshall F-88: accuracy and precision of a new device for discontinuous finger blood pressure measurement. , 1993, Journal of human hypertension.

[7]  D. Phillips,et al.  Accuracy of automated finger blood pressure devices. , 1996, Family medicine.

[8]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[9]  G. Parati,et al.  Effects of graded vasoconstriction upon the measurement of finger arterial pressure. , 1992, Journal of hypertension.

[10]  J. J. Settels,et al.  Physiocal, calibrating finger vascular physiology for finapres , 1995 .

[11]  Short-term cardiovascular responses to a step decrease in peripheral conductance in humans. , 1994, The American journal of physiology.

[12]  E Länsimies,et al.  Continuous non-invasive measurement of mean blood pressure in fingers by volume-clamp and differential oscillometric method. , 1996, Clinical physiology.

[13]  M. Irwin,et al.  The Finapres 2300e finger cuff , 1993, Anaesthesia.

[14]  G D Pinna,et al.  Estimation of arterial blood pressure variability by spectral analysis: comparison between Finapres and invasive measurements. , 1996, Physiological measurement.

[15]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[16]  M. Eriksen,et al.  Peripheral vasoconstriction shortly after onset of moderate exercise in humans. , 1994, Journal of applied physiology.