Acoustic power calibration of high-intensity focused ultrasound transducers using a radiation force technique.

To address the challenges associated with measuring the ultrasonic power from high-intensity focused ultrasound transducers via radiation force, a technique based on pulsed measurements was developed and analyzed. Two focused ultrasound transducers were characterized in terms of an effective duty factor, which was then used to calculate the power during the pulse at high applied power levels. Two absorbing target designs were used, and both gave comparable results and displayed no damage and minimal temperature rise if placed near the transducer and away from the focus. The method yielded reproducible results up to the maximum pulse power generated of approximately 230 W, thus allowing the radiated power to be calibrated in terms of the peak-to-peak voltage applied to the transducer.

[1]  Adam Shaw How to Measure HIFU Output Power Properly , 2006 .

[2]  Lawrence A Crum,et al.  Effects of nonlinear propagation, cavitation, and boiling in lesion formation by high intensity focused ultrasound in a gel phantom. , 2006, The Journal of the Acoustical Society of America.

[3]  G. Harris,et al.  Progress in medical ultrasound exposimetry , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[4]  C. Lafon,et al.  Safety Issues for HIFU Transducer Design , 2005 .

[5]  Gregory T. Clement,et al.  Perspectives in clinical uses of high-intensity focused ultrasound. , 2004, Ultrasonics.

[6]  B. Zeqiri,et al.  Measurement of ultrasonic power using an acoustically absorbing well. , 2003, Ultrasound in medicine & biology.

[7]  M.C. Ziskin Specification of acoustic output level and measurement uncertainty in ultrasonic exposimetry , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[8]  K. Hynynen,et al.  Interlaboratory Acoustic Power Measurement , 2003, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[9]  K. Beissner IEC measurement standards for ultrasonic hydrophones and radiation force balances , 2002 .

[10]  G. Wong,et al.  High power ultrasound standard. , 2002, The Journal of the Acoustical Society of America.

[11]  M Hodnett,et al.  Validated ultrasonic power measurements up to 20 W. , 2001, Ultrasound in medicine & biology.

[12]  B Zeqiri,et al.  A new anechoic material for medical ultrasonic applications. , 2000, Ultrasound in medicine & biology.

[13]  S. Fick Ultrasound power measurement by pulsed radiation pressure , 1999 .

[14]  George S. K. Wong,et al.  Speed of sound in pure water as a function of temperature , 1993 .

[15]  K. Hynynen,et al.  Acoustic power calibrations of cylindrical intracavitary ultrasound hyperthermia applicators. , 1993, Medical physics.

[16]  K. Beissner Minimum target size in radiation force measurements , 1984 .

[17]  R. Robinson Performance Evaluation of a Digital Readout Hyperthermia Range Ultrasonic Wattmeter , 1984, IEEE Transactions on Sonics and Ultrasonics.

[18]  G. Kossoff Balance Technique for the Measurement of Very Low Ultrasonic Power Outputs , 1965 .