Reduction of motion axis by sensitivity calibration of sensor in shape measurement instrument using normal vector tracing.

High-precision free-form surface mirrors are required for synchrotron radiation facilities in the scientific field and semiconductor lithography systems in the industrial field. Previously, we developed a nano-profiler with the goal of achieving a measurement accuracy of 30 nm. The nano-profiler scanned and measured the slope angle of the surface to be measured with laser light and calculated the shape from the angle information. By driving the optical head and surface to be measured with four rotation axes and one translation axis, the surface could be scanned while keeping the optical path length constant. Although the rotation axis was controlled by a high-precision rotary encoder, pitching and yawing errors occurred in the translation axis. In this study, we attempt to eliminate the error of pitching and yawing from the conventional measurement operation of four axes of rotation and one axis of translation to the drive of only four axes of rotation. If the translation drive is eliminated, the optical path length will not be constant, and the sensitivity of the light-receiving element will change. Therefore, we propose a new method to calibrate the sensitivity of the receiving element and perform a comparative measurement with the conventional measurement method. Comparing the measured shapes obtained by both measurement results, it was found that the shapes had a maximum peak to valley difference of 6.2 nm. Thus, the proposed novel measurement method allows a significant reduction in pitching and yawing errors.

[1]  C. Elster,et al.  Round robin comparison study on the form measurement of optical freeform surfaces , 2020 .

[2]  T. Ishikawa,et al.  Nanofocusing of X-ray free-electron laser using wavefront-corrected multilayer focusing mirrors , 2018, Scientific Reports.

[3]  Valeriy V. Yashchuk,et al.  The ALS OSMS: Optical Surface Measuring System for high accuracy two-dimensional slope metrology with state-of-the-art x-ray mirrors , 2018, Optical Engineering + Applications.

[4]  R. Henselmans,et al.  NANOMEFOS non-contact measurement machine for aspheric and freeform optics , 2017, International Conference on Space Optics.

[5]  Ryota Kudo,et al.  High-accuracy three-dimensional aspheric mirror measurement with nanoprofiler based on normal vector tracing method , 2017 .

[6]  Li Li,et al.  Extreme ultraviolet resist materials for sub-7 nm patterning. , 2017, Chemical Society reviews.

[7]  Banqiu Wu,et al.  Extreme ultraviolet lithography and three dimensional integrated circuit—A review , 2014 .

[8]  G. Berger,et al.  Non-contact metrology of aspheric surfaces based on MWLI technology , 2013, Other Conferences.

[9]  Takuya Kojima,et al.  Development of a nanoprofiler using the follow-up normal vector to the surface for next-generation ultraprecise mirrors , 2012, Optical Systems Design.

[10]  Takuya Kojima,et al.  Effects of a laser beam profile to measure an aspheric mirror on a high-speed nanoprofiler using normal vector tracing method , 2012 .

[11]  Wolfgang Osten,et al.  Systematic analysis of the measurement of cone angles using high line density computer-generated holograms , 2011 .

[12]  Kenji Ueno,et al.  Surface gradient integrated profiler for x-ray and EUV optics: self calibration method of measured position for an off-axis parabolic mirror (f=150mm) measurement , 2007, SPIE Optical Engineering + Applications.

[13]  Ping Zhou,et al.  Fabrication error analysis and experimental demonstration for computer-generated holograms. , 2007, Applied optics.

[14]  Hans J. Tiziani,et al.  Computer-generated holograms in interferometric testing , 2004 .

[15]  Fengzhou Fang,et al.  Manufacturing and measurement of freeform optics , 2013 .