Blinded multi-laboratory evaluation of an in-house dot-blot ELISA kit for diagnosis of human parastrongyliasis.

An in-house membrane dot-blot enzyme-linked immunosorbent assay kit, for the detection of specific antibody to a 31 kDa protein of Parastrongylus cantonensis was evaluated in a blinded multi-laboratory study. The kit was provided to technologists, who were trained in its use, from nine regional laboratories of Thailand. With an identical set of 20 coded serum samples and reference positive and negative controls, the kit was found to have a diagnostic sensitivity of 100% and a diagnostic specificity of 100% in all the laboratories. There was no obvious variation in quality among five lots of the antigen-coated nitrocellulose strips evaluated. The shelf life of the kit was > or = 6 months when the test components were stored at 4 degrees C and 18 months at -20 degrees C. The enzyme-linked immunoassay dot technique is easy to perform and does not require sophisticated electrical equipment; the result is available within 3 hours. If appropriate technical training is included, the application of this dot-blot ELISA kit in clinical laboratories throughout Thailand should be possible.

[1]  P. Eamsobhana,et al.  Purification of a specific immunodiagnostic Parastrongylus cantonensis antigen by electroelution from SDS-polyacrylamide gels. , 2001, The Southeast Asian journal of tropical medicine and public health.

[2]  K. Sawanyawisuth,et al.  Horizontal conjugate gaze palsy in eosinophilic meningitis. , 1999, Southeast Asian Journal of Tropical Medicine and Public Health.

[3]  N. M. Dung,et al.  Evaluation of a Rapid Immunochromatographic Test for Diagnosis of Dengue Virus Infection , 1998, Journal of Clinical Microbiology.

[4]  W. Abeyewickreme,et al.  Second case of ocular parastrongyliasis from Sri Lanka. , 1998, Transactions of the Royal Society of Tropical Medicine and Hygiene.

[5]  B. Masquelier,et al.  Multicenter evaluating of a commercially available PCR assay for diagnosing enterovirus infection in a panel of cerebrospinal fluid specimens , 1996, Journal of clinical microbiology.

[6]  M. Kliks,et al.  Eosinophilic meningitis beyond the Pacific Basin: the global dispersal of a peridomestic zoonosis caused by Angiostrongylus cantonensis, the nematode lungworm of rats. , 1992, Social science & medicine.

[7]  J. Alicata The discovery of Angiostrongylus cantonensis as a cause of human eosinophilic meningitis. , 1991, Parasitology today.

[8]  J. Cross Public health importance of Angiostrongylus cantonensis and its relatives. , 1987, Parasitology today.

[9]  R. Galen,et al.  Predictive value and efficiency of laboratory testing. , 1980, Pediatric clinics of North America.