Performance of FBK SiPMs coupled to PETA3 read-out ASIC for PET application

Abstract In this paper we show the energy and timing resolution performances of FBK SiPMs coupled to the PETA3 ASIC for PET application. We developed a measurement set-up to characterize single SiPMs coupled to scintillator exploiting the detector stack developed within the HYPERImage project. In this way we are able to characterize the combined SiPM/ASIC performance with the same signal chain (from the sensor to the ASIC board) used in the PET system. We show that using two scintillator detectors, composed of a 3 × 3 × 5 mm 3 LYSO crystal coupled to a 3 × 3 mm 2 SiPM, an intra-stack CRT of about 200 ps FWHM can be obtained.