RF Head Coil Design With Improved RF Magnetic Near-Fields Uniformity for Magnetic Resonance Imaging (MRI) Systems

Higher magnetic field strength in magnetic resonance imaging (MRI) systems offers higher signal-to-noise ratio, contrast, and spatial resolution in magnetic resonance (MR) images. However, the wavelength in ultrahigh fields (7 T and beyond) becomes shorter than the human body at the Larmor frequency with increasing static magnetic field (B0) of the MRI system. At short wavelengths, an interference effect appears, resulting in nonuniformity of the RF magnetic near-field (B1) over the subject and MR images may have spatially anomalous contrast. The B1 near-field generated by the transverse electromagnetic RF coil's microstrip line element has a maximum near the center of its length and falls off towards both ends. In this study, a double trapezoidal-shaped microstrip transmission line element is proposed to obtain uniform B1 field distribution by gradual impedance variation. Two multi-channel RF head coils with uniform and trapezoidal shape elements were built and tested with a phantom at 7-T MRI scanner for comparison. The simulation and experimental results show stronger and more uniform B1+ near-field with the trapezoidal shape.

[1]  P. Silvester,et al.  TEM wave properties of microstrip transmission lines , 1968 .

[2]  L. Bolinger,et al.  Mapping of the Radiofrequency Field , 1993 .

[3]  Ravi S. Menon,et al.  Imaging at high magnetic fields: initial experiences at 4 T. , 1993, Magnetic resonance quarterly.

[4]  T. Ibrahim,et al.  Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator. , 2001, Magnetic resonance imaging.

[5]  R. Goebel,et al.  7T vs. 4T: RF power, homogeneity, and signal‐to‐noise comparison in head images , 2001, Magnetic resonance in medicine.

[6]  K. Uğurbil,et al.  Analysis of wave behavior in lossy dielectric samples at high field , 2002, Magnetic resonance in medicine.

[7]  A. Sbarbati,et al.  High field MRI in preclinical research. , 2003, European journal of radiology.

[8]  K. Uğurbil,et al.  Efficient high‐frequency body coil for high‐field MRI , 2004, Magnetic resonance in medicine.

[9]  P. Schmalbrock,et al.  Electromagnetic perspective on the operation of RF coils at 1.5–11.7 Tesla , 2005, Magnetic resonance in medicine.

[10]  Fritz Schick,et al.  Whole-body MRI at high field: technical limits and clinical potential , 2005, European Radiology.

[11]  Michael B. Smith,et al.  Central brightening due to constructive interference with, without, and despite dielectric resonance , 2005, Journal of magnetic resonance imaging : JMRI.

[12]  Peter Andersen,et al.  9.4T human MRI: Preliminary results , 2006, Magnetic resonance in medicine.

[13]  Kamil Ugurbil,et al.  Alternating impedance multi-channel transmission line resonators for high field magnetic resonance imaging , 2010, 2010 IEEE MTT-S International Microwave Symposium.

[14]  F. Mühlbauer Ultra-high-field magnetic resonance imaging , 2008 .

[15]  A. Omar,et al.  An investigation of alternating impedance microstrip transceiver coil arrays for MRI at 7T , 2011, 2011 IEEE MTT-S International Microwave Symposium.

[16]  A. Gopinath,et al.  Auto-tuning of the RF transmission line coil for high-fields magnetic resonance imaging (MRI) systems , 2011, 2011 IEEE MTT-S International Microwave Symposium.

[17]  J. Griffiths,et al.  RF Coils for MRI , 2012 .

[18]  Jeff H. Duyn,et al.  The future of ultra-high field MRI and fMRI for study of the human brain , 2012, NeuroImage.

[19]  A. Gopinath,et al.  RF multi-channel head coil design with improved B1+ Fields uniformity for high field MRI systems , 2012, 2012 IEEE/MTT-S International Microwave Symposium Digest.

[20]  A. Gopinath,et al.  8-Channel RF head coil of MRI with automatic tuning and matching , 2013, 2013 IEEE MTT-S International Microwave Symposium Digest (MTT).