Microfocus X-ray sources for 3D microtomography

Abstract An analytic model for the performance of cone beam microtomography is described. The maximum power of a microfocus X-ray source is assumed to be approximately proportional to the focal spot size. Radiation flux penetrating the specimen is predicted by a semi-empirical relation which is valid for X-ray energies less than 20 keV. Good signal to noise ratio is predicted for bone specimens of 0.1 to 10 mm when scanned at the optimal energy. A flux of about 1 × 10 10 photons/mm 2 /s is identified for 0.2 mm specimens. Cone beam volumetric microtomography is found to compare favorably with synchrotron based methods.

[1]  Stephen J. Riederer,et al.  Medical imaging [6] , 1995 .

[2]  S. Goldstein,et al.  Evaluation of a microcomputed tomography system to study trabecular bone structure , 1990, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[3]  S J Riederer,et al.  Noise Due to Photon Counting Statistics in Computed X‐Ray Tomography , 1977, Journal of computer assisted tomography.

[4]  Stephen Whitaker,et al.  X-Ray Anode Surface Temperatures: The Effect Of Volume Heating , 1988, Medical Imaging.

[5]  S. Goldstein,et al.  The direct examination of three‐dimensional bone architecture in vitro by computed tomography , 1989, Journal of bone and mineral research : the official journal of the American Society for Bone and Mineral Research.

[6]  R. H. Schneider Medical imaging II: 31 January–5 February 1988, Newport Beach CA-USA , 1989 .

[7]  H. Moseley,et al.  LXXX. The high-frequency spectra of the elements. Part II , 1914 .

[8]  W.-R. Dix,et al.  A system for dual energy microtomography of bones , 1989 .

[9]  H. Moseley,et al.  XCIII. The high-frequency spectra of the elements , 1913 .

[10]  S Webb,et al.  Three-dimensional x-ray microtomography for medical and biological applications. , 1990, Physics in medicine and biology.

[11]  B. Flannery,et al.  Three-Dimensional X-ray Microtomography , 1987, Science.

[12]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[13]  L. Grodzins,et al.  Optimum energies for x-ray transmission tomography of small samples. Applications of synchrotron radiation to computerized tomography I , 1983 .

[14]  M. C. Nichols,et al.  X‐ray microtomography on beamline X at SSRL , 1989 .

[15]  K. Göhler,et al.  Characteristic fluorescence correction for layered samples part I: Closed solution for film‐substrate combination , 1991 .