Ex vivo animal-model assessment of a non-invasive system for loss of resistance detection during epidural blockade

During recent decades epidural analgesia has gained widespread recognition in many applications. In this complex procedure, anaesthetist uses a specific needle to inject anesthetic into the epidural space. It is crucial the appropriate insertion of the needle through inhomogeneous tissues placed between the skin and the epidural space to minimize anesthetic-related complications (e.g., nausea, headache, and dural puncture). Usually, anaesthetists perform the procedure without any supporting tools, and stop pushing the syringe when they sense a loss of resistance (LOR). This phenomenon is caused by the physical properties of the epidural space: the needle breaks the ligamentum flavum and reaches the epidural space, in this stage the anaesthetist perceives a LOR because the epidural space is much softer than the ligamentum flavum. To support the clinician in this maneuver we designed a non-invasive system able to detect the LOR by measuring the pressure exerted on the syringe plunger to push the needle up to the epidural space. In a previous work we described the system and its assessment during in vitro tests. This work aims at assessing the feasibility of the system for LOR detection on a more realistic model (ex vivo pig model). The system was assessed by analyzing: its ability to hold a constant value (saturation condition) during the insertion of the needle, and its ability to detect the entrance within the epidural space by a decrease of the system's output. Lastly, the anaesthetist was asked to assess how the ex vivo procedure mimics a clinical scenario. The system reached the saturation condition during the needle insertion; this feature is critical to avoid false positive during the procedure. However, it was not easy to detect the entrance within the epidural space due to its small volume in the animal model. Lastly, the practitioner found real the model, and performed the procedures in a conventional manner because the system did not influence his actions.

[1]  E Schena,et al.  Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal airway: an in vivo study. , 2013, Minerva anestesiologica.

[2]  Edward T Riley,et al.  The Episure syringe: a novel loss of resistance syringe for locating the epidural space. , 2007, Anesthesia and analgesia.

[3]  Ronald Melzack,et al.  The myth of painless childbirth (The John J. Bonica Lecture) , 1984, Pain.

[4]  L. J. Teunissen,et al.  Clinical Results with the Acoustic Puncture Assist Device, a New Acoustic Device to Identify the Epidural Space , 2003, Anesthesia and analgesia.

[5]  Terrence K. Allen,et al.  A Pilot Study to Compare the EpisureTM AutodetectTM Syringe With the Glass Syringe for Identification of the Epidural Space in Parturients , 2008 .

[6]  E Schena,et al.  Bronchial blockers under pressure: in vitro model and ex vivo model. , 2016, British journal of anaesthesia.

[7]  Emiliano Schena,et al.  Flow measurement in mechanical ventilation: a review. , 2015, Medical engineering & physics.

[8]  John A. Cowan,et al.  Efficacy of postoperative epidural analgesia: a meta-analysis. , 2003, JAMA.

[9]  M. Seeberger,et al.  Epidural abscess complicating epidural anesthesia and analgesia: An analysis of the literature , 1998, Acta anaesthesiologica Scandinavica.

[10]  B. Covino,et al.  Local Anesthetics , 1987, Handbook of Experimental Pharmacology.

[11]  Emiliano Schena,et al.  Design and Feasibility Assessment of a Magnetic Resonance-Compatible Smart Textile Based on Fiber Bragg Grating Sensors for Respiratory Monitoring , 2016, IEEE Sensors Journal.

[12]  Timo J. Lechner,et al.  The Use of a Sound-enabled Device to Measure Pressure During Insertion of an Epidural Catheter in Women in Labor , 2012 .

[13]  Sergio Silvestri,et al.  Design and characterization of a measurement system for monitoring pressure exerted by bronchial blockers: In vitro trials , 2015, 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[14]  Emiliano Schena,et al.  A cost-effective, non-invasive system for pressure monitoring during epidural needle insertion: Design, development and bench tests , 2016, 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[15]  A. Habib,et al.  A Pilot Study to Compare the Episure™ Autodetect™ Syringe with the Glass Syringe for Identification of the Epidural Space in Parturients , 2008, Anesthesia and analgesia.

[16]  D. Formica,et al.  Smart Textile Based on Fiber Bragg Grating Sensors for Respiratory Monitoring: Design and Preliminary Trials , 2015, Biosensors.

[17]  Emiliano Schena,et al.  Medical Smart Textiles Based on Fiber Optic Technology: An Overview , 2015, Journal of functional biomaterials.

[18]  B M Hibbard,et al.  Serious non-fatal complications associated with extradural block in obstetric practice. , 1990, British journal of anaesthesia.

[19]  Catton Dv,et al.  Epidural analgesia for labor and delivery. , 1969 .

[20]  H. Wulf,et al.  Epidural anaesthesia and spinal haematoma , 1996, Canadian journal of anaesthesia = Journal canadien d'anesthesie.

[21]  T. Lechner,et al.  Clinical results with a new acoustic device to identify the epidural space , 2002, Anaesthesia.

[22]  A Wantman,et al.  Techniques for identifying the epidural space: a survey of practice amongst anaesthetists in the UK , 2006, Anaesthesia.