Rapid Deflation of the Bronchial Cuff of the Double-Lumen Tube After Decreasing the Concentration of Inspired Nitrous Oxide

Deflationary phenomena of the endotracheal tube cuff may occur after inspired nitrous oxide (N2O) concentrations are reduced, but deflationary phenomena of the double-lumen tube (DLT) cuff have not been investigated. In this study, tracheal and bronchial cuffs of left-sided Mallinckrodt (Athlone, Ireland) DLTs were inflated with air, 40% N2O, or 67% N2O (Air, N40, or N67 groups, respectively) in 24 patients undergoing thoracic surgery; 40 min later, O2 was substituted for N2O in some of the patients in the N40 group (N40-c group). Intracuff gas volumes, N2O concentrations, and cuff compliance were also measured. Both tracheal and bronchial cuff pressures significantly increased in the Air group but decreased in the N67 group. Neither pressure significantly changed in the N40 group, but both decreased in the N40-c group after terminating N2O anesthesia; the time required for bronchial cuff pressures to decrease by half (12.0 ± 5.5 min) was less than that for tracheal cuff pressures (31.2 ± 11.0 min, P < 0.01). The volume change in the N40-c group was not significantly different between the tracheal and bronchial cuffs, but tracheal cuff compliance was significantly higher than bronchial compliance. Therefore, filling DLT cuffs with 40% N2O stabilizes cuff pressure during anesthesia with 67% N2O, but bronchial cuffs deflate more quickly than tracheal cuffs after cessation of N2O administration through smaller compliance.

[1]  F. Karasawa,et al.  Repeated Deflation of a Gas-Barrier Cuff to Stabilize Cuff Pressure During Nitrous Oxide Anesthesia , 2002, Anesthesia and analgesia.

[2]  T. Satoh,et al.  An assessment of a method of inflating cuffs with a nitrous oxide gas mixture to prevent an increase in intracuff pressure in five different tracheal tube designs , 2001, Anaesthesia.

[3]  T. Satoh,et al.  Deflationary Phenomenon of the Nitrous Oxide-Filled Endotracheal Tube Cuff After Cessation of Nitrous Oxide Administration , 2001, Anesthesia and analgesia.

[4]  T. Satoh,et al.  The effect on intracuff pressure of various nitrous oxide concentrations used for inflating an endotracheal tube cuff. , 2000, Anesthesia and analgesia.

[5]  M. Gomes,et al.  A Comparison of the Reliability of Two Techniques of Left Double-Lumen Tube Bronchial Cuff Inflation in Producing Water-Tight Seal of the Left Mainstem Bronchus , 1998, Anesthesia and analgesia.

[6]  P. Silverman,et al.  Evaluation of an approach to choosing a left double-lumen tube size based on chest computed tomographic scan measurement of left mainstem bronchial diameter. , 1997, Journal of cardiothoracic and vascular anesthesia.

[7]  H. Uno,et al.  The Effect of Thoracic Epidural Anesthesia on Hypoxic Pulmonary Vasoconstriction in Dogs: An Analysis of the Pressure-Flow Curve , 1996, Anesthesia and analgesia.

[8]  G. Hedenstierna,et al.  Nitric Oxide Modulation of Pulmonary Blood Flow Distribution in Lobar Hypoxia , 1995, Anesthesiology.

[9]  U. Ruttimann,et al.  The relationship between left mainstem bronchial diameter and patient size. , 1995, Journal of cardiothoracic and vascular anesthesia.

[10]  J. Benumof,et al.  The Resting Volume and Compliance Characteristics of the Bronchial Cuff of Left Polyvinyl Chloride Double‐Lumen Endobronchial Tubes , 1993, Anesthesia and analgesia.

[11]  J. Benumof,et al.  Comparison of Three Techniques to Inflate the Bronchial Cuff of Left Polyvinylchloride Double‐Lumen Tubes , 1993, Anesthesia and analgesia.

[12]  P. Slinger,et al.  A Clinical Comparison of Bronchial Cuff Pressures in Three Different Designs of Left Double‐Lumen Tubes , 1993, Anesthesia and Analgesia.

[13]  R. Vaughan,et al.  Endobronchial cuff pressures. , 1993, British Journal of Anaesthesia.

[14]  G. Jacobsen,et al.  The utility of a double-lumen tube for one-lung ventilation in a variety of noncardiac thoracic surgical procedures. , 1992, Journal of cardiothoracic and vascular anesthesia.

[15]  D. Gaba,et al.  Bronchial Cuff Pressures of Double‐Lumen Tubes , 1989, Anesthesia and analgesia.

[16]  M. Hannallah,et al.  Bronchial rupture associated with the use of a double-lumen tube in a small adult. , 1989, Anesthesiology.

[17]  N. Burton,et al.  Rupture of the left main-stem bronchus with a polyvinylchloride double-lumen tube. , 1983, Chest.

[18]  T. Hornbein,et al.  Failure of a high-compliance low-pressure cuff to prevent aspiration. , 1975, Anesthesiology.

[19]  T. Stanley,et al.  Effects of nitrous oxide on volume and pressure of endotracheal tube cuffs. , 1974, Anesthesiology.