Performance Evaluation of a Multichannel All-In-One Phantom Dosimeter for Dose Measurement of Diagnostic X-ray Beam

We developed a multichannel all-in-one phantom dosimeter system composed of nine sensing probes, a chest phantom, an image intensifier, and a complementary metal-oxide semiconductor (CMOS) image sensor to measure the dose distribution of an X-ray beam used in radiation diagnosis. Nine sensing probes of the phantom dosimeter were fabricated identically by connecting a plastic scintillating fiber (PSF) to a plastic optical fiber (POF). To measure the planar dose distribution on a chest phantom according to exposure parameters used in clinical practice, we divided the top of the chest phantom into nine equal parts virtually and then installed the nine sensing probes at each center of the nine equal parts on the top of the chest phantom as measuring points. Each scintillation signal generated in the nine sensing probes was transmitted through the POFs and then intensified by the image intensifier because the scintillation signal normally has a very low light intensity. Real-time scintillation images (RSIs) containing the intensified scintillation signals were taken by the CMOS image sensor with a single lens optical system and displayed through a software program. Under variation of the exposure parameters, we measured RSIs containing dose information using the multichannel all-in-one phantom dosimeter and compared the results with the absorbed doses obtained by using a semiconductor dosimeter (SCD). From the experimental results of this study, the light intensities of nine regions of interest (ROI) in the RSI measured by the phantom dosimeter were similar to the dose distribution obtained using the SCD. In conclusion, we demonstrated that the planar dose distribution including the entrance surface dose (ESD) can be easily measured by using the proposed phantom dosimeter system.

[1]  J Zoetelief,et al.  Survey of posteroanterior chest radiography in The Netherlands: patient dose and image quality. , 2003, The British journal of radiology.

[2]  Amber J Gislason-Lee,et al.  Dose optimization in cardiac x-ray imaging. , 2013, Medical physics.

[3]  S. Shin,et al.  Development of a fiber-optic dosimeter based on modified direct measurement for real-time dosimetry during radiation diagnosis , 2013 .

[4]  Wook Jae Yoo,et al.  Development of a scintillating fiber-optic dosimeter for measuring the entrance surface dose in diagnostic radiology , 2013 .

[5]  A. Jones,et al.  Potential clinical utility of a fibre optic-coupled dosemeter for dose measurements in diagnostic radiology. , 2008, Radiation protection dosimetry.

[6]  Filippo Attivissimo,et al.  Dose Optimization in Chest Radiography: System and Model Characterization via Experimental Investigation , 2014, IEEE Transactions on Instrumentation and Measurement.

[7]  D. Hyer,et al.  Characterization of a water-equivalent fiber-optic coupled dosimeter for use in diagnostic radiology. , 2009, Medical physics.

[8]  Mitchell M. Goodsitt,et al.  Standardized Methods for Measuring Diagnostic X-Ray Exposures , 1990 .

[9]  A. Huston,et al.  Characterization of a fiber-optic-coupled radioluminescent detector for application in the mammography energy range. , 2007, Medical physics.

[10]  T. R. Fewell,et al.  Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinations. , 1984, Medical physics.

[11]  I. Tsalafoutas,et al.  Radiation doses to patients undergoing standard radiographic examinations: a comparison between two methods. , 2007, The British journal of radiology.

[12]  Wook Jae Yoo,et al.  Measurement of Entrance Surface Dose on an Anthropomorphic Thorax Phantom Using a Miniature Fiber-Optic Dosimeter , 2014, Sensors.

[13]  G. Ardran,et al.  Checking diagnostic x-ray beam quality. , 1968, The British journal of radiology.