Magnetic confinement of electron and photon radiotherapy dose: a Monte Carlo simulation with a nonuniform longitudinal magnetic field.

It recently has been shown experimentally that the focusing provided by a longitudinal nonuniform high magnetic field can significantly improve electron beam dose profiles. This could permit precise targeting of tumors near critical areas and minimize the radiation dose to surrounding healthy tissue. The experimental results together with Monte Carlo simulations suggest that the magnetic confinement of electron radiotherapy beams may provide an alternative to proton or heavy ion radiation therapy in some cases. In the present work, the external magnetic field capability of the Monte Carlo code PENELOPE was utilized by providing a subroutine that modeled the actual field produced by the solenoid magnet used in the experimental studies. The magnetic field in our simulation covered the region from the vacuum exit window to the phantom including surrounding air. In a longitudinal nonuniform magnetic field, it is observed that the electron dose can be focused in both the transverse and longitudinal directions. The measured dose profiles of the electron beam are generally reproduced in the Monte Carlo simulations to within a few percent in the region of interest provided that the geometry and the energy of the incident electron beam are accurately known. Comparisons for the photon beam dose profiles with and without the magnetic field are also made. The experimental results are qualitatively reproduced in the simulation. Our simulation shows that the excessive dose at the beam entrance is due to the magnetic field trapping and focusing scattered secondary electrons that were produced in the air by the incident photon beam. The simulations also show that the electron dose profile can be manipulated by the appropriate control of the beam energy together with the strength and displacement of the longitudinal magnetic field.

[1]  D. Jette Magnetic fields with photon beams: Monte Carlo calculations for a model magnetic field. , 2000, Medical physics.

[2]  Lijun Ma Dosimetric properties of magnetically collimated electron beams for radiation therapy. , 2004, Medical Physics (Lancaster).

[3]  J A Purdy,et al.  Magnetic enhancement of electron dose distribution in a phantom. , 1977, Medical physics.

[4]  D. Whitmire,et al.  Magnetic modification of the electron-dose distribution in tissue and lung phantoms. , 1978, Medical physics.

[5]  S. Naqvi,et al.  Control of photon beam dose profiles by localized transverse magnetic fields. , 2000, Physics in medicine and biology.

[6]  R. Stern,et al.  Tests of a large air‐core superconducting solenoid as a nuclear‐reaction‐product spectrometer , 1987 .

[7]  B. Paliwal,et al.  Magnetic field modification of electron-beam dose distributions in inhomogeneous media. , 1978, Medical physics.

[8]  D. McShan,et al.  MAGNETIC CONFINEMENT OF RADIOTHERAPY BEAM-DOSE PROFILES , 2001 .

[9]  J. Baró,et al.  PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter , 1995 .

[10]  A F Bielajew,et al.  The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beams. , 1993, Medical physics.

[11]  R Nath,et al.  Response of mammalian cells irradiated with 30 MV X-rays in the presence of a uniform 20-kilogauss magnetic field. , 1980, International journal of radiation biology and related studies in physics, chemistry, and medicine.

[12]  C C Shih,et al.  High energy electron radiotherapy in a magnetic field. , 1975, Medical physics.

[13]  S. Rockwell,et al.  Influence of a 1400-gauss magnetic field on the radiosensitivity and recovery of EMT6 cells in vitro. , 1977, International journal of radiation biology and related studies in physics, chemistry, and medicine.

[14]  Wenzhong Liu,et al.  Production and Use of Radioactive Ion Beams for Measurements of Nuclear Reactions. , 1990 .

[15]  R. Nath,et al.  Modification of electron-beam dose distributions by transverse magnetic fields. , 1978, Medical physics.

[16]  J. Baró,et al.  An algorithm for Monte Carlo simulation of coupled electron-photon transport , 1997 .

[17]  W. H. Bostick Possible Techniques in Direct-Electron-Beam Tumor The apy , 1950 .

[18]  Indrin J Chetty,et al.  Experimental validation of the DPM Monte Carlo code using minimally scattered electron beams in heterogeneous media. , 2002, Physics in medicine and biology.

[19]  S. Naqvi,et al.  Evaluating the effectiveness of a longitudinal magnetic field in reducing underdosing of the regions around upper respiratory cavities irradiated with photon beams--a Monte Carlo study. , 2001, Medical physics.

[20]  J. Sempau,et al.  Monte Carlo simulation of electron beams from an accelerator head using PENELOPE , 2001, Physics in Medicine and Biology.

[21]  R Nath,et al.  Enhancement of electron beam dose distributions by longitudinal magnetic fields: Monte Carlo simulations and magnet system optimization. , 1985, Medical physics.

[22]  D. J. Strom,et al.  Microdosimetric properties of ionizing electrons in water: a test of the PENELOPE code system. , 2002, Physics in medicine and biology.

[23]  X. Li,et al.  Reducing loss in lateral charged-particle equilibrium due to air cavities present in x-ray irradiated media by using longitudinal magnetic fields. , 2001, Medical physics.

[24]  D. Sheikh-Bagheri,et al.  Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters. , 2002, Medical physics.