Loosening of the femoral component after total hip replacement. The thin black line and the sinking hip.
暂无分享,去创建一个
Review of the radiographs of 882 consecutive metal-to-polyethylene hip replacements after a mean follow-up of 6.8 years showed 72 cases (8%) of aseptic femoral loosening, all of which showed a lucent zone or black line between metal and cement. In 97% of the series this appeared within two years of the arthroplasty. Progress of loosening was unpredictable and only 24% had significant symptoms. Improved technique with methylmethacrylate cement in Stanmore prostheses had reduced loosening after two years from 7.9% to 0.4%. Methods of detection of early loosening are discussed.