4D ultrasound speckle tracking of intra-fraction prostate motion: a phantom-based comparison with x-ray fiducial tracking using CyberKnife

This study investigates the use of a mechanically-swept 3D ultrasound (3D-US) probe for soft-tissue displacement monitoring during prostate irradiation, with emphasis on quantifying the accuracy relative to CyberKnife® x-ray fiducial tracking. An US phantom, implanted with x-ray fiducial markers was placed on a motion platform and translated in 3D using five real prostate motion traces acquired using the Calypso system. Motion traces were representative of all types of motion as classified by studying Calypso data for 22 patients. The phantom was imaged using a 3D swept linear-array probe (to mimic trans-perineal imaging) and, subsequently, the kV x-ray imaging system on CyberKnife. A 3D cross-correlation block-matching algorithm was used to track speckle in the ultrasound data. Fiducial and US data were each compared with known phantom displacement. Trans-perineal 3D-US imaging could track superior-inferior (SI) and anterior-posterior (AP) motion to ≤0.81 mm root-mean-square error (RMSE) at a 1.7 Hz volume rate. The maximum kV x-ray tracking RMSE was 0.74 mm, however the prostate motion was sampled at a significantly lower imaging rate (mean: 0.04 Hz). Initial elevational (right-left; RL) US displacement estimates showed reduced accuracy but could be improved (RMSE <2.0 mm) using a correlation threshold in the ultrasound tracking code to remove erroneous inter-volume displacement estimates. Mechanically-swept 3D-US can track the major components of intra-fraction prostate motion accurately but exhibits some limitations. The largest US RMSE was for elevational (RL) motion. For the AP and SI axes, accuracy was sub-millimetre. It may be feasible to track prostate motion in 2D only. 3D-US also has the potential to improve high tracking accuracy for all motion types. It would be advisable to use US in conjunction with a small (∼2.0 mm) centre-of-mass displacement threshold in which case it would be possible to take full advantage of the accuracy and high imaging rate capability.

[1]  S L Schoeppel,et al.  Treatment planning issues related to prostate movement in response to differential filling of the rectum and bladder. , 1991, International journal of radiation oncology, biology, physics.

[2]  David J Brenner,et al.  What hypofractionated protocols should be tested for prostate cancer? , 2002, International journal of radiation oncology, biology, physics.

[3]  J. Wong,et al.  Flat-panel cone-beam computed tomography for image-guided radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[4]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[5]  Alexander Markovic,et al.  Quality assurance of U.S.-guided external beam radiotherapy for prostate cancer: report of AAPM Task Group 154. , 2011, Medical physics.

[6]  T. Peters,et al.  Poly(vinyl alcohol) cryogel phantoms for use in ultrasound and MR imaging , 2004, Physics in medicine and biology.

[7]  A. Hanlon,et al.  Ultrasound-based stereotactic guidance of precision conformal external beam radiation therapy in clinically localized prostate cancer. , 2000, Urology.

[8]  Gig S Mageras,et al.  Planning in the IGRT context: closing the loop. , 2007, Seminars in radiation oncology.

[9]  M. Terris,et al.  Comparison of ultrasound imaging in patients undergoing transperineal and transrectal prostate ultrasound. , 1998, Urology.

[10]  Christos Antypas,et al.  Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system , 2008, Physics in medicine and biology.

[11]  M. Lachaine,et al.  INTRAFRACTIONAL PROSTATE MOTION MANAGEMENT WITH THE CLARITY AUTOSCAN SYSTEM , 2013 .

[12]  James M. Balter,et al.  The Influence of Intrafraction Movement on Margins for Prostate Radiotherapy , 2005 .

[13]  Graham M. Treece,et al.  Processing and visualizing three-dimensional ultrasound data. , 2004, British Journal of Radiology.

[14]  S. Korreman Motion in radiotherapy: photon therapy , 2012, Physics in medicine and biology.

[15]  Rick Chappell,et al.  Is α/β for prostate tumors really low? , 2001 .

[16]  D. Jaffray,et al.  Image-guided radiotherapy: has it influenced patient outcomes? , 2012, Seminars in radiation oncology.

[17]  W. Lee,et al.  Hypofractionation for prostate cancer: a critical review. , 2008, Seminars in radiation oncology.

[18]  K. Shinohara,et al.  Technique for implantation of fiducial markers in the prostate. , 2008, Urology.

[19]  O T Von Ramm,et al.  3D Ultrasound Tissue Motion Tracking Using Correlation Search , 1998, Ultrasonic imaging.

[20]  Patrick A Kupelian,et al.  Influence of intrafraction motion on margins for prostate radiotherapy. , 2006, International journal of radiation oncology, biology, physics.

[21]  John T. Wei,et al.  Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. , 2006, International journal of radiation oncology, biology, physics.

[22]  T. Kron,et al.  Rectal filling at planning does not predict stability of the prostate gland during a course of radical radiotherapy if patients with large rectal filling are re-imaged. , 2009, Clinical oncology (Royal College of Radiologists (Great Britain)).

[23]  Brett Byram,et al.  3-D phantom and in vivo cardiac speckle tracking using a matrix array and raw echo data , 2010, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[24]  A.A. Morsy,et al.  FLASH correlation: a new method for 3-D ultrasound tissue motion tracking and blood velocity estimation , 1999, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[25]  Mary Feng,et al.  Potential Use of Ultrasound Speckle Tracking for Motion Management During Radiotherapy , 2012, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[26]  Avid,et al.  INTRAFRACTIONAL MOTION OF THE PROSTATE DURING HYPOFRACTIONATED RADIOTHERAPY , 2009 .

[27]  P. Evans,et al.  In vivo liver tracking with a high volume rate 4D ultrasound scanner and a 2D matrix array probe , 2012, Physics in medicine and biology.

[28]  J. Adler,et al.  CyberKnife Radiotherapy for Localized Prostate Cancer: Rationale and Technical Feasibility , 2003, Technology in cancer research & treatment.

[29]  C. Maurer,et al.  The CyberKnife® Robotic Radiosurgery System in 2010 , 2010, Technology in cancer research & treatment.

[30]  Steve B. Jiang,et al.  The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. , 2007, Medical physics.

[31]  George Starkschall,et al.  Late rectal toxicity: dose-volume effects of conformal radiotherapy for prostate cancer. , 2002, International journal of radiation oncology, biology, physics.

[32]  Dimitre Hristov,et al.  Telerobotic system concept for real-time soft-tissue imaging during radiotherapy beam delivery. , 2010, Medical physics.

[33]  D P Dearnaley,et al.  Evaluating the effect of rectal distension and rectal movement on prostate gland position using cine MRI. , 1999, International journal of radiation oncology, biology, physics.

[34]  Jean-François Aubry,et al.  Measurements of intrafraction motion and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers. , 2004, International journal of radiation oncology, biology, physics.

[35]  C. Pelizzari,et al.  Evaluation of changes in the size and location of the prostate, seminal vesicles, bladder, and rectum during a course of external beam radiation therapy. , 1995, International journal of radiation oncology, biology, physics.

[36]  M. V. van Herk,et al.  The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[37]  S. Webb Motion effects in (intensity modulated) radiation therapy: a review , 2006, Physics in medicine and biology.

[38]  M. Ritter,et al.  The α/β ratio for prostate cancer: What is it, really? , 2005 .

[39]  S. Webb,et al.  Feasibility of using ultrasound for real-time tracking during radiotherapy. , 2005, Medical physics.

[40]  John Wong,et al.  Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients. , 2004, International journal of radiation oncology, biology, physics.

[41]  Jeffrey C Bamber,et al.  Performance of ultrasound based measurement of 3D displacement using a curvilinear probe for organ motion tracking. , 2007, Physics in medicine and biology.

[42]  Sanyam Mehta,et al.  A high-performance parallel implementation of Sum of Absolute Differences algorithm for motion estimation using CUDA , 2010 .