First experimental-based characterization of oxygen ion beam depth dose distributions at the Heidelberg Ion-Beam Therapy Center

Over the last decades, the application of proton and heavy-ion beams to external beam radiotherapy has rapidly increased. Due to the favourable lateral and depth dose profile, the superposition of narrow ion pencil beams may enable a highly conformal dose delivery to the tumour, with better sparing of the surrounding healthy tissue in comparison to conventional radiation therapy with photons. To fully exploit the promised clinical advantages of ion beams, an accurate planning of the patient treatments is required. The clinical treatment planning system (TPS) at the Heidelberg Ion-Beam Therapy Center (HIT) is based on a fast performing analytical algorithm for dose calculation, relying, among others, on laterally integrated depth dose distributions (DDDs) simulated with the FLUKA Monte Carlo (MC) code. Important input parameters of these simulations need to be derived from a comparison of the simulated DDDs with measurements. In this work, the first measurements of (16)O ion DDDs at HIT are presented with a focus on the determined Bragg peak positions and the understanding of factors influencing the shape of the distributions. The measurements are compared to different simulation approaches aiming to reproduce the acquired data at best. A simplified geometrical model is first used to optimize important input parameters, not known a priori, in the simulations. This method is then compared to a more realistic, but also more time-consuming simulation approach better accounting for the experimental set-up and the measuring process. The results of this work contributed to a pre-clinical oxygen ion beam database, which is currently used by a research TPS for corresponding radio-biological cell experiments. A future extension to a clinical database used by the clinical TPS at HIT is foreseen. As a side effect, the performed investigations showed that the typical water equivalent calibration approach of experimental data acquired with water column systems leads to slight deviations between the experimentally determined and the real Bragg peak positions. For improved accuracy, the energy dependence of the stopping power, and herewith the water equivalent thickness, of the material downstream of the water tank should be considered in the analysis of measured data.

[1]  K Parodi,et al.  Monte Carlo simulations to support start-up and treatment planning of scanned proton and carbon ion therapy at a synchrotron-based facility , 2012, Physics in medicine and biology.

[2]  Takashi Akagi,et al.  Determination of the mean excitation energy of water from proton beam ranges , 2007 .

[3]  O Jäkel,et al.  Treatment planning for heavy-ion radiotherapy: physical beam model and dose optimization. , 2000, Physics in medicine and biology.

[4]  H. Bichsel,et al.  Aspects of Fast-Ion Dosimetry , 2000, Radiation research.

[5]  A Mairani,et al.  Benchmarking nuclear models of FLUKA and GEANT4 for carbon ion therapy , 2010, Physics in medicine and biology.

[6]  A. Ferrari,et al.  FLUKA: A Multi-Particle Transport Code , 2005 .

[7]  J Fieres,et al.  Experimental characterization of lateral profiles of scanned proton and carbon ion pencil beams for improved beam models in ion therapy treatment planning , 2011, Physics in medicine and biology.

[8]  Katia Parodi,et al.  Heidelberg Ion Therapy Center (HIT): Initial clinical experience in the first 80 patients , 2010, Acta oncologica.

[9]  O Jäkel,et al.  The Heidelberg Ion Therapy Center. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[10]  M Scholz,et al.  RBE for carbon track-segment irradiation in cell lines of differing repair capacity. , 1999, International journal of radiation biology.

[11]  F. Cerutti,et al.  The FLUKA code: Description and benchmarking , 2007 .

[12]  D. Schardt,et al.  Magnetic scanning system for heavy ion therapy , 1993 .

[13]  Katia Parodi,et al.  Investigating the accuracy of the FLUKA code for transport of therapeutic ion beams in matter , 2006, Physics in medicine and biology.

[14]  M Scholz,et al.  The FLUKA Monte Carlo code coupled with the local effect model for biological calculations in carbon ion therapy , 2010, Physics in medicine and biology.

[15]  Harald Paganetti,et al.  Relative biological effectiveness (RBE) values for proton beam therapy. , 2002, International journal of radiation oncology, biology, physics.

[16]  G Kraft,et al.  Design and construction of a ripple filter for a smoothed depth dose distribution in conformal particle therapy. , 1999, Physics in medicine and biology.

[17]  E. Rietzel,et al.  Experimental study of the water-to-air stopping power ratio of monoenergetic carbon ion beams for particle therapy , 2012, Physics in medicine and biology.

[18]  J. Deasy ICRU Report 49, Stopping Powers and Ranges for Protons and Alph Particles , 1994 .

[19]  R. Wilson Radiological use of fast protons. , 1946, Radiology.

[20]  D. Schardt,et al.  Nuclear fragmentation of high-energy heavy-ion beams in water. , 1996, Advances in space research : the official journal of the Committee on Space Research.