Comparison of MPEG digital video with super VHS tape for diagnostic echocardiographic readings.

BACKGROUND Digital recording of echocardiographic studies is on the clinical horizon. However, full digital capture of complete echocardiographic studies in traditional video format is impractical, given current storage capacity and network bandwidth. To overcome these constraints, we evaluated the diagnostic image quality of digital video by using MPEG (Motion Picture Experts Group) compression. METHODS AND RESULTS Fifty-eight complete, consecutive studies were recorded simultaneously with the use of MPEG-1 and sVHS videotape. Each matched MPEG and sVHS study pair was reviewed by two from a total of six readers, and findings were recorded with the use of a detailed, computerized reporting tool. Intrareader and interreader discrepancies were characterized as major or minor and analyzed in total and for specific subgroups of findings (left and right ventricular parameters, valvular insufficiency, and left ventricular regional wall motion). Intrareader discrepancies were reviewed by a consensus panel for agreement with either MPEG or sVHS findings. There was an exact concordance between MPEG and sVHS readings in 83% of findings. The majority of discrepancies were minor, with major discrepancies in only 2.7% of findings. There was no difference in the rate of consensus panel agreement with MPEG or sVHS for instances of intrareader discrepancy, either in total or for any subgroup of findings. Interreader discrepancy rates were nearly identical for both MPEG and sVHS. CONCLUSIONS MPEG-1 digital video is equivalent to sVHS videotape for diagnostic echocardiography. MPEG increases the range of practical options for digital echocardiography and offers, for the first time, the advantages of digital recording in a familiar video format.

[1]  Joan L. Mitchell,et al.  MPEG Video: Compression Standard , 1996 .

[2]  Harvey Feigenbaum,et al.  Practical Approach to a Digital Echocardiographic Laboratory , 1997, Echocardiography.

[3]  J D Thomas,et al.  Digital formatting standards in medical imaging: a primer for echocardiographers. , 1994, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[4]  J D Thomas,et al.  Digital storage of echocardiograms offers superior image quality to analog storage, even with 20:1 digital compression: results of the Digital Echo Record Access Study. , 1996, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[5]  J B Seward,et al.  Is review of videotape necessary after review of digitized cine-loop images in stress echocardiography? A prospective study in 306 patients. , 1997, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[6]  Joseph P. Murgo,et al.  Is a Full Digital Echocardiography Laboratory Feasible for Routine Daily Use? , 1996, Echocardiography.

[7]  D S Segar,et al.  Comparison of digital with videotape echocardiography in patients with chest pain in the emergency department. , 1996, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[8]  J D Thomas,et al.  Digital compression of echocardiograms: impact on quantitative interpretation of color Doppler velocity. , 1996, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[9]  J D Thomas,et al.  JPEG compression of digital echocardiographic images: impact on image quality. , 1995, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[10]  N. Reichek,et al.  Recommendations for quantitation of the left ventricle by two-dimensional echocardiography. American Society of Echocardiography Committee on Standards, Subcommittee on Quantitation of Two-Dimensional Echocardiograms. , 1989, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.