Dosimetric characterization of the iBEAM evo carbon fiber couch for radiotherapy.

PURPOSE This study characterizes the dosimetric properties of the iBEAM evo carbon fiber couch manufactured by Medical Intelligence and examines the accuracy of the CMS XiO and Nucletron Oncentra Masterplan (OMP) treatment planning systems for calculating beam attenuation due to the presence of the couch. METHODS To assess the homogeneity of the couch, it was CT scanned at isocentric height and a number of signal intensity profiles were generated and analyzed. To simplify experimental procedures, surface dose and central axis depth dose measurements were performed in a solid water slab phantom using Gafchromic film for 6 and 10 MV photon beams at gantry angles of 0 degree (normal incidence), 30 degrees, and 60 degrees with an inverted iBEAM couch placed on top of the phantom. Attenuation measurements were performed in a cylindrical solid water phantom with an ionization chamber positioned at the isocenter. Measurements were taken for gantry angles from 0 degree to 90 degrees in 10 degrees increments for both 6 and 10 MV photon beams. This setup was replicated in the XiO and OMP treatment planning systems. Dose was calculated using the pencil beam, collapsed cone, convolution, and superposition algorithms. RESULTS The CT scan of the couch showed that it was uniformly constructed. Surface dose increased by (510 +/-0)% for a 6 MV beam and (600 +/- 20)% for a 10 MV beam passing through the couch at normal incidence. Obliquely incident beams resulted in a higher surface dose compared to normally incident beams for both open fields and fields with the couch present. Depth dose curves showed that the presence of the couch resulted in an increase in dose in the build up region. For 6 and 10 MV beams incident at 60 degrees, nearly all skin sparing was lost. Attenuation measurements derived using the ionization chamber varied from 2.7% (0 degree) to a maximum of 4.6% (50 degrees) for a 6 MV beam and from 1.9% (0 degree) to a maximum of 4.0% (50 degrees) for a 10 MV beam. The pencil beam and convolution algorithms failed to accurately calculate couch attenuation. The collapsed cone and superposition algorithms calculated attenuation within an absolute error of +/- 1.2% for 6 MV and +/- 0.8% for 10 MV for gantry angles from 0 degree to 40 degrees. Some differences in attenuation were observed dependent on how the couch was contoured. CONCLUSIONS These results demonstrate that the presence of the iBEAM evo carbon fiber couch increases the surface dose and dose in the build up region. The inclusion of the couch in the planning scan is limited by the field of view employed and the couch height at the time of CT scanning.

[1]  Luca Cozzi,et al.  The impact of treatment couch modelling on RapidArc , 2009, Physics in medicine and biology.

[2]  David Wilkins,et al.  Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop , 2006, Journal of applied clinical medical physics.

[3]  Martin J Butso,et al.  Variations in skin dose associated with linac bed material at 6 MV x-ray energy. , 2002, Physics in medicine and biology.

[4]  A G Glendinning,et al.  Dosimetric characteristics of the Elekta Beam Modulator™ , 2005, Physics in medicine and biology.

[5]  Luca Cozzi,et al.  Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations , 2006, Physics in medicine and biology.

[6]  B. Schaeken,et al.  The effect of carbon fibre inserts on the build-up and attenuation of high energy photon beams. , 1997, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[7]  Emiliano Spezi,et al.  Evaluating the influence of the Siemens IGRT carbon fibre tabletop in head and neck IMRT. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  L. Mooy The use of carbon fibres in radiotherapy. , 1991 .

[9]  F. Khan,et al.  Plane-parallel ionization chamber response in the buildup region of obliquely incident photon beams. , 1997, Medical physics.

[10]  J. Galvin,et al.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. , 1998, Medical physics.

[11]  K. Langmack,et al.  An investigation into the use of carbon fibre for megavoltage radiotherapy applications. , 1998, Physics in medicine and biology.

[12]  Emiliano Spezi,et al.  Dosimetric characteristics of the Siemens IGRT carbon fiber tabletop. , 2007, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[13]  J. Seuntjens,et al.  Accurate skin dose measurements using radiochromic film in clinical applications. , 2006, Medical physics.

[14]  A. Ahnesjö Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media. , 1989, Medical physics.

[15]  Barbara Dobler,et al.  Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX—a feasibility study , 2010, Physics in medicine and biology.

[16]  L A DeWerd,et al.  An electron density calibration phantom for CT-based treatment planning computers. , 1992, Medical physics.

[17]  M Niedbala,et al.  Dose perturbations by two carbon fiber treatment couches and the ability of a commercial treatment planning system to predict these effects. , 2009, Medical physics.

[18]  E. Moros,et al.  Modeling of carbon fiber couch attenuation properties with a commercial treatment planning system. , 2008, Medical physics.

[19]  Leen Paelinck,et al.  Evaluation of the Sinmed Mastercouch as replacement for a standard couch. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[20]  Armand Djouguela,et al.  The Effect of a Carbon-Fiber Couch on the Depth-Dose Curves and Transmission Properties for Megavoltage Photon Beams , 2007, Strahlentherapie und Onkologie.

[21]  M. Butson,et al.  Megavoltage x-ray skin dose variation with an angle using grid carbon fibre couch tops , 2007, Physics in medicine and biology.

[22]  The effect of carbon fiber couch inserts on surface dose with beam size variation. , 2001, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[23]  T. Knöös,et al.  The dosimetric verification of a pencil beam based treatment planning system. , 1994, Physics in medicine and biology.

[24]  M Miften,et al.  Implementation of FFT convolution and multigrid superposition models in the FOCUS RTP system. , 2000, Physics in medicine and biology.

[25]  Jennifer Diffey,et al.  The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. , 2005, Medical physics.

[26]  S C Lillicrap,et al.  Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service , 1990 .