Simultaneous Multislice Echo Planar Imaging With Blipped Controlled Aliasing in Parallel Imaging Results in Higher Acceleration: A Promising Technique for Accelerated Diffusion Tensor Imaging of Skeletal Muscle

ObjectiveThe aim of this study was to investigate the feasibility of accelerated diffusion tensor imaging (DTI) of skeletal muscle using echo planar imaging (EPI) applying simultaneous multislice excitation with a blipped controlled aliasing in parallel imaging results in higher acceleration unaliasing technique. Materials and MethodsAfter federal ethics board approval, the lower leg muscles of 8 healthy volunteers (mean [SD] age, 29.4 [2.9] years) were examined in a clinical 3-T magnetic resonance scanner using a 15-channel knee coil. The EPI was performed at a b value of 500 s/mm2 without slice acceleration (conventional DTI) as well as with 2-fold and 3-fold acceleration. Fractional anisotropy (FA) and mean diffusivity (MD) were measured in all 3 acquisitions. Fiber tracking performance was compared between the acquisitions regarding the number of tracks, average track length, and anatomical precision using multivariate analysis of variance and Mann-Whitney U tests. ResultsAcquisition time was 7:24 minutes for conventional DTI, 3:53 minutes for 2-fold acceleration, and 2:38 minutes for 3-fold acceleration. Overall FA and MD values ranged from 0.220 to 0.378 and 1.595 to 1.829 mm2/s, respectively. Two-fold acceleration yielded similar FA and MD values (P ≥ 0.901) and similar fiber tracking performance compared with conventional DTI. Three-fold acceleration resulted in comparable MD (P = 0.199) but higher FA values (P = 0.006) and significantly impaired fiber tracking in the soleus and tibialis anterior muscles (number of tracks, P < 0.001; anatomical precision, P ⩽ 0.005). ConclusionsSimultaneous multislice EPI with blipped controlled aliasing in parallel imaging results in higher acceleration can remarkably reduce acquisition time in DTI of skeletal muscle with similar image quality and quantification accuracy of diffusion parameters. This may increase the clinical applicability of muscle anisotropy measurements.

[1]  L. Axel,et al.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. , 1990, Medical physics.

[2]  Klaus Hahn,et al.  Fast low‐angle shot diffusion tensor imaging with stimulated echo encoding in the muscle of rabbit shank , 2014, NMR in biomedicine.

[3]  Sharmila Majumdar,et al.  Considerations in high‐resolution skeletal muscle diffusion tensor imaging using single‐shot echo planar imaging with stimulated‐echo preparation and sensitivity encoding , 2012, NMR in biomedicine.

[4]  N. Paragios,et al.  Assessment of calf muscle contraction by diffusion tensor imaging , 2008, European Radiology.

[5]  J. R. Landis,et al.  The measurement of observer agreement for categorical data. , 1977, Biometrics.

[6]  Thorsten Feiweier,et al.  Comparison of twice refocused spin echo versus stimulated echo diffusion tensor imaging for tracking muscle fibers , 2015, Journal of magnetic resonance imaging : JMRI.

[7]  Zhaohua Ding,et al.  Quantitative diffusion tensor MRI-based fiber tracking of human skeletal muscle. , 2007, Journal of applied physiology.

[8]  L. White,et al.  Diffusion tensor imaging and fiber tractography of skeletal muscle: optimization of B value for imaging at 1.5 T. , 2009, AJR. American journal of roentgenology.

[9]  A. Cotten,et al.  Diffusion tensor imaging in musculoskeletal disorders. , 2014, Radiographics : a review publication of the Radiological Society of North America, Inc.

[10]  F. Carletti,et al.  Feasibility of Diffusion Tensor Tractography of Brachial Plexus Injuries at 1.5 T , 2013, Investigative radiology.

[11]  Damien Rohmer,et al.  Reconstruction and Visualization of Fiber and Laminar Structure in the Normal Human Heart from Ex Vivo Diffusion Tensor Magnetic Resonance Imaging (DTMRI) Data , 2007, Investigative radiology.

[12]  Zhaohua Ding,et al.  Quantitative assessment of DTI‐based muscle fiber tracking and optimal tracking parameters , 2009, Magnetic resonance in medicine.

[13]  Derek K. Jones,et al.  The effect of gradient sampling schemes on measures derived from diffusion tensor MRI: A Monte Carlo study † , 2004, Magnetic resonance in medicine.

[14]  Kawin Setsompop,et al.  Improving the spatial resolution of magnetic resonance inverse imaging via the blipped-CAIPI acquisition scheme , 2014, NeuroImage.

[15]  M. Moseley,et al.  In vivo MR tractography using diffusion imaging. , 2003, European journal of radiology.

[16]  M. Ladd,et al.  Diffusive sensitivity to muscle architecture: a magnetic resonance diffusion tensor imaging study of the human calf , 2004, European Journal of Applied Physiology.

[17]  C. Liang,et al.  Grading of rabbit skeletal muscle trauma by diffusion tensor imaging and tractography on magnetic resonance imaging. , 2006, Chinese medical sciences journal = Chung-kuo i hsueh k'o hsueh tsa chih.

[18]  M. Bronskill,et al.  T1, T2 relaxation and magnetization transfer in tissue at 3T , 2005, Magnetic resonance in medicine.

[19]  C. Hancart,et al.  Diffusion tensor imaging and tractography of the median nerve in carpal tunnel syndrome: preliminary results , 2008, European Radiology.

[20]  Elizabeth M Tunnicliffe,et al.  Accelerated human cardiac diffusion tensor imaging using simultaneous multislice imaging , 2015, Magnetic resonance in medicine.

[21]  L. White,et al.  Diffusion Tensor Magnetic Resonance Imaging of the Human Calf: Comparison Between 1.5 T and 3.0 T—Preliminary Results , 2008, Investigative radiology.

[22]  Robin M Heidemann,et al.  Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi‐slice imaging , 2005, Magnetic resonance in medicine.

[23]  Zhaohua Ding,et al.  Repeatability of DTI‐based skeletal muscle fiber tracking , 2010, NMR in biomedicine.

[24]  P. Basser,et al.  In vivo fiber tractography using DT‐MRI data , 2000, Magnetic resonance in medicine.

[25]  M. Tarnopolsky,et al.  Diffusion tensor MRI to assess skeletal muscle disruption following eccentric exercise , 2012, Muscle & nerve.

[26]  J. Polimeni,et al.  Blipped‐controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g‐factor penalty , 2012, Magnetic resonance in medicine.

[27]  A. Anderson,et al.  Validation of diffusion tensor MRI‐based muscle fiber tracking , 2002, Magnetic resonance in medicine.

[29]  Usha Sinha,et al.  In vivo diffusion tensor imaging of the human calf muscle , 2006, Journal of magnetic resonance imaging : JMRI.

[30]  Dinesh Kumbhare,et al.  Diffusion tensor imaging in evaluation of human skeletal muscle injury , 2006, Journal of magnetic resonance imaging : JMRI.

[31]  N. Papadakis,et al.  Minimal gradient encoding for robust estimation of diffusion anisotropy. , 2000, Magnetic resonance imaging.

[32]  P. Boesiger,et al.  SENSE: Sensitivity encoding for fast MRI , 1999, Magnetic resonance in medicine.

[33]  M. Ladd,et al.  A diffusion tensor imaging analysis of gender differences in water diffusivity within human skeletal muscle , 2005, NMR in biomedicine.

[34]  P. Basser,et al.  MR diffusion tensor spectroscopy and imaging. , 1994, Biophysical journal.

[35]  Anna Vilanova,et al.  Determination of mouse skeletal muscle architecture using three‐dimensional diffusion tensor imaging , 2005, Magnetic resonance in medicine.

[36]  Reuben H. Fan,et al.  Compartmental relaxation and diffusion tensor imaging measurements in vivo in λ‐carrageenan‐induced edema in rat skeletal muscle , 2008, NMR in biomedicine.

[37]  D. Feinberg,et al.  Contiguous thin multisection MR imaging by two-dimensional Fourier transform techniques. , 1986, Radiology.

[38]  Martijn Froeling,et al.  DTI of human skeletal muscle: the effects of diffusion encoding parameters, signal‐to‐noise ratio and T2 on tensor indices and fiber tracts , 2013, NMR in Biomedicine.

[39]  Bruce M Damon,et al.  Effects of image noise in muscle diffusion tensor (DT)‐MRI assessed using numerical simulations , 2008, Magnetic resonance in medicine.

[40]  Petros Martirosian,et al.  Diffusion tensor imaging of the human calf muscle: distinct changes in fractional anisotropy and mean diffusion due to passive muscle shortening and stretching , 2009, NMR in biomedicine.

[41]  Himanshu Bhat,et al.  Slice accelerated gradient‐echo spin‐echo dynamic susceptibility contrast imaging with blipped CAIPI for increased slice coverage , 2014, Magnetic resonance in medicine.

[42]  Zhaohua Ding,et al.  DTI‐based muscle fiber tracking of the quadriceps mechanism in lateral patellar dislocation , 2009, Journal of magnetic resonance imaging : JMRI.

[43]  Martijn Froeling,et al.  Muscle changes detected with diffusion-tensor imaging after long-distance running. , 2015, Radiology.

[44]  Mara Cercignani,et al.  Twenty‐five pitfalls in the analysis of diffusion MRI data , 2010, NMR in biomedicine.