Human Factor and Usability Testing of a Binocular Optical Coherence Tomography System

Purpose To perform usability testing of a binocular optical coherence tomography (OCT) prototype to predict its function in a clinical setting, and to identify any potential user errors, especially in an elderly and visually impaired population. Methods Forty-five participants with chronic eye disease (mean age 62.7 years) and 15 healthy controls (mean age 53 years) underwent automated eye examination using the prototype. Examination included ‘whole-eye' OCT, ocular motility, visual acuity measurement, perimetry, and pupillometry. Interviews were conducted to assess the subjective appeal and ease of use for this cohort of first-time users. Results All participants completed the full suite of tests. Eighty-one percent of the chronic eye disease group, and 79% of healthy controls, found the prototype easier to use than common technologies, such as smartphones. Overall, 86% described the device to be appealing for use in a clinical setting. There was no statistically significant difference in the total time taken to complete the examination between participants with chronic eye disease (median 702 seconds) and healthy volunteers (median 637 seconds) (P = 0.81). Conclusion On their first use, elderly and visually impaired users completed the automated examination without assistance. Binocular OCT has the potential to perform a comprehensive eye examination in an automated manner, and thus improve the efficiency and quality of eye care. Translational Relevance A usable binocular OCT system has been developed that can be administered in an automated manner. We have identified areas that would benefit from further development to guide the translation of this technology into clinical practice.

[1]  Ian Robinson,et al.  Benefits of and barriers to involving users in medical device technology development and evaluation , 2007, International Journal of Technology Assessment in Health Care.

[2]  Bridget A. Lewis Book Review: Usability Testing of Medical Devices (2nd ed.) , 2018 .

[3]  Chris A Johnson,et al.  A computerized method of visual acuity testing: adaptation of the early treatment of diabetic retinopathy study testing protocol. , 2003, American journal of ophthalmology.

[4]  J. Coughlin,et al.  PERSPECTIVE: Older Adults' Adoption of Technology: An Integrated Approach to Identifying Determinants and Barriers , 2015 .

[5]  P. Netland,et al.  Wait time as a driver of overall patient satisfaction in an ophthalmology clinic , 2013, Clinical ophthalmology.

[6]  G. Ripandelli,et al.  Optical coherence tomography. , 1998, Seminars in ophthalmology.

[7]  Shanhui Fan,et al.  Optical coherence tomography for whole eye segment imaging. , 2012, Optics express.

[8]  Jonathan Kendler,et al.  Usability Testing of Medical Devices , 2010 .

[9]  P. Keane,et al.  Automated Analysis of Vitreous Inflammation Using Spectral-Domain Optical Coherence Tomography. , 2015, Translational vision science & technology.

[10]  Alexander C Walsh,et al.  Binocular optical coherence tomography. , 2011, Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye.

[11]  Chen D. Lu,et al.  Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers , 2012, Biomedical optics express.

[12]  Elizabeth Murphy,et al.  Medical device development: the challenge for ergonomics. , 2008, Applied ergonomics.

[13]  Srinivas R Sadda,et al.  Retinal imaging in the twenty-first century: state of the art and future directions. , 2014, Ophthalmology.

[14]  L. Faulkner Beyond the five-user assumption: Benefits of increased sample sizes in usability testing , 2003, Behavior research methods, instruments, & computers : a journal of the Psychonomic Society, Inc.

[15]  Walter Sermeus,et al.  Exploring the relation between process design and efficiency in high-volume cataract pathways from a lean thinking perspective. , 2011, International journal for quality in health care : journal of the International Society for Quality in Health Care.

[16]  P. Dugel,et al.  Development of an activity-based costing model to evaluate physician office practice profitability. , 2011, Ophthalmology.

[17]  J. Wild,et al.  The influence of the learning effect on automated perimetry in patients with suspected glaucoma , 1989, Acta ophthalmologica.

[18]  D. Menon,et al.  Role of patient and public participation in health technology assessment and coverage decisions , 2011, Expert review of pharmacoeconomics & outcomes research.

[19]  Natalia F. Callaway,et al.  THINKING LEAN: Improving Vitreoretinal Clinic Efficiency by Decentralizing Optical Coherence Tomography , 2016, Retina.

[20]  Paul Tornambe,et al.  Evaluation of economic efficiencies in clinical retina practice: activity-based cost analysis and modeling to determine impacts of changes in patient management , 2011, Clinical ophthalmology.