Analysis of a novel expanded tip wire (ETW) antenna for microwave ablation of cardiac arrhythmias

A novel expanded tip wire (ETW) catheter antenna is proposed for microwave ablation for the treatment of atrial fibrillation (AF). The antenna is designed as an integral part of coaxial cable so that it can be inserted via a 6F catheter. A numerical model based on the rotationally symmetric finite-difference time-domain technique incorporating the generalized perfectly matched layer as the absorbing boundary condition has been utilized to accurately model the interaction between the antenna and the myocardium. Numerical and in-vitro experimental results are presented for specific absorption rate, return loss and heating pattern produced by the antenna. Both numerical modeling and in-vitro experimentation show that the proposed ETW antenna produces a well-defined electric field distribution that provides continuous long and linear lesions for the treatment of AF.

[1]  Pierre Savard,et al.  A finite element model of a microwave catheter for cardiac ablation , 1996 .

[2]  D. Haines,et al.  Microwave catheter ablation of myocardium in vitro. Assessment of the characteristics of tissue heating and injury. , 1994, Circulation.

[3]  Paolo Bernardi,et al.  A 915-MHz antenna for microwave thermal ablation treatment: physical design, computer modeling and experimental measurement , 2001, IEEE Transactions on Biomedical Engineering.

[5]  Zhonghua Wu,et al.  Generalized perfectly matched layer for the absorption of propagating and evanescent waves in lossless and lossy media , 1996 .

[6]  Jean-Pierre Berenger,et al.  A perfectly matched layer for the absorption of electromagnetic waves , 1994 .

[7]  J. C. Lin,et al.  Catheter microwave ablation therapy for cardiac arrhythmias. , 1999, Bioelectromagnetics.

[8]  P. J. Wang,et al.  Development and experimental verification of the wide-aperture catheter-based microwave cardiac ablation antenna , 2000 .

[9]  J.C. Lin,et al.  The cap-choke catheter antenna for microwave ablation treatment , 1996, IEEE Transactions on Biomedical Engineering.

[10]  S. Nattel Therapeutic implications of atrial fibrillation mechanisms: can mechanistic insights be used to improve AF management? , 2002, Cardiovascular research.

[11]  Antonije R. Djordjevic,et al.  Analysis and synthesis of wire-antennas , 1982 .

[12]  A. Rosen,et al.  A method for the in vitro testing of cardiac ablation catheters , 1996 .

[13]  P. Stauffer,et al.  Evaluation of microwave and radio frequency catheter ablation in a myocardium-equivalent phantom model , 1992, IEEE Transactions on Biomedical Engineering.

[14]  Ananda Sanagavarapu Mohan,et al.  Rotationally symmetric FDTD for wideband performance prediction of TM01 DR filters , 2002 .

[15]  L. Roy,et al.  Monopole antennas for microwave catheter ablation , 1996 .

[16]  A. V. Vorst,et al.  Applications of RF/microwaves in medicine , 2002 .

[17]  K. Fujimoto,et al.  Small Antennas , 1987 .

[18]  J. R. Carl,et al.  Microwave catheter design , 1998, IEEE Transactions on Biomedical Engineering.