Using Cherenkov imaging to monitor the match line between photon and electron radiation therapy fields on biological tissue phantoms

Abstract. Significance: Due to patients’ respiratory movement or involuntary body movements during breast cancer radiotherapy, the mismatched adjacent fields in surface exposure regions could result in insufficient dosage or overdose in these regions, which would lead to tissue injury, excessive skin burns, and potential death. Cherenkov luminescence imaging (CLI) could be used to effectively detect the matching information of adjacent radiation fields without extra radiation or invasive imaging. Aim: Our objective was to provide a biological experimental basis for monitoring matching of adjacent radiation fields between photon and electron fields due to introduced shifts during radiotherapy by CLI technique. Approach: A medical accelerator was used to generate photon and electron fields. An industrial camera system was adopted to image the excited CLI signal during irradiation of chicken tissue with yellow (group A and group C experiments) or black color (group B experiment). The following introduced shifts were tested: 10, 5, 2, and 0 mm toward superior or inferior direction. A model was introduced to deal with matching error analysis of adjacent radiation fields due to introduced shifts with adapted plans used to treat neoplasms of the right breast with supraclavicular nodes or internal mammary lymph node. Results: The matching values between photon and electron fields were consistent with the tested introduced shifts during yellow chicken irradiation. In group A, average discrepancies were 0.59  ±  0.35  mm and 0.68  ±  0.37  mm for photon fields and electron fields in anterior/posterior (AP) direction, with 87% and 75% of measurement within 1 mm, respectively. In group C, average discrepancies were 0.80  ±  0.65  mm and 1.07  ±  0.57  mm for oblique photon field with gantry angles of 330 deg and 150 deg, with 66% and 65% of measurement within 1 mm, respectively. The average discrepancies were 0.44  ±  0.30  mm for electron field in the AP direction, with 94% of measurement within 1 mm. The matching error introduced by the proposed method was less than 1.5 mm for AP fields and 2 mm for oblique incidence fields. However, the field matching could not be monitored with black chicken tissue irradiation due to a weak CLI signal that could hardly be extracted from background noise in group B. Conclusions: CLI is demonstrated for the quantitative monitoring of the field match line on light biological tissue phantoms and has potential for monitoring of field matching in surface tissue during breast cancer radiotherapy.

[1]  Brian W. Pogue,et al.  Čerenkov radiation emission and excited luminescence (CREL) sensitivity during external beam radiation therapy: Monte Carlo and tissue oxygenation phantom studies , 2012, Biomedical optics express.

[2]  K. R. Muralidhar,et al.  Commissioning and quality assurance of Calypso four-dimensional target localization system in linear accelerator facility , 2013, Journal of medical physics.

[3]  Jan Grimm,et al.  Cerenkov Luminescence Imaging of Medical Isotopes , 2010, Journal of Nuclear Medicine.

[4]  Peter Hoskin,et al.  Dosimetry and field matching for radiotherapy to the breast and supraclavicular fossa. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  M. Baum,et al.  Targeted radiotherapy for early breast cancer , 2018, The Lancet.

[6]  Brian W Pogue,et al.  Superficial dosimetry imaging of Čerenkov emission in electron beam radiotherapy of phantoms , 2013, Physics in medicine and biology.

[7]  P. Franco,et al.  Three-Dimensional Ultrasound-Based Image-Guided Hypofractionated Radiotherapy for Intermediate-Risk Prostate Cancer: Results of a Consecutive Case Series , 2015, Cancer investigation.

[8]  D. Pasquier,et al.  Surface imaging, laser positioning or volumetric imaging for breast cancer with nodal involvement treated by helical TomoTherapy , 2016, Journal of applied clinical medical physics.

[9]  S R Cherry,et al.  Optical imaging of Cerenkov light generation from positron-emitting radiotracers , 2009, Physics in medicine and biology.

[10]  Sergei A. Vinogradov,et al.  Cherenkov excited phosphorescence-based pO2 estimation during multi-beam radiation therapy: phantom and simulation studies , 2014, Physics in medicine and biology.

[11]  M. Roumeliotis,et al.  Including internal mammary lymph nodes in radiation therapy for synchronous bilateral breast cancer: an international survey of treatment technique and clinical priorities , 2018, Breast Cancer Research and Treatment.

[12]  C. Wuu,et al.  An investigation of clinical treatment field delivery verification using cherenkov imaging: IMRT positioning shifts and field matching , 2018, Medical physics.

[13]  Nicola Belcari,et al.  Cerenkov luminescence imaging: physics principles and potential applications in biomedical sciences , 2017, EJNMMI Physics.

[14]  J. O'Daniel,et al.  Accuracy and efficiency of image-guided radiation therapy (IGRT) for preoperative partial breast radiosurgery. , 2020, Journal of radiosurgery and SBRT.

[15]  Andrea Sbarbati,et al.  Cerenkov radiation allows in vivo optical imaging of positron emitting radiotracers , 2010, Physics in medicine and biology.

[16]  Brian W Pogue,et al.  Camera selection for real-time in vivo radiation treatment verification systems using Cherenkov imaging. , 2015, Medical physics.

[17]  Louis Archambault,et al.  Transient noise characterization and filtration in CCD cameras exposed to stray radiation from a medical linear accelerator. , 2008, Medical physics.

[18]  P. Jacobs,et al.  Intrafraction monitoring of prostate motion during radiotherapy using the Clarity® Autoscan Transperineal Ultrasound (TPUS) system. , 2017, Radiography.

[19]  Timothy Solberg,et al.  Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy. , 2007, International journal of radiation oncology, biology, physics.

[20]  Brian W. Pogue,et al.  A GAMOS plug-in for GEANT4 based Monte Carlo simulation of radiation-induced light transport in biological media , 2013, Biomedical optics express.

[21]  P. A. Čerenkov Visible radiation produced by electrons moving in a medium with velocities exceeding that of light , 1937 .

[22]  J. Yarnold Selective avoidance of lymphatic irradiation in the conservative management of breast cancer. , 1984, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  Christopher G Willett,et al.  Quantification of respiration-induced abdominal tumor motion and its impact on IMRT dose distributions. , 2004, International journal of radiation oncology, biology, physics.

[24]  D van Eeden,et al.  Set-up error validation with EPID images: Measurements vs Egs_cbct simulation. , 2019, Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology.

[25]  A. Pacifici,et al.  How accurate is CBCT in measuring bone density? A comparative CBCT-CT in vitro study. , 2014, Clinical implant dentistry and related research.

[26]  Davide Fontanarosa,et al.  Review of ultrasound image guidance in external beam radiotherapy part II: intra-fraction motion management and novel applications , 2016, Physics in medicine and biology.

[27]  Brian W Pogue,et al.  Superficial dosimetry imaging based on Čerenkov emission for external beam radiotherapy with megavoltage x-ray beam. , 2013, Medical physics.

[28]  Brian W Pogue,et al.  Cherenkoscopy based patient positioning validation and movement tracking during post-lumpectomy whole breast radiation therapy , 2015, Physics in medicine and biology.

[29]  Sonja Dieterich,et al.  Can a commercially available EPID dosimetry system detect small daily patient setup errors for cranial IMRT/SRS? , 2017, Practical radiation oncology.

[30]  Scott C Davis,et al.  Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms. , 2012, Optics letters.

[31]  Brian W Pogue,et al.  Real-time in vivo Cherenkoscopy imaging during external beam radiation therapy , 2013, Journal of biomedical optics.

[32]  R. Austin,et al.  Accuracy of CBCT for volumetric measurement of simulated periapical lesions. , 2013, International endodontic journal.

[33]  Brian W Pogue,et al.  Cherenkov video imaging allows for the first visualization of radiation therapy in real time. , 2014, International journal of radiation oncology, biology, physics.

[34]  C. Belka,et al.  Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the CatalystTM/SentinelTM system for deep inspiration breath-hold (DIBH) , 2016, Radiation Oncology.

[35]  Brian W Pogue,et al.  Beam and tissue factors affecting Cherenkov image intensity for quantitative entrance and exit dosimetry on human tissue , 2017, Journal of biophotonics.

[36]  John T. Wei,et al.  Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. , 2006, International journal of radiation oncology, biology, physics.

[37]  C. Belka,et al.  Real-time intra-fraction motion management in breast cancer radiotherapy: analysis of 2028 treatment sessions , 2018, Radiation Oncology.