Force decay and dimensional changes of thermoplastic and novel thermoset elastomeric ligatures.

OBJECTIVES To compare over a period of 8 weeks (1) the force decay and (2) the dimensional changes between thermoplastic (TP) and thermoset (TS) elastomeric ligatures. MATERIALS AND METHODS TP and TS elastomeric ligatures were obtained from Rocky Mountain Orthodontics™. The TS ligatures were custom made specifically for this study. The sample included 72 clear TP and 72 clear TS elastomeric ligatures. The experiment was performed in a simulated oral environment (pH of 6.75) at 37°C. The remaining forces and the dimensional changes were measured at different time points over a period of 8 weeks. RESULTS Student's t-tests revealed significant differences in percent force loss, percent change in outer diameter, percent change in inner diameter, and percent change in wall thickness between TP and TS elastomeric ligatures across all time points (P < .001). The difference in percent change in width between TP and TS elastomeric ligatures was not significant at all time points (P > .05). The mean difference in force loss between TP and TS across all time points was 22.91%. The TP and TS specimens exhibited 93.04% and 77.41% force loss, respectively, at the 28th day. CONCLUSIONS This novel TS elastomeric ligature showed significantly less force decay and dimensional changes over time; therefore, it might be superior during initial leveling and aligning and during finishing stages. Using a transfer jig to prevent relaxation of the specimens before force measurement showed that force decay of commercially available elastomeric ligatures was greater than that described in previous publications.

[1]  E. Begole,et al.  Force decay evaluation of thermoplastic and thermoset elastomeric chains: A mechanical design comparison. , 2014, The Angle orthodontist.

[2]  J. Hagan,et al.  Force Decay of Elastomeric Chains – a Mechanical Design and Product Comparison Study , 2011, Journal of orthodontics.

[3]  S. Rosenstein,et al.  Force decay of elastomeric ligatures: influence on unloading force compared to self-ligation. , 2009, The Angle orthodontist.

[4]  A. Kuijpers-Jagtman,et al.  Optimum force magnitude for orthodontic tooth movement: a systematic literature review. , 2009, The Angle orthodontist.

[5]  R. Kusy,et al.  Force application and decay characteristics of untreated and treated polyurethane elastomeric chains. , 2009, The Angle orthodontist.

[6]  C. Flores‐Mir,et al.  In vivo comparison of force decay between injection molded and die-cut stamped elastomers. , 2006, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[7]  R. Ricketts The wisdom of the bioprogressive philosophy. , 1998, Seminars in orthodontics.

[8]  E. D. Rekow,et al.  Force degradation in elastomeric chains. , 1997, Seminars in orthodontics.

[9]  W. N. Wang,et al.  Force decay of elastomeric chain--a serial study. Part II. , 1993, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[10]  J. P. Ferriter,et al.  The effect of hydrogen ion concentration on the force-degradation rate of orthodontic polyurethane chain elastics. , 1990, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[11]  L. Nunez,et al.  Observations on the Elastic Behavior of a Synthetic Orthodontic Elastomer , 1990, Journal of dental research.

[12]  W. Rock,et al.  The Effect of Bracket Type and Ligation Method upon Forces Exerted by Orthodontic Archwires , 1989, British journal of orthodontics.

[13]  W. Rock,et al.  Forces Exerted by Orthodontic Aligning Archwires , 1988, British journal of orthodontics.

[14]  J. Powers,et al.  Effects of brackets and ties on stiffness of an arch wire. , 1987, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.

[15]  B. Ingervall,et al.  Laboratory and intra-oral tests of the degradation of elastic chains. , 1986, European journal of orthodontics.

[16]  R. Shaye,et al.  Force degradation of orthodontic elastomeric chains--a product comparison study. , 1985, American journal of orthodontics.

[17]  R. Nikolai,et al.  Relaxation of Orthodontic Elastomeric Chains and Modules In Vitro and In Vivo , 1978, Journal of dental research.

[18]  G. Andreasen,et al.  Comparison of alastik chains with elastics involved with intra-arch molar to molar forces. , 1971, The Angle orthodontist.

[19]  G. Andreasen,et al.  A comparison of time related forces between plastic alastiks and latex elastics. , 1970, The Angle orthodontist.

[20]  G. Andreasen,et al.  Comparison of alastik chains of elastics involved with intra-arch molar-to-molar forces. , 1970, American journal of orthodontics.

[21]  L Lorton,et al.  Force decay and deformation of orthodontic elastomeric ligatures. , 1997, American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics.